Comparison of keratometric and topographic cylinder and axis measurements on normal corneas with low astigmatism.

C H Karabatsas1, I Papaefthymiou, I M Aslanides, D Z Chatzoulis.   

Abstract

PURPOSE: To evaluate agreement in measurements of astigmatic axis power and location between keratometry and computer assisted videokeratography (corneal topography) on normal corneas with less than 1.50 D of idiopathic astigmatism.
METHODS: Keratometric readings with the 10 SL/O Zeiss ophthalmometer and corneal topographic maps with the TMS-1 were obtained by two independent examiners on 32 normal corneas. Measurement agreement between the two instruments was evaluated in regard to steep and flat meridian power and location, and in astigmatism magnitude (D).
RESULTS: The limits of agreement (d-2 SD to d+2 SD) between the two instruments were found to be broad for clinical purposes in measuring the steep meridian power (-0.16 to -1.20 D), flat meridian power (0.43 to -1.25 D), and astigmatism (0.60 to -1.12 D). A constant bias of the TMS-1 towards the 10 SL/O Zeiss ophthalmometer was found, in measuring steeper both principal meridians and higher amount of astigmatism. Mean location difference was 19 degrees (+/-190) for the steep meridian and 17 degrees (+/-20 degrees) for the flat meridian.
CONCLUSIONS: Despite the differences seen in measurements between the 10 SL/O ophthalmometer and the TMS-1, these differences may be clinically small enough for the methods to be used interchangeably in measuring only the magnitude of astigmatism on normal corneas. However, the disagreement in astigmatism axes is too great to be ignored.

Entities:  

Mesh:

Year:  2005        PMID: 15751233     DOI: 10.1177/112067210501500102

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  4 in total

1.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

2.  Color light-emitting diode reflection topography: validation of keratometric repeatability in a large sample of wide cylindrical-range corneas.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2015-02-05

3.  A comparison of IOLMaster 500 and IOLMaster 700 in the measurement of ocular biometric parameters in cataract patients.

Authors:  Jianhong Jiang; Xiaojing Pan; Mingming Zhou; Xiaoyun Wang; Hai Zhu; Dongfang Li
Journal:  Sci Rep       Date:  2022-07-27       Impact factor: 4.996

4.  Distribution and Repeatability of Corneal Astigmatism Measurements (Magnitude and Axis) Evaluated With Color Light Emitting Diode Reflection Topography.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Cornea       Date:  2015-08       Impact factor: 2.651

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.