Gairik Kundu1, Rohit Shetty1, Pooja Khamar2, Sneha Gupta2, Ritika Mullick1, Vaitheeswaran L Ganesan3, Sharon D'Souza1.
Abstract
Purpose: To assess impact of tear optics on repeatability of a Scheimpflug device with a Hartmann Shack aberrometer and a ray tracing aberrometer.Entities:
Keywords: Aberrations; OSI; Pentacam AXL wave; iTrace; repeatability
Mesh:
Year: 2022 PMID: 35326004 PMCID: PMC9240491 DOI: 10.4103/ijo.IJO_2153_21
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 2.969
Figure 1Representative image of tear film analysis with an optical quality analysis system (OQAS). Objective scatter index (OSI) is shown over a period of time and with blinks noted. The OSI is less than 1 suggestive of good tear film optics
Figure 2Representative image of tear film analysis with an optical quality analysis system (OQAS) with the objective scatter index (OSI) more than 1 suggestive of poor tear film optics
Mean and the standard error of mean values of Schirmer and TBUT in both healthy and postrefractive eyes with a mean OSI <1 and mean OSI ≥1
| Healthy eyes ( | Post Refractive eyes ( | |||
|---|---|---|---|---|
|
|
| |||
| Mean OSI <1 ( | Mean OSI ≥1 ( | Mean OSI <1 ( | Mean OSI ≥1 ( | |
| TBUT (s) | 9±0.21 | 7.9±0.48 | 8.6±0.31 | 5.1±0.24 |
| Schirmer (mm) | 25±0.47 | 20±0.98 | 27±0.34 | 22±0.47 |
TBUT – Tear film break-up time, OSI – objective scatter index
Figure 3Graph showing negative correlation between tear film break-up time (TBUT) and mean objective scatter index (OSI)
Repeatability (with 95% CI) of wavefront refraction, flat keratometry, steep keratometry, TCT, RMS HOA, RMS LOA, RMS COMA, and SA derived from Pentacam AXL wave in healthy eyes with those having a mean OSI<1
| Sw | TRT | COV (%) | ICC | |
|---|---|---|---|---|
| Sphere | 0.65 (0.40-0.84) | 1.80 (1.48-2.44) | 10.2 (8.6-13.9) | 0.89 |
| Cylinder | 0.66 (0.43-0.86) | 1.82 (1.52-2.39) | 14.5 (11.02-17.9) | 0.88 |
| Axis | 19.2 (13.7-23.5) | 53.18 (41.6-64.3) | 19.7 (15.6-24.3) | 0.83 |
| Flat keratometry | 0.22 (0.19-0.27) | 0.60 (0.51-0.73) | 0.2 (0.1-0.6) | 0.98 |
| Steep keratometry | 0.11 (0.08-0.13) | 0.31 (0.25-0.36) | 1.1 (0.5-1.8) | 0.96 |
| TCT measurement | 9.23 (7.11-10.54) | 25.56 (21.36-28.50) | 2.5 (0.2-3.6) | 0.97 |
| RMS HOA | 0.024 (0.016-0.033) | 0.066 (0.042-0.079) | 8.3 (6.4-11.7) | 0.88 |
| RMS LOA | 0.112 (0.07-0.23) | 0.31 (0.1-0.46) | 4.4 (2.6-6.9) | 0.92 |
| RMS COMA | 0.016 (0.013-0.026) | 0.044 (0.024-0.067) | 5.5 (3.6-7.8) | 0.89 |
| SA | 0.01387 (0.0093-0.031) | 0.038 (0.023-0.054) | 5.1 (3.1-7.6) | 0.91 |
TCT – thinnest corneal thickness, RMS HOA – root means square of higher-order aberrations, RMS LOA – Root mean square of lower-order aberrations, RMS COMA – Root mean square of coma, SA – Spherical aberration, Sw – within-subject standard deviation, TRT – test–retest variability, COV – within-subject coefficient of variation, ICC – intraclass correlation coefficient, CI – confidence interval, OSI – objective scatter index
Comparison of repeatability of total aberrations derived from Pentacam AXL wave and iTrace in healthy eyes with a mean OSI <1
| Parameter | Pentacam AXL wave | iTrace |
|---|---|---|
| RMS COMA | ||
| Sw (D) | 0.016 (0.013-0.026) | 0.023 (0.01-0.045) |
| TRT (D) | 0.044 (0.024-0.067) | 0.063 (0.041-0.089) |
| COV (%) | 5.5 (3.6-7.8) | 6.8 (4.8-9.6) |
| ICC | 0.89 | 0.84 |
| RMS HOA | ||
| Sw (D) | 0.024 (0.016-0.033) | 0.034 (0.022-0.048) |
| TRT (D) | 0.066 (0.042-0.079) | 0.094 (0.045-0.14) |
| COV (%) | 8.3 (6.4-11.7) | 9.7 (6.32-12.2) |
| ICC | 0.88 | 0.82 |
| RMS LOA | ||
| Sw (D) | 0.112 (0.07-0.23) | 0.159 (0.07-0.22) |
| TRT (D) | 0.31 (0.1-0.46) | 0.44 (0.31-0.57) |
| COV (%) | 4.4 (2.6-6.9) | 6.2 (3.3-8.7) |
| ICC | 0.92 | 0.89 |
| SA | ||
| Sw (D) | 0.01387 (0.0093-0.031) | 0.023 (0.012-0.038) |
| TRT (D) | 0.038 (0.023-0.054) | 0.064 (0.039-0.080) |
| COV (%) | 5.1 (3.1-7.6) | 7.04 (4.63-9.25) |
| ICC | 0.91 | 0.88 |
RMS HOA – Root means square of higher-order aberrations, RMS LOA – Root mean square of lower-order aberrations, RMS COMA – Root mean square of coma, SA – Spherical aberration, Sw – within-subject standard deviation, TRT – test–retest variability, COV – within-subject coefficient of variation, ICC – Intraclass Correlation Coefficient, OSI – objective scatter index
Repeatability (with 95% CI) of wavefront refraction, flat keratometry, steep keratometry, TCT, RMS HOA, RMS LOA, RMS COMA, and SA derived from Pentacam AXL wave in healthy eyes with those having a mean OSI ≥1
| Sw | TRT | COV (%) | ICC | |
|---|---|---|---|---|
| Sphere | 0.78 (0.38-0.92) | 2.1 (1.34-3.24) | 12.2 (8.4-14.6) | 0.88 |
| Cylinder | 0.79 (0.40-0.96) | 2.2 (1.62-3.5) | 16.5 (10.1-19.4) | 0.86 |
| Axis | 23.2 (11.8-24.5) | 64.2 (38.5-72.3) | 20.3 (14.9-26.6) | 0.82 |
| Flat keratometry | 0.27 (0.16-0.38) | 0.74 (0.42-0.83) | 0.3 (0.2-0.6) | 0.93 |
| Steep keratometry | 0.16 (0.06-0.21) | 0.44 (0.22-0.52) | 1.3 (0.4-2.4) | 0.94 |
| TCT measurement | 10.1 (6.32-13.29) | 27.9 (20.68-30.4) | 3.2 (0.9-4.9) | 0.90 |
| RMS HOA | 0.037 (0.013-0.049) | 0.102 (0.036-0.163) | 9.4 (5.9-13.2) | 0.77 |
| RMS LOA | 0.152 (0.09-0.38) | 0.42 (0.08-0.53) | 4.9 (2.5-7.8) | 0.84 |
| RMS COMA | 0.028 (0.014-0.041) | 0.077 (0.022-0.089) | 6.1 (3.3-8.9) | 0.79 |
| SA | 0.0192 (0.0089-0.039) | 0.052 (0.019-0.069) | 5.9 (3.4-8.6) | 0.83 |
TCT – thinnest corneal thickness, RMS HOA – Root means square of higher-order aberrations, RMS LOA – Root mean square of lower-order aberrations, RMS COMA – Root mean square of coma, SA – Spherical aberration, Sw – within-subject standard deviation, TRT – test–retest variability, COV – within-subject coefficient of variation, ICC – intraclass correlation coefficient, CI – confidence interval, OSI – objective scatter index
Comparison of repeatability of total aberrations derived from Pentacam AXL wave and iTrace in healthy eyes with a mean OSI ≥1
| Parameter | Pentacam AXL wave | iTrace |
|---|---|---|
| RMS COMA | ||
| Sw (D) | 0.028 (0.014-0.041) | 0.032 (0.02-0.056) |
| TRT (D) | 0.077 (0.022-0.089) | 0.063 (0.035-0.092) |
| COV (%) | 6.1 (3.3-8.9) | 7.3 (4.7-9.9) |
| ICC | 0.79 | 0.76 |
| RMS HOA | ||
| Sw (D) | 0.037 (0.013-0.049) | 0.048 (0.024-0.061) |
| TRT (D) | 0.102 (0.036-0.163) | 0.094 (0.049-0.17) |
| COV (%) | 9.4 (5.9-13.2) | 10.2 (6.5-13.8) |
| ICC | 0.77 | 0.74 |
| RMS LOA | ||
| Sw (D) | 0.152 (0.09-0.38) | 0.18 (0.07-0.22) |
| TRT (D) | 0.42 (0.08-0.53) | 0.44 (0.11-0.62) |
| COV (%) | 4.9 (2.5-7.8) | 7.3 (3.5-9.8) |
| ICC | 0.84 | 0.81 |
| SA | ||
| Sw (D) | 0.0192 (0.0089-0.039) | 0.031 (0.012-0.043) |
| TRT (D) | 0.052 (0.019-0.069) | 0.064 (0.034-0.092) |
| COV (%) | 5.9 (3.4-8.6) | 7.8 (4.2-10.6) |
| ICC | 0.83 | 0.79 |
RMS HOA – Root means square of higher-order aberrations, RMS LOA – Root mean square of lower-order aberrations, RMS COMA – Root mean square of coma, SA – Spherical aberration, Sw – within-subject standard deviation, TRT – test–retest variability, COV – within-subject coefficient of variation, ICC – intraclass correlation coefficient, OSI – objective scatter index
Repeatability (with 95% CI) of flat keratometry, steep keratometry, TCT, RMS HOA, RMS LOA, RMS COMA, and SA derived from Pentacam AXL wave in postrefractive eyes with a mean OSI<1
| Sw | TRT | COV (%) | ICC | |
|---|---|---|---|---|
| Flat keratometry | 0.27 (0.12-0.32) | 0.74 (0.51-0.83) | 0.3 (0.0-0.8) | 0.94 |
| Steep keratometry | 0.14 (0.05-0.15) | 0.38 (0.25-0.47 | 1.4 (0.3-1.9) | 0.92 |
| TCT measurement | 10.12 (7.64-10.92) | 28.03 (21.36-31.6) | 2.9 (0.1-4.2) | 0.94 |
| RMS HOA | 0.031 (0.013-0.043) | 0.085 (0.042-0.103) | 9.1 (5.9-12.2) | 0.83 |
| RMS LOA | 0.109 (0.06-0.28) | 0.31 (0.1-0.49) | 4.5 (2.8-7.2) | 0.90 |
| RMS COMA | 0.022 (0.014-0.039) | 0.055 (0.022-0.072) | 5.7 (3.4-8.2) | 0.87 |
| SA | 0.015 (0.008-0.034) | 0.041 (0.021-0.059) | 5.4 (2.8-7.9) | 0.89 |
TCT – thinnest corneal thickness, RMS HOA – Root means square of higher-order aberrations, RMS LOA – Root mean square of lower-order aberrations, RMS COMA – Root mean square of coma, SA – Spherical aberration, Sw – within-subject standard deviation, TRT – test–retest variability, COV – within-subject coefficient of variation, ICC – intraclass correlation coefficient , CI – confidence interval, OSI – objective scatter index
Comparison of repeatability of total aberrations derived from Pentacam AXL wave and iTrace in postrefractive eyes with a mean OSI <1
| Parameter | Pentacam AXL wave | iTrace |
|---|---|---|
| RMS COMA | ||
| Sw (D) | 0.022 (0.014-0.039) | 0.034 (0.4-0.059) |
| TRT (D) | 0.055 (0.022-0.072) | 0.094 (0.031-0.12) |
| COV (%) | 5.7 (3.4-8.2) | 7.1 (4.2-10.4) |
| ICC | 0.87 | 0.85 |
| RMS HOA | ||
| Sw (D) | 0.031 (0.013-0.043) | 0.045 (0.02-0.062) |
| TRT (D) | 0.085 (0.042-0.103) | 0.12 (0.039-0.17) |
| COV (%) | 9.1 (5.9-12.2) | 9.9 (6.1-13.4) |
| ICC | 0.83 | 0.80 |
| RMS LOA | ||
| Sw (D) | 0.109 (0.06-0.28) | 0.18 (0.06-0.26) |
| TRT (D) | 0.31 (0.1-0.49) | 0.49 (0.29-0.78) |
| COV (%) | 4.4 (2.6-6.9) | 6.5 (3.2-9.1) |
| ICC | 0.90 | 0.88 |
| SA | ||
| Sw (D) | 0.015 (0.008-0.034) | 0.026 (0.011-0.043) |
| TRT (D) | 0.041 (0.021-0.059) | 0.072 (0.039-0.094) |
| COV (%) | 5.4 (2.8-7.9) | 7.8 (4.3-9.6) |
| ICC | 0.89 | 0.86 |
RMS HOA – Root means square of higher-order aberrations, RMS LOA – Root mean square of lower-order aberrations, RMS COMA – Root mean square of coma, SA – Spherical aberration, Sw – within-subject standard deviation, TRT – test–retest variability, COV – within-subject coefficient of variation, ICC – intraclass correlation coefficient, OSI – objective scatter index
Repeatability (with 95% CI) of flat keratometry, steep keratometry, TCT, RMS HOA, RMS LOA, RMS COMA, and SA derived from Pentacam AXL wave in postrefractive eyes with a mean OSI ≥1
| Sw | TRT | COV (%) | ICC | |
|---|---|---|---|---|
| Flat keratometry | 0.32 (0.13-0.46) | 0.88 (0.48-0.99) | 0.5 (0.0-0.9) | 0.92 |
| Steep keratometry | 0.18 (0.05-0.22) | 0.49 (0.23-0.68) | 1.8 (0.5-2.2) | 0.90 |
| TCT measurement | 10.65 (7.3-12.23) | 29.5 (20.6-41.2) | 3.3 (0.3-4.9) | 0.93 |
| RMS HOA | 0.046 (0.016-0.062) | 0.12 (0.06-0.18) | 9.8 (5.2-14.1) | 0.78 |
| RMS LOA | 0.14 (0.07-0.32) | 0.38 (0.11-0.56) | 4.8 (2.4-7.6) | 0.84 |
| RMS COMA | 0.037 (0.018-0.051) | 0.102 (0.06-0.19) | 6.2 (3.6-9.4) | 0.79 |
| SA | 0.021 (0.011-0.039) | 0.058 (0.023-0.076) | 6.1 (3.1-9.6) | 0.80 |
TCT – thinnest corneal thickness, RMS HOA – Root means square of higher-order aberrations, RMS LOA – Root mean square of lower-order aberrations, RMS COMA – Root mean square of coma, SA – Spherical aberration, Sw – within-subject standard deviation, TRT – test–retest variability, COV – within-subject coefficient of variation, ICC – intraclass correlation coefficient, CI – confidence interval, OSI – objective scatter index
Comparison of repeatability of total aberrations derived from Pentacam AXL wave and iTrace in postrefractive eyes with a mean OSI ≥1
| Parameter | Pentacam AXL wave | iTrace |
|---|---|---|
| RMS COMA | ||
| Sw (D) | 0.037 (0.018-0.051) | 0.041 (0.3-0.062) |
| TRT (D) | 0.102 (0.06-0.19) | 0.113 (0.034-0.22) |
| COV (%) | 6.2 (3.6-9.4) | 7.5 (4.1-11.2) |
| ICC | 0.79 | 0.78 |
| RMS HOA | ||
| Sw (D) | 0.046 (0.016-0.062) | 0.059 (0.03-0.076) |
| TRT (D) | 0.12 (0.06-0.18) | 0.16 (0.032-0.24) |
| COV (%) | 9.8 (5.2-14.1) | 10.1 (6.3-14.6) |
| ICC | 0.78 | 0.76 |
| RMS LOA | ||
| Sw (D) | 0.14 (0.07-0.32) | 0.23 (0.05-0.36) |
| TRT (D) | 0.38 (0.11-0.56) | 0.63 (0.24-0.86) |
| COV (%) | 4.8 (2.4-7.6) | 6.9 (3.4-9.7) |
| ICC | 0.84 | 0.81 |
| SA | ||
| Sw (D) | 0.021 (0.011-0.039) | 0.035 (0.013-0.052) |
| TRT (D) | 0.058 (0.023-0.076) | 0.096 (0.04-0.13) |
| COV (%) | 6.1 (3.1-9.6) | 7.4 (4.1-10.2) |
| ICC | 0.80 | 0.79 |
RMS HOA – Root means square of higher-order aberrations, RMS LOA – Root mean square of lower-order aberrations, RMS COMA – Root mean square of coma, SA –Spherical aberration, Sw – within-subject standard deviation, TRT – test–retest variability, COV – within-subject coefficient of variation, ICC – intraclass correlation coefficient, OSI – objective scatter index