OBJECTIVE: To investigate image quality and radiation dose of CT colonography (CTC) with adaptive iterative dose reduction three-dimensional (AIDR3D).Entities:
Mesh:
Year: 2015 PMID: 25635839 PMCID: PMC4311968 DOI: 10.1371/journal.pone.0117116
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Quantitative evaluation of CTC image quality with FBP and AIDR3D.
| Evaluation item | 10 mAs with FBP | 10 mAs with AIDR3D | 20 mAs with FBP | 20 mAs with AIDR3D | 30 mAs with FBP | 30 mAs with AIDR3D | 40 mAs with FBP | 40 mAs with AIDR3D | 50 mAs with FBP | 50 mAs with AIDR3D |
|---|---|---|---|---|---|---|---|---|---|---|
| Noise | 48.0±3.1 | 21.2±2.5 | 32.5±3.2 | 16.4±1.6 | 25.2±2.6 | 14.0±1.8 | 23.1±3.1 | 13.0±2.0 | 16.8±2.6 | 11.8±1.4 |
| SNR | 0.9±0.1 | 1.7±0.3 | 1.3±0.2 | 1.9±0.2 | 1.4±0.2 | 2.1±0.4 | 1.6±0.4 | 2.7±0.3 | 1.9±0.4 | 3.0±0.4 |
| CNR | 21.4±1.3 | 48.4±5.7 | 31.8±3.4 | 62.3±6.8 | 40.9±4.6 | 73.2±9.8 | 54.7±5.9 | 79.7±11.7 | 62.0±6.2 | 85.8±10.9 |
Qualitative evaluation of CTC image quality with FBP and AIDR3D.
| Reconstruction technique | 10 mAs | 20 mAs | 30 mAs | 40 mAs | 50 mAs |
|---|---|---|---|---|---|
| FBP | 1.9 | 2.5 | 3.7 | 4.1 | 4.3 |
| AIDR3D | 2.8 | 4.1 | 4.5 | 4.7 | 4.8 |
Fig 1A-JComparison of qualitative CTC images of porcine colons using 5 current doses with 2 reconstructions.
CTC images of 10 mAs with FBP, 10 mAs with AIDR3D, 20 mAs with FBP, 20 mAs with AIDR3D, 30 mAs with FBP, 30 mAs with AIDR3D, 40 mAs with FBP, 40 mAs with AIDR3D, 50 mAs with FBP, 50 mAs with AIDR3D, respectively; their qualitative evaluation scores are 1.9, 2.8, 2.5, 4.1, 3.7, 4.5, 4.1, 4.7, 4.3, 4.8, respectively.