| Literature DB >> 35861648 |
Dustin J Jacqmin1,2, Jessica R Miller1,2, Brendan A Barraclough1,3, Zacariah E Labby1.
Abstract
PURPOSE: The purpose of this work is to evaluate the Standard Imaging Exradin W2 plastic scintillation detector (W2) for use in the types of fields used for stereotactic radiosurgery.Entities:
Keywords: Exradin W2; plastic scintillation detector; radiation measurement; small field dosimetry; stereotactic radiosurgery
Mesh:
Substances:
Year: 2022 PMID: 35861648 PMCID: PMC9359019 DOI: 10.1002/acm2.13728
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
FIGURE 1Radiation fields used to measure Čerenkov light ratio. The top row shows the minimum fiber length configurations, while the bottom row shows the maximum fiber length configurations. (a) 10 × 1 cm rectangle method, (b) 5.5 × 10 cm rectangle method, (c) 3 × 5 cm rectangle method, (d) 2 × 3 cm rectangle method. Jaw positions are shown with a faint yellow rectangle; MLC positions are displayed as opaque blue rectangles
Characteristics of radiosurgery plans
| Plan Number | Indication | Target volume (cc) | Equivalent sphere diameter (cm) | Aperture definition |
|---|---|---|---|---|
| 1 | Brain metastasis | 3.11 | 1.8 | MLCs |
| 2 | Brain metastasis | 0.15 | 0.7 | 7.5‐, 10‐mm cones |
| 3 | Brain metastasis | 0.52 | 1.0 | MLCs |
| 4 | Brain metastasis | 0.06 | 0.5 | 5‐, 7.5‐mm cones |
| 5 | Brain metastasis | 3.8 | 1.9 | MLCs |
| 6 | Brain metastasis | 0.06 | 0.5 | 5‐, 7.5‐mm cones |
| 7 | Trigeminal neuralgia | N/A | N/A | 4‐mm cone |
FIGURE 2The measurement geometry used for radiosurgery plans
Čerenkov light ratio (CLR) and gain values measured using the four CLR measurement techniques for 6 MV FFF
| Measurement technique | CLR | Gain |
|---|---|---|
| 10 × 1 cm rectangle method | 0.970 | 2.397 |
| 5.5 × 10 cm rectangle method | 0.972 | 2.398 |
| 3 × 5 cm rectangle method | 0.970 | 2.375 |
| 2 × 3 cm rectangle method | 0.946 | 2.344 |
Summary of general characterization testing, with results from Carrasco et al. and Galavis et al. for comparison
| Test | Our result (W2 detector) | Carrasco et al. 2015 (W1 detector) | Galavis et al. 2019 (W2 detector) |
|---|---|---|---|
| Energy dependence | Max. deviation 2.3% | Negligible (k = 2) | N/A |
| Settling behavior |
RMSGreen = 0.20% RMSBlue = 0.8% | N/A |
|
| Short‐term repeatability |
|
| N/A |
| Dose‐response linearity |
RMS6MV = 0.11% RMS6 MV FFF = 0.13% | RMS = 0.61% | “dose linearity readings were within 0.05%” |
| Angular dependence (axial) | RMS = 0.24% | RMS = 0.21% | N/A |
| Angular dependence (polar) | RMS = 1.48% | N/A | N/A |
| Temperature dependence | −0.180%°C–1 | −0.225%°C–1 | −0.170%°C–1 |
| Dose per pulse dependence | RMS = 0.35% | RMS = 0.38% | N/A |
| Repetition rate dependence |
RMS6MV = 0.66% RMS6 MV FFF = 0.19% | RMS = 0.53% | “linear above 200 MU/min” |
Note: Our results are for 6 MV FFF unless otherwise stated. Carrasco combined measurements for 6 and 15 MV photons when reporting σ and RMS values. Galavis’ results are for 6 MV.
Abbreviation: RMS, root‐mean‐square.
The percent difference between the dose measured with the W2 scintillator and the dose measured with the ion chamber
| 6 MV FFF Cal | 6 MV Cal | 10 MV FFF Cal | 10 MV Cal | |
|---|---|---|---|---|
|
| 0.0 ± 0.6 | −1.3 ± 0.2 | −1.8 ± 0.2 | −2.3 ± 0.2 |
|
| 0.9 ± 0.3 | 0.0 ± 0.3 | −0.7 ± 0.2 | −1.0 ± 0.4 |
|
| 1.6 ± 0.3 | 0.5 ± 0.2 | 0.0 ± 0.1 | −0.5 ± 0.2 |
|
| 1.9 ± 0.2 | 0.6 ± 0.2 | 0.2 ± 0.1 | −0.1 ± 0.2 |
Note: Each row represents output measurements for a given beam energy using the four different calibrations (columns). The uncertainty is expressed with k = 2.
FIGURE 3Polar angle dependence of the W2 scintillator, normalized at perpendicular incidence
Gamma analysis results comparing different scanning techniques for relative profile measurements of a 1 × 1 cm MLC‐shaped field
| Scantechnique | Integration time (s) | Speed (cm/s) | 0.5%G/0.5 mm | 1%G/0.5 mm | 2%G/0.5mm | 1%G/1 mm |
|---|---|---|---|---|---|---|
| Scanning | 0.3 | 96.9 | 100 | 100 | 100 | |
| Scanning | 0.6 | 95.7 | 99.4 | 100 | 100 | |
| Scanning | 0.9 | 91.2 | 97.2 | 99.5 | 100 | |
| Scanning | 1.2 | 85.1 | 93.1 | 99.2 | 98.5 | |
| Scanning | 1.5 | 87.3 | 95.1 | 97.5 | 100 | |
| Step by Step | 0.2 | 97.8 | 99.5 | 100 | 100 | |
| Step by Step | 0.5 | 97 | 99.8 | 100 | 100 | |
| Step by Step | 1 | 99 | 100 | 100 | 100 | |
| Step by Step | 2 |
|
|
|
|
Note: The step‐by‐step profile acquired with 2 s integration was used as the gold standard.
Gamma analysis results comparing different scanning techniques for relative profile measurements of a 4‐mm radiosurgery cone
| Scantechnique | Integration time (s) | Speed (cm/s) | 0.5%G/0.5 mm | 1%G/0.5 mm | 2%G/0.5 mm | 1%G/1 mm |
|---|---|---|---|---|---|---|
| Scanning | 0.3 | 99.8 | 100 | 100 | 100 | |
| Scanning | 1.5 | 89 | 96.6 | 99.4 | 100 | |
| Step by step | 0.2 | 99.2 | 100 | 100 | 100 | |
| Step by step | 0.5 | 99 | 100 | 100 | 100 | |
| Step by step | 1 | 100 | 100 | 100 | 100 | |
| Step by step | 2 |
|
|
|
|
Note: The step‐by‐step profile acquired with 2 s integration was used as the gold standard.
FIGURE 4Inline profile measurements of a 1 × 1 cm2 MLC‐shaped field. Both measurements were acquired with the W2 plastic scintillator using different scan settings (black = step‐by‐step measurement with 2 s integration time, red = scanning measurement with a scanning speed of 0.3 cm/s). The gamma values from a 1D gamma analysis are plotted as well (1% global dose difference, 0.5 mm distance‐to‐agreement)
Gamma analysis results comparing four Čerenkov light ratio (CLR) measurement methods for relative profile measurements
| CLR | Scan direction | 0.5%G/0.5 mm | 1%G/0.5 mm | 2%G/0.5 mm | 1%G/1 mm |
|---|---|---|---|---|---|
| 5.5 × 10 | Crossline |
|
|
|
|
| Inline |
|
|
|
| |
| 3 × 5 | Crossline | 100 | 100 | 100 | 100 |
| Inline | 99.2 | 100 | 100 | 100 | |
| 2 × 3 | Crossline | 99.4 | 100 | 100 | 100 |
| Inline | 100 | 100 | 100 | 100 | |
| 10 × 1 | Crossline | 98.8 | 100 | 100 | 100 |
| Inline | 99.5 | 100 | 100 | 100 |
Note: Gamma analysis results for the W2 comparing four CLR measurement methods for the 6FFF 4 mm cone field. The profiles acquired with the “5.5 × 10 cm rectangle method” are used as the gold standard.
Gamma analysis results for the comparison of the W2 detector with the IBA RAZOR diode
| Aperture | Scan direction | 0.5%G/0.5 mm | 1%G/0.5 mm | 2%G/0.5 mm | 1%G/1 mm |
|---|---|---|---|---|---|
| 1 × 1 MLC | Crossline | 93.6 | 100 | 100 | 100 |
| Inline | 87.9 | 99.5 | 100 | 99.7 | |
| 4 mm Cone | Crossline | 90.3 | 99.1 | 100 | 100 |
| Inline | 64.3 | 96.1 | 100 | 100 |
FIGURE 5Crossline profile measurements of a 4‐mm cone‐shaped field. The measurements were acquired with the W2 plastic scintillator (black) and an IBA RAZOR diode (red). The gamma values from a 1D gamma analysis are plotted as well (1% global dose difference, 0.5 mm distance‐to‐agreement)
MLC and cone output factors measured with the W2 and the RAZOR diode
| MLC‐defined field size (cm x cm) | Cone diameter (mm) | W2 output factor | RAZOR output factor | % Difference |
|---|---|---|---|---|
| 1 × 1 | – | 0.755 | 0.742 | 1.8% |
| 2 × 2 | – | 0.832 | 0.820 | 1.5% |
| 3 × 3 | – | 0.872 | 0.861 | 1.3% |
| – | 4 | 0.581 | 0.572 | 1.5% |
| – | 5 | 0.637 | 0.628 | 1.4% |
| – | 7.5 | 0.728 | 0.718 | 1.3% |
| – | 10 | 0.784 | 0.774 | 1.3% |
| – | 12.5 | 0.82 | 0.809 | 1.3% |
| – | 15 | 0.846 | 0.834 | 1.5% |
| – | 17.5 | 0.864 | 0.851 | 1.6% |
Cone output factors computed with four different Čerenkov light ratio (CLR) measurement methods
| RAZOR | 5.5 × 10 | 3 × 5 | 2 × 3 | 10 × 1 | |||||
|---|---|---|---|---|---|---|---|---|---|
| Cone diameter (mm) | OF | OF | % Diff | OF | % Diff | OF | % Diff | OF | % Diff |
| 4 | 0.572 | 0.581 | 1.5% | 0.580 | 1.4% | 0.575 | 0.5% | 0.580 | 1.4% |
| 5 | 0.628 | 0.637 | 1.4% | 0.636 | 1.3% | 0.631 | 0.4% | 0.636 | 1.3% |
| 7.5 | 0.718 | 0.728 | 1.3% | 0.727 | 1.2% | 0.721 | 0.4% | 0.727 | 1.2% |
| 10 | 0.774 | 0.784 | 1.3% | 0.784 | 1.2% | 0.777 | 0.4% | 0.783 | 1.2% |
| 12.5 | 0.809 | 0.820 | 1.3% | 0.819 | 1.3% | 0.813 | 0.4% | 0.819 | 1.2% |
| 15 | 0.834 | 0.846 | 1.5% | 0.846 | 1.4% | 0.839 | 0.6% | 0.846 | 1.4% |
| 17.5 | 0.851 | 0.864 | 1.6% | 0.864 | 1.5% | 0.857 | 0.7% | 0.864 | 1.5% |
The effect of Čerenkov light ratio (CLR) on the composite treatment plans
| Difference between measurement and calculation (%) | |||||
|---|---|---|---|---|---|
| Plan Number | Equivalent squarediameter (cm) | 5.5 × 10 | 3 × 5 | 2 × 3 | 10 × 1 cm |
| 1 | 1.8 | 0.29% | 0.29% | 0.22% | 0.29% |
| 2 | 0.7 | 1.15% | 1.12% | 0.72% | 1.10% |
| 3 | 1.0 | 2.18% | 2.16% | 1.92% | 2.15% |
| 4 | 0.5 | 0.56% | 0.53% | 0.21% | 0.51% |
| 5 | 1.9 | 0.01% | 0.01% | ‐0.07% | 0.00% |
| 6 | 0.5 | 0.82% | 0.79% | 0.47% | 0.78% |
| 7 | N/A | 1.19% | 1.15% | 0.64% | 1.13% |
The effect of polar angle correction on the composite treatment plans
| Equivalent square diameter (cm) | Difference between measurement and calculation (%) | ||
|---|---|---|---|
| Plan number | No polar angle correction | With polar angle correction | |
| 1 | 1.8 | 0.29% | −0.13% |
| 2 | 0.7 | 1.15% | 0.92% |
| 3 | 1.0 | 2.18% | 1.55% |
| 4 | 0.5 | 0.56% | −0.16% |
| 5 | 1.9 | 0.01% | −0.51% |
| 6 | 0.5 | 0.82% | 0.41% |
| 7 | N/A | 1.19% | 1.24% |