Clinical evaluation of keratometry and computerised videokeratography: intraobserver and interobserver variability on normal and astigmatic corneas.

C H Karabatsas1, S D Cook, J Papaefthymiou, P Turner, J M Sparrow.   

Abstract

AIMS: To evaluate intra- and interobserver variability in measurements on normal and astigmatic corneas with keratometry and computerized videokeratography.
METHODS: Keratometric readings with the 10 SL/O Zeiss keratometer and topographic maps with the TMS-1 were obtained by two independent examiners on 32 normal and 33 postkeratoplasty corneas. Inter- and intraobserver coefficients of variability (COR) for measurements of steep and flat meridian power and location, in addition to the magnitude of astigmatism, were assessed.
RESULTS: Compared with TMS-1, the 10 SL/O keratometer showed a superior repeatability in measuring normal corneas (intraobserver COR for keratometry and TMS-1 respectively: 0.22 and 0.30 D for steep meridian power; 0.18 and 0.44 D for flat meridian power; 0.26 and 0.40 D for astigmatism; 5 degrees and 26 degrees for steep meridian location; 5 degrees and 13 degrees for flat meridian location). Astigmatism intraobserver COR (0.20 D and 0.26 D for the two observers) and interobserver COR (0.28 D) of the keratometer for normal corneas was very good and not affected by observers' experience. Repeatability of the TMS-1 on normal corneas was found to be: (a) observer related, and (b) astigmatism related. A novice observer showed a much greater COR (1.62 D for astigmatism, 30 degrees for flat meridian location) compared with the experienced examiner (0.40 D for astigmatism, 13 degrees for flat meridian location). Higher deviation scores were observed for corneas with higher astigmatism. For the postkeratoplasty corneas, again the keratometer achieved superior reproducibility (astigmatism interobserver COR 1.12 D for keratometry, 4.06 D for TMS-1; steep meridian location interobserver COR 10 degrees for keratometry, 34 degrees for TMS-1).
CONCLUSION: Keratometric readings are more reproducible than topographic data both for normal and postkeratoplasty corneas. The two instruments should not be used interchangeably especially on highly astigmatic corneas. For the TMS-1, users with the same level of experience should be employed in clinical or experimental studies.

Entities:  

Mesh:

Year:  1998        PMID: 9797664      PMCID: PMC1722645          DOI: 10.1136/bjo.82.6.637

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  15 in total

1.  The validity of some existing methods of measuring corneal contour compared with suggested new methods.

Authors:  J STONE
Journal:  Br J Physiol Opt       Date:  1962 Oct-Dec

2.  Analysis of the effects of astigmatism and misalignment on corneal surface reconstruction from photokeratoscopic data.

Authors:  J Y Wang; D A Rice; S D Klyce
Journal:  Refract Corneal Surg       Date:  1991 Mar-Apr

3.  A comparison of the EH-270 corneal topographer with conventional keratometry.

Authors:  L J Davis; M S Dresner
Journal:  CLAO J       Date:  1991-07

4.  3-D corneal modeling system.

Authors:  R J Mammone; M Gersten; D J Gormley; R S Koplin; V L Lubkin
Journal:  IEEE Trans Biomed Eng       Date:  1990-01       Impact factor: 4.538

5.  Reproducibility of normal corneal power measurements with a keratometer, photokeratoscope, and video imaging system.

Authors:  S B Hannush; S L Crawford; G O Waring; M C Gemmill; M J Lynn; A Nizam
Journal:  Arch Ophthalmol       Date:  1990-04

6.  Accuracy and precision of keratometry, photokeratoscopy, and corneal modeling on calibrated steel balls.

Authors:  S B Hannush; S L Crawford; G O Waring; M C Gemmill; M J Lynn; A Nizam
Journal:  Arch Ophthalmol       Date:  1989-08

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Evaluating the reproducibility of topography systems on spherical surfaces.

Authors:  L J Maguire; S E Wilson; J J Camp; S Verity
Journal:  Arch Ophthalmol       Date:  1993-02

9.  Computer-assisted corneal topography: accuracy and reproducibility of the topographic modeling system.

Authors:  J M Legeais; Q Ren; G Simon; J M Parel
Journal:  Refract Corneal Surg       Date:  1993 Sep-Oct

10.  The effect of poor fixation on computer-assisted topographic corneal analysis. Pseudokeratoconus.

Authors:  R E Hubbe; G N Foulks
Journal:  Ophthalmology       Date:  1994-10       Impact factor: 12.079

View more
  2 in total

1.  Color light-emitting diode reflection topography: validation of keratometric repeatability in a large sample of wide cylindrical-range corneas.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2015-02-05

2.  Distribution and Repeatability of Corneal Astigmatism Measurements (Magnitude and Axis) Evaluated With Color Light Emitting Diode Reflection Topography.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Cornea       Date:  2015-08       Impact factor: 2.651

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.