Effects of characteristic x rays on the noise power spectra and detective quantum efficiency of photoconductive x-ray detectors.

W Zhao1, W G Ji, J A Rowlands.   

Abstract

The effects of K fluorescence on the imaging performance of photoconductor-based x-ray imaging systems are investigated. A cascaded linear systems model was developed, where a parallel cascaded process was implemented to take into account the effect of K-fluorescence reabsorption on the modulation transfer function (MTF), noise power spectrum (NPS), and the spatial frequency dependent detective quantum efficiency [DQE(f)] of an imaging system. The investigation was focused on amorphous selenium (a-Se), which is the most highly developed photoconductor material for x-ray imaging. The results were compared to those obtained with Monte Carlo simulation using the same imaging condition and detector parameters, so that the validity of the cascaded linear system model could be confirmed. Our results revealed that K-fluorescence reabsorption in a-Se is responsible for a 18% drop in NPS at high spatial frequencies with an incident x-ray photon energy of E=20 keV (which is just above the K edge of 12.5 keV). When E increases to 60 keV, the effects of K-fluorescence reabsorption on NPS decrease to approximately 12% at high spatial frequencies. Because the high frequency drop is present in both MTF and NPS, the effect of K fluorescence on DQE(f) is minimal, especially for E that is much higher than the K edge. We also applied the cascaded linear system model to a newly developed compound photoconductor, lead iodide (PbI2), and found that at 60 keV there is a high frequency drop in NPS of 19%. The calculated NPS were compared to previously published measurements of PbI2 detectors.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11695766     DOI: 10.1118/1.1405845

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  17 in total

1.  Solid-state fluoroscopic imager for high-resolution angiography: parallel-cascaded linear systems analysis.

Authors:  Srinivasan Vedantham; Andrew Karellas; Sankararaman Suryanarayanan
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

2.  Cascaded systems analysis of the 3D noise transfer characteristics of flat-panel cone-beam CT.

Authors:  Daniel J Tward; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

3.  Three-dimensional linear system analysis for breast tomosynthesis.

Authors:  Bo Zhao; Wei Zhao
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

4.  Noise aliasing and the 3D NEQ of flat-panel cone-beam CT: effect of 2D/3D apertures and sampling.

Authors:  Daniel J Tward; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

5.  Accurate MTF measurement in digital radiography using noise response.

Authors:  Andrew Kuhls-Gilcrist; Amit Jain; Daniel R Bednarek; Kenneth R Hoffmann; Stephen Rudin
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

6.  A theoretical and experimental evaluation of the microangiographic fluoroscope: A high-resolution region-of-interest x-ray imager.

Authors:  Amit Jain; D R Bednarek; Ciprian Ionita; S Rudin
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

7.  Progress in electron-multiplying CCD (EMCCD) based, high-resolution, high-sensitivity x-ray detector for fluoroscopy and radiography.

Authors:  Andrew T Kuhls; Girijesh Yadava; Vikas Patel; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2007

8.  The effect of amorphous selenium detector thickness on dual-energy digital breast imaging.

Authors:  Yue-Houng Hu; Wei Zhao
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

9.  Fundamental relationship between the noise properties of grating-based differential phase contrast CT and absorption CT: theoretical framework using a cascaded system model and experimental validation.

Authors:  Ke Li; Nicholas Bevins; Joseph Zambelli; Guang-Hong Chen
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

10.  Signal and noise characteristics of a CdTe-based photon counting detector: Cascaded systems analysis and experimental studies.

Authors:  Xu Ji; Ran Zhang; Yongshuai Ge; Guang-Hong Chen; Ke Li
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.