Literature DB >> 12101472

Multi-slice CT-colonography in low-dose technique--Preliminary results.

M Cohnen1, C Vogt, V Aurich, A Beck, D Häussinger, U Mödder.   

Abstract

PURPOSE: Assessment of Multi-Slice CT-colonography (MSCTC) with standard and low-dose technique in comparison to conventional colonoscopy (CC).
MATERIALS AND METHODS: After standardized bowel cleansing, 28 men and 24 women (mean age 60,3 years) underwent CC within two hours after MSCTC had been performed (4 x 2.5 mm collimation, 3 mm slice width, 17.5 mm table feed, 120 kVp). In 19 patients, MSCTC was performed with 120 mAs (CTDI w, eff : 11.28 mGy), and in 33 patients with 10 mAs (CTDI w, eff : 0.94 mGy), respectively. After noise reduction using mathematical algorithms by a dedicated software (ECCET), images were analyzed by two blinded readers in simultaneously displayed 2D- and 3D-modes. Findings in MSCTC were compared to the results obtained with CC.
RESULTS: Calculated effective doses ranged between 9 and 12 mSv for 120 mAs, and between 0.75 and 1 mSv for the low-dose technique. In standard dose MSCTC, 20 of 25 lesions (80 %) were found: all 7 tumors (100 %), one large polyp (> 10 mm; 100 %), 6 of 7 medium sized polyps (9.9 to 5 mm; 86 %) and 6 of 10 small polyps (< 5 mm; 60 %). In low-dose MSCTC, 32 of 44 lesions (73 %) were seen: all three large polyps (100 %), 13 of 14 medium sized polyps (93 %) and 16 of 27 small polyps (59 %). An overall sensitivity of 78 % with a specificity of 68 % was calculated. False positive findings were seen mostly for small lesions (12 for standard dose, 11 for low-dose).
CONCLUSION: Despite a 12-fold reduction of radiation exposure, MSCTC in low-dose technique performs comparably well to standard dose MSCTC. After mathematical noise reduction, more than 90 % of polyps larger than 5 mm can be detected.

Entities:  

Mesh:

Year:  2002        PMID: 12101472     DOI: 10.1055/s-2002-32691

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  4 in total

1.  Radiation-related cancer risks from CT colonography screening: a risk-benefit analysis.

Authors:  Amy Berrington de González; Kwang Pyo Kim; Amy B Knudsen; Iris Lansdorp-Vogelaar; Carolyn M Rutter; Rebecca Smith-Bindman; Judy Yee; Karen M Kuntz; Marjolein van Ballegooijen; Ann G Zauber; Christine D Berg
Journal:  AJR Am J Roentgenol       Date:  2011-04       Impact factor: 3.959

2.  Evaluation of dose exposure in 64-slice CT colonography.

Authors:  O Luz; M Buchgeister; M Klabunde; T Trabold; A F Kopp; C D Claussen; M Heuschmid
Journal:  Eur Radiol       Date:  2007-02-07       Impact factor: 5.315

3.  Computed tomographic colonography: an evidence-based analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2003-10-01

4.  Pilot study on image quality and radiation dose of CT colonography with adaptive iterative dose reduction three-dimensional.

Authors:  Hesong Shen; Dan Liang; Mingyue Luo; Chaijie Duan; Wenli Cai; Shanshan Zhu; Jianping Qiu; Wenru Li
Journal:  PLoS One       Date:  2015-01-30       Impact factor: 3.240

  4 in total