Literature DB >> 16129291

Retinal image quality after microincision intraocular lens implantation.

Jorge L Alió1, Patricia Schimchak, Robert Montés-Micó, Ahmed Galal.   

Abstract

PURPOSE: To evaluate the modulation-transfer function (MTF) in eyes implanted with a conventional intraocular lens (IOL) and 2 IOLs designed for microincision cataract surgery (MICS).
SETTING: Research, Development, and Innovation Department, Instituto Oftalmológico de Alicante, Alicante, Spain.
METHODS: This prospective nonrandomized consecutive series comprised 30 eyes implanted with 1 of the following IOLs: conventional acrylic foldable (AcrySof MA60BM, Alcon Laboratories) or the UltraChoice 1.0 ThinOptX (ThinOptX Inc.) or the Acri. Smart 48S (Acri.Tech) MICS IOL. The 0.5 MTF and 0.1 MTF following MICS were calculated 3 months after implantation with the Optical Quality Analysis System (OQAS) for a 5 mm pupil. The differences were statistically analyzed with the Mann-Whitney U test.
RESULTS: The values of 0.5 MTF for AcrySof, UltraChoice, ThinOptX, and Acri. Smart IOLs were, respectively, 2.647 cycles per degree (cpd) +/- 0.833 (SD), 2.601 +/- 0.986 cpd, and 3.453 +/- 0.778 cpd. The mean 0.1 MTF values for the same IOLs were 8.720 +/- 3.074 cpd, 8.814 +/- 4.380 cpd, and 11.418 +/- 2.574 cpd, respectively. Statistical analysis did not show significant differences in 0.5 MTF and 0.1 MTF between the conventional IOL and MICS IOLs.
CONCLUSIONS: Microincision cataract surgery IOLs showed excellent MTF performance when implanted after cataract surgery, equal to that of conventional IOLs. An in vivo MTF study may be an excellent option to evaluate IOL performance in the eye.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16129291     DOI: 10.1016/j.jcrs.2005.01.026

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  15 in total

1.  [Microincision intraocular lens with plate haptic design. Evaluation of rotational stability and centering of a microincision intraocular lens with plate haptic design in 12-19 months of follow-up].

Authors:  W Wehner
Journal:  Ophthalmologe       Date:  2007-05       Impact factor: 1.059

2.  Non-cycloplegic spherical equivalent refraction in adults: comparison of the double-pass system, retinoscopy, subjective refraction and a table-mounted autorefractor.

Authors:  Meritxell Vilaseca; Montserrat Arjona; Jaume Pujol; Elvira Peris; Vanessa Martínez
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

3.  [Intraocular lenses for microincisional cataract surgery].

Authors:  T Kohnen; O K Klaproth
Journal:  Ophthalmologe       Date:  2010-02       Impact factor: 1.059

4.  Visual quality assessment after presbyopic laser in-situ keratomileusis.

Authors:  Dong Hui Lim; Eui-Sang Chung; Myoung Joon Kim; Tae-Young Chung
Journal:  Int J Ophthalmol       Date:  2018-03-18       Impact factor: 1.779

5.  Early outcomes of vision and objective visual quality analysis after cataract surgery with trifocal intraocular lens implantation.

Authors:  Mei-Fang Chu; Na Hui; Cong-Yi Wang; Lei Yu; Bo Ma; Yan Li; Cheng Pei
Journal:  Int J Ophthalmol       Date:  2019-10-18       Impact factor: 1.779

6.  Optical quality of toric intraocular lens implantation in cataract surgery.

Authors:  Xian-Wen Xiao; Jing Hao; Hong Zhang; Fang Tian
Journal:  Int J Ophthalmol       Date:  2015-02-18       Impact factor: 1.779

7.  Corneal Optical Quality Following Sub 1.8 mm Micro-Incision Cataract Surgery vs. 2.2 mm Mini-Incision Coaxial Phacoemulsification.

Authors:  Jorge L Alió; Bassam Elkady; Dolores Ortiz
Journal:  Middle East Afr J Ophthalmol       Date:  2010-01

8.  Wavefront analysis and modulation transfer function of three multifocal intraocular lenses.

Authors:  Marcony R Santhiago; Marcelo V Netto; Jackson Barreto; Beatriz A F Gomes; Arthur Schaefer; Newton Kara-Junior
Journal:  Indian J Ophthalmol       Date:  2010 Mar-Apr       Impact factor: 1.848

Review 9.  Cataract surgery with implantation of an artificial lens.

Authors:  Thomas Kohnen; Martin Baumeister; Daniel Kook; Oliver K Klaproth; Christian Ohrloff
Journal:  Dtsch Arztebl Int       Date:  2009-10-23       Impact factor: 5.594

10.  A contralateral eye study comparing apodized diffractive and full diffractive lenses: wavefront analysis and distance and near uncorrected visual acuity.

Authors:  Marcony Rodrigues de Santhiago; Marcelo Vieira Netto; Jackson Barreto; Beatriz de Abreu Fiuza Gomes; Arthur Schaefer; Newton Kara-Junior
Journal:  Clinics (Sao Paulo)       Date:  2009       Impact factor: 2.365