Literature DB >> 25471979

Side readout of long scintillation crystal elements with digital SiPM for TOF-DOI PET.

Jung Yeol Yeom1, Ruud Vinke2, Craig S Levin3.   

Abstract

PURPOSE: Side readout of scintillation light from crystal elements in positron emission tomography (PET) is an alternative to conventional end-readout configurations, with the benefit of being able to provide accurate depth-of-interaction (DOI) information and good energy resolution while achieving excellent timing resolution required for time-of-flight PET. This paper explores different readout geometries of scintillation crystal elements with the goal of achieving a detector that simultaneously achieves excellent timing resolution, energy resolution, spatial resolution, and photon sensitivity.
METHODS: The performance of discrete LYSO scintillation elements of different lengths read out from the end/side with digital silicon photomultipliers (dSiPMs) has been assessed.
RESULTS: Compared to 3 × 3 × 20 mm(3) LYSO crystals read out from their ends with a coincidence resolving time (CRT) of 162 ± 6 ps FWHM and saturated energy spectra, a side-readout configuration achieved an excellent CRT of 144 ± 2 ps FWHM after correcting for timing skews within the dSiPM and an energy resolution of 11.8% ± 0.2% without requiring energy saturation correction. Using a maximum likelihood estimation method on individual dSiPM pixel response that corresponds to different 511 keV photon interaction positions, the DOI resolution of this 3 × 3 × 20 mm(3) crystal side-readout configuration was computed to be 0.8 mm FWHM with negligible artifacts at the crystal ends. On the other hand, with smaller 3 × 3 × 5 mm(3) LYSO crystals that can also be tiled/stacked to provide DOI information, a timing resolution of 134 ± 6 ps was attained but produced highly saturated energy spectra.
CONCLUSIONS: The energy, timing, and DOI resolution information extracted from the side of long scintillation crystal elements coupled to dSiPM have been acquired for the first time. The authors conclude in this proof of concept study that such detector configuration has the potential to enable outstanding detector performance in terms of timing, energy, and DOI resolution.

Mesh:

Substances:

Year:  2014        PMID: 25471979      PMCID: PMC4247369          DOI: 10.1118/1.4901524

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

Review 1.  Promising new photon detection concepts for high-resolution clinical and preclinical PET.

Authors:  Craig S Levin
Journal:  J Nucl Med       Date:  2012-02       Impact factor: 10.057

2.  Optimization of a LSO-Based Detector Module for Time-of-Flight PET.

Authors:  W W Moses; M Janecek; M A Spurrier; P Szupryczynski; W-S Choong; C L Melcher; M Andreaco
Journal:  IEEE Trans Nucl Sci       Date:  2010-06-01       Impact factor: 1.679

3.  First characterization of a digital SiPM based time-of-flight PET detector with 1 mm spatial resolution.

Authors:  Stefan Seifert; Gerben van der Lei; Herman T van Dam; Dennis R Schaart
Journal:  Phys Med Biol       Date:  2013-04-15       Impact factor: 3.609

4.  A method to achieve spatial linearity and uniform resolution at the edges of monolithic scintillation crystal detectors.

Authors:  Ruud Vinke; Craig S Levin
Journal:  Phys Med Biol       Date:  2014-05-19       Impact factor: 3.609

Review 5.  Instrumentation aspects of animal PET.

Authors:  Yuan-Chuan Tai; Richard Laforest
Journal:  Annu Rev Biomed Eng       Date:  2005       Impact factor: 9.590

6.  Compact pulse width modulation circuitry for silicon photomultiplier readout.

Authors:  M F Bieniosek; P D Olcott; C S Levin
Journal:  Phys Med Biol       Date:  2013-07-08       Impact factor: 3.609

7.  Study of a high-resolution, 3D positioning cadmium zinc telluride detector for PET.

Authors:  Y Gu; J L Matteson; R T Skelton; A C Deal; E A Stephan; F Duttweiler; T M Gasaway; C S Levin
Journal:  Phys Med Biol       Date:  2011-02-18       Impact factor: 3.609

Review 8.  Recent development in PET instrumentation.

Authors:  By Hao Peng; Craig S Levin
Journal:  Curr Pharm Biotechnol       Date:  2010-09-01       Impact factor: 2.837

9.  Continuous depth-of-interaction encoding using phosphor-coated scintillators.

Authors:  Huini Du; Yongfeng Yang; Jarek Glodo; Yibao Wu; Kanai Shah; Simon R Cherry
Journal:  Phys Med Biol       Date:  2009-03-03       Impact factor: 3.609

10.  Photo-detectors for time of flight positron emission tomography (ToF-PET).

Authors:  Virginia Ch Spanoudaki; Craig S Levin
Journal:  Sensors (Basel)       Date:  2010-11-18       Impact factor: 3.576

View more
  9 in total

1.  Analytical calculation of the lower bound on timing resolution for PET scintillation detectors comprising high-aspect-ratio crystal elements.

Authors:  Joshua W Cates; Ruud Vinke; Craig S Levin
Journal:  Phys Med Biol       Date:  2015-06-17       Impact factor: 3.609

2.  A multiplexed TOF and DOI capable PET detector using a binary position sensitive network.

Authors:  M F Bieniosek; J W Cates; C S Levin
Journal:  Phys Med Biol       Date:  2016-10-14       Impact factor: 3.609

3.  Evaluation of a clinical TOF-PET detector design that achieves ⩽100 ps coincidence time resolution.

Authors:  Joshua W Cates; Craig S Levin
Journal:  Phys Med Biol       Date:  2018-06-07       Impact factor: 3.609

4.  Characterization of stacked-crystal PET detector designs for measurement of both TOF and DOI.

Authors:  Jeffrey P Schmall; Suleman Surti; Joel S Karp
Journal:  Phys Med Biol       Date:  2015-04-10       Impact factor: 3.609

5.  Study of optical reflectors for a 100ps coincidence time resolution TOF-PET detector design.

Authors:  Andrea Gonzalez-Montoro; Shirin Pourashraf; Min Sun Lee; Joshua W Cates; Craig S Levin
Journal:  Biomed Phys Eng Express       Date:  2021-09-15

6.  Parallax error in long-axial field-of-view PET scanners-a simulation study.

Authors:  Jeffrey P Schmall; Joel S Karp; Matt Werner; Suleman Surti
Journal:  Phys Med Biol       Date:  2016-07-01       Impact factor: 3.609

7.  A layered single-side readout depth of interaction time-of-flight-PET detector.

Authors:  L Bläckberg; S Sajedi; G El Fakhri; H Sabet
Journal:  Phys Med Biol       Date:  2021-02-11       Impact factor: 3.609

Review 8.  Silicon photomultiplier signal readout and multiplexing techniques for positron emission tomography: a review.

Authors:  Haewook Park; Minseok Yi; Jae Sung Lee
Journal:  Biomed Eng Lett       Date:  2022-07-16

9.  Development and initial characterization of a high-resolution PET detector module with DOI.

Authors:  Mohan Li; Yuli Wang; Shiva Abbaszadeh
Journal:  Biomed Phys Eng Express       Date:  2020-10-14
  9 in total