PURPOSE: To evaluate the repeatability and reproducibility of a double-pass instrument (OQASII, Visiomereics SL, Spain), which objectively measures overall optical quality of the human eyes.Entities:
Mesh:
Year: 2015 PMID: 25646846 PMCID: PMC4315520 DOI: 10.1371/journal.pone.0117587
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Intra-observer repeatability among three tests in each session for the parameters provided by OQASII.
| Mean value | Sw | Precision | CV(%) | ICC | |
|---|---|---|---|---|---|
| MTF cutoff(cpd) | 39.32±9.75 | 2.18 | 4.27 | 5.96 | 0.94 |
| Strehl ratio | 0.22±0.06 | 0.02 | 0.03 | 7.98 | 0.88 |
| OV100% | 1.31±0.33 | 0.07 | 0.14 | 5.94 | 0.94 |
| OV20% | 1.33±0.39 | 0.09 | 0.18 | 7.22 | 0.92 |
| OV9% | 1.33±0.41 | 0.10 | 0.20 | 8.02 | 0.90 |
| OSI | 0.60±0.42 | 0.05 | 0.09 | 9.49 | 0.98 |
Results of inter-observer reproducibility of OQASII.
| Mean Difference | Sw | Precision | CV(%) | ICC | |
|---|---|---|---|---|---|
| MTF cutoff(cpd) | 0.37±3.27 | 1.90 | 3.73 | 5.33 | 0.95 |
| Strehl ratio | 0.00±0.03 | 0.02 | 0.03 | 6.98 | 0.90 |
| OV100% | 0.01±0.11 | 0.06 | 0.12 | 5.30 | 0.95 |
| OV20% | 0.01±0.16 | 0.09 | 0.17 | 6.52 | 0.93 |
| OV9% | 0.01±0.01 | 0.11 | 0.22 | 8.56 | 0.88 |
| OSI | -0.01±0.11 | 0.05 | 0.10 | 9.55 | 0.97 |
Results of inter-visit reproducibility of OQASII.
| Mean Difference | Sw | Precision | CV(%) | ICC | |
|---|---|---|---|---|---|
| MTF cut off(cpd) | 0.43±3.57 | 2.05 | 4.01 | 5.59 | 0.94 |
| Strehl ratio | 0.00±0.03 | 0.02 | 0.03 | 7.44 | 0.88 |
| OV100% | 0.01±0.12 | 0.07 | 0.13 | 5.50 | 0.94 |
| OV20% | 0.01±0.16 | 0.09 | 0.17 | 6.78 | 0.92 |
| OV9% | 0.00±0.19 | 0.10 | 0.19 | 7.09 | 0.91 |
| OSI | -0.01±0.12 | 0.06 | 0.12 | 11.06 | 0.96 |
Fig 1Bland-Altman plots of the mean differences between session A and B (inter-observer) were compared for parameters: MTF cut off (A), Strehl ratio (B), OV 100% (C), OV 20% (D), OV 9% (E), and OSI (F).
The 95% limits of agreement are shown with dashed lines (ranged from-6.04 to 6.78cpd, -0.05 to 0.05, -0.20 to 0.23, -0.29 to 0.32, -0.40 to 0.42, -0.23 to 0.21, respectively), and the solid line represents the mean difference between these measurements.
Fig 2Bland-Altman plots of the mean differences between session A and C (inter-visit) were compared for parameters: MTF cut off (A), Strehl ratio (B), OV 100% (C), OV 20% (D), OV 9% (E) and OSI (F).
The 95% limits of agreement are shown with dashed lines (ranged from-6.56 to 7.42cpd, -0.06 to 0.06, -0.22 to 0.24, -0.30 to 0.32, -0.35 to 0.34, -0.24 to 0.23, respectively), and the solid line represents the mean difference between these measurements.