Clinical comparison of the Icare tonometer and Goldmann applanation tonometry.

Nima Pakrou1, Tim Gray, Richard Mills, John Landers, Jamie Craig.   

Abstract

PURPOSE: To compare a new method of intraocular pressure (IOP) measurement, using the Icare tonometer, with Goldmann applanation tonometry (GAT). PATIENTS AND METHODS: Two observers obtained IOP readings in 292 eyes (143 right and 149 left) of 153 subjects, using the Icare without topical anesthetic. A GAT reading was subsequently obtained by a consultant ophthalmologist, without the knowledge of the Icare readings. Central corneal thickness (CCT) was obtained on all eyes with ultrasound pachymetry. Patient comfort after IOP measurement was assessed in a consecutive subset of patients.
RESULTS: The intraclass correlation coefficient between the 2 modalities of IOP measurement was r=0.95 for the right and r=0.93 for the left eye. The mean difference (Icare-GAT) between the IOP measured by the 2 methods was 0.4 mm Hg in the right eye (SD 3.0, 95% confidence interval -5.5 to 6.3), and 0.8 mm Hg in the left eye (SD 3.0, confidence interval -4.7 to 6.2). GAT measurements did not vary with CCT [correlation coefficient=0.09 (P=0.25) right and 0.14 (P=0.09) left eyes]. However, IOP measured with Icare tonometry increased with increasing CCT [correlation coefficient=0.16 (P=0.05) right and 0.21 (P=0.01) left eyes]. For every 100-microm increase in CCT, the difference (Icare-GAT) increased by 1 mm Hg. Of the 38 consecutive patients surveyed, 28 (73.7%) rated the Icare more comfortable than GAT, with only 2 (5.3%) rating it less comfortable (P<0.001).
CONCLUSIONS: There is good correlation between the 2 methods of IOP measurement, even at extremes of IOP. The Icare instrument was easy to use and recorded rapid and consistent readings with minimal training. It seems to be more comfortable than GAT and obviates the need for topical anesthesia.

Entities:  

Mesh:

Year:  2008        PMID: 18303384     DOI: 10.1097/IJG.0b013e318133fb32

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  38 in total

1.  Age variations in intraocular pressure in a cohort of healthy Austrian school children.

Authors:  W A Dusek; B K Pierscionek; J F McClelland
Journal:  Eye (Lond)       Date:  2012-03-23       Impact factor: 3.775

2.  Early and frequent development of ocular hypertension in children with nephrotic syndrome.

Authors:  Emi Kawaguchi; Kenji Ishikura; Riku Hamada; Yoshinobu Nagaoka; Yoshihiko Morikawa; Tomoyuki Sakai; Yuko Hamasaki; Hiroshi Hataya; Eiichiro Noda; Masaru Miura; Takashi Ando; Masataka Honda
Journal:  Pediatr Nephrol       Date:  2014-05-24       Impact factor: 3.714

3.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

4.  Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study.

Authors:  Senay Asik Nacaroglu; Emine Seker Un; Mehmet Giray Ersoz; Yelda Tasci
Journal:  Int J Ophthalmol       Date:  2015-06-18       Impact factor: 1.779

5.  Limbal rebound tonometry: clinical comparisons and applications.

Authors:  Georgios Bontzos; Michail Agiorgiotakis; Zoi Kapsala; Efstathios Detorakis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-06-28       Impact factor: 3.117

6.  [Evaluation of the self-tonometer Icare ONE in comparison to Goldmann applanation tonometry].

Authors:  V Witte; Ä Glass; R Beck; R Guthoff
Journal:  Ophthalmologe       Date:  2012-10       Impact factor: 1.059

7.  Does patient comfort influence the choice of tonometer for the measurement of intraocular pressure?

Authors:  Mary O Ugalahi; Mukaila A Seidu; Bolutife A Olusanya; Aderonke M Baiyeroju
Journal:  Int Ophthalmol       Date:  2015-10-16       Impact factor: 2.031

8.  Error in measurement of intraocular pressure with the Icare and IcarePRO.

Authors:  Mitsuya Otsuka; Naoki Tojo; Atsushi Hayashi
Journal:  Int Ophthalmol       Date:  2019-11-06       Impact factor: 2.031

9.  Measurement of Intraocular Pressure by Patients With Glaucoma.

Authors:  Savva Pronin; Lyndsay Brown; Roly Megaw; Andrew J Tatham
Journal:  JAMA Ophthalmol       Date:  2017-10-01       Impact factor: 7.389

10.  Immediate changes in intraocular pressure after clear corneal micro-incision versus small-incision cataract surgery.

Authors:  Ken Hayashi; Motoaki Yoshida; Koichi Yoshimura
Journal:  Jpn J Ophthalmol       Date:  2014-06-28       Impact factor: 2.447

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.