Halil Huseyin Cagatay1, Metin Ekinci1, Zeliha Yazar1, Gokcen Gokce2, Erdinc Ceylan3.
Abstract
PURPOSE: To compare intraocular pressure (IOP) measurements with the Goldmann applanation tonometer (GAT) and the ICare rebound tonometer (RBT) in high myopic eyes. PATIENTS AND METHODS: This randomized prospective study included 40 eyes of 40 patients with high myopia. All patients' central corneal thickness (CCT), anterior chamber depth (ACD), axial length (AXL), keratometry, and refractive measurements were recorded and followed by IOP measurement with RBT and GAT.Entities:
Mesh:
Year: 2014 PMID: 25379549 PMCID: PMC4212535 DOI: 10.1155/2014/869460
Source DB: PubMed Journal: ScientificWorldJournal ISSN: 1537-744X
IOP measurement results.
| RBT readings (mmHg) | GAT readings (mmHg) | CORRECTED GAT readings (mmHg) | GAT-ICARE (mmHg) | |
|---|---|---|---|---|
| Mean | 17.18 | 16.48 | 17.49 | −0.71 |
| Standard deviation | 3.72 | 3.19 | 3.01 | 2.35 |
| Range | 11.10–27.00 | 10–25 | 11.25–24.05 | −4.50–6.00 |
Refractive errors and keratometry values.
|
|
| Mean | Spherical value (D) | Cyclindiri | Spherical equivalent (D) | |
|---|---|---|---|---|---|---|
| Mean | 42.46 | 44.08 | 43.27 | −10.44 | −1.75 | −11.31 |
| Standard deviation | 1.48 | 1.5 | 1.4 | 4.28 | 0.96 | 4.30 |
| Range | 40.25–45.75 | 10–25 | 11.25–24.05 | −6.0–−23.0 | −0.25–3.75 | −6.38–24.50 |
Figure 1Correlation between CCT and the deviation of the RBT measurements from the corrected GAT values calculated according to the formula derived from the studies of Doughty and Zaman [15].
Figure 2Bland-Altman analysis showing the distribution of differences in IOP (GAT tonometer value minus RBT value, mmHg) (y-axis) and the mean IOP value of the tonometers (x-axis) for each eye measured.