The influence of corneal properties on rebound tonometry.

Wan-sang Chui1, Andrew Lam, Davie Chen, Roger Chiu.   

Abstract

PURPOSE: To determine the influence of corneal thickness, corneal hysteresis (CH), and corneal resistance factor (CRF) on intraocular pressure (IOP) measurements obtained using the iCARE rebound tonometer.
DESIGN: Cross-sectional study. PARTICIPANTS: One hundred twenty-five eyes of 125 normal healthy subjects.
METHODS: Intraocular pressure was determined in all subjects using the iCARE at both central and temporal corneal regions. Corneal thickness at these 2 regions was determined by ultrasonic pachymetry. Corneal hysteresis and corneal resistance factor were measured using an ocular response analyzer. Goldmann applanation tonometry (GAT) was also performed on all eyes. MAIN OUTCOME MEASURES: Intraocular pressure, corneal thickness, corneal hysteresis, and corneal resistance factor.
RESULTS: Although the peripheral cornea was significantly thicker than the central cornea, IOPs obtained by the iCARE at the 2 regions were similar and significantly higher than GAT measurements. There were an overestimation at higher IOPs and an underestimation at lower IOPs relative to GAT measurements. Multiple regression analysis showed a significant correlation between the iCARE and CH (partial correlation r = -0.67, P<0.01) and CRF (partial correlation r = 0.82, P<0.01). The correlation between the iCARE and central corneal thickness was not significant.
CONCLUSIONS: Rebound tonometry is affected by corneal properties including CH and CRF but not corneal thickness.

Entities:  

Mesh:

Year:  2007        PMID: 17573114     DOI: 10.1016/j.ophtha.2007.03.061

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  43 in total

1.  Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects.

Authors:  Ioannis Halkiadakis; Aimilianos Stratos; George Stergiopoulos; Eleni Patsea; Sotiris Skouriotis; Panagiotis Mitropoulos; Dimitrios Papaconstantinou; Gerasimos Georgopoulos
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-08       Impact factor: 3.117

2.  Comparation of the new rebound tonometer IOPen and the Goldmann tonometer, and their relationship to corneal properties.

Authors:  J Moreno-Montañés; I Gosende; J Caire; M García-Granero
Journal:  Eye (Lond)       Date:  2010-10-01       Impact factor: 3.775

3.  Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Fatih Özcura; Nilgün Yildirim; Afsun Şahin; Ertuğrul Çolak
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

4.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

5.  Dynamic contour tonometry over silicone hydrogel contact lens.

Authors:  Andrew K C Lam; Jimmy S H Tse
Journal:  J Optom       Date:  2013-08-28

6.  [Comparison of the iCare tonometer with the Goldmann tonometer in Malawi].

Authors:  J Hohmann; M Schulze-Schwering; T Chirambo Nyaka; V Moyo; P C Kayange; D Doycheva; N H Batumba; M S Spitzer
Journal:  Ophthalmologe       Date:  2012-11       Impact factor: 1.059

7.  [Evaluation of the self-tonometer Icare ONE in comparison to Goldmann applanation tonometry].

Authors:  V Witte; Ä Glass; R Beck; R Guthoff
Journal:  Ophthalmologe       Date:  2012-10       Impact factor: 1.059

8.  Validation of the Icare® TONOVET plus rebound tonometer in normal rabbit eyes.

Authors:  Shawna Gloe; Abby Rothering; Julie A Kiland; Gillian J McLellan
Journal:  Exp Eye Res       Date:  2019-06-12       Impact factor: 3.467

9.  The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes.

Authors:  Mustafa Değer Bilgeç; Eray Atalay; Ömer Sözer; Hüseyin Gürsoy; Muzaffer Bilgin; Nilgün Yıldırım
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

10.  Measurement of Intraocular Pressure by Patients With Glaucoma.

Authors:  Savva Pronin; Lyndsay Brown; Roly Megaw; Andrew J Tatham
Journal:  JAMA Ophthalmol       Date:  2017-10-01       Impact factor: 7.389

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.