| Literature DB >> 35856911 |
Maude Gondré1, Mireille Conrad1, Véronique Vallet1, Jean Bourhis2, François Bochud1, Raphaël Moeckli1.
Abstract
BACKGROUND: RaySearch (AB, Stockholm) has released a module for CyberKnife (CK) planning within its RayStation (RS) treatment planning system (TPS).Entities:
Keywords: CyberKnife; Monte Carlo; collapsed cone; commissioning; treatment planning system; validation measurements
Mesh:
Year: 2022 PMID: 35856911 PMCID: PMC9359029 DOI: 10.1002/acm2.13732
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
Parameters available for modeling in RayStation (RS)
| Parameters | Collimator | Indicator | Manual/auto‐modeling |
|---|---|---|---|
| Energy spectrum | All | PDD | Auto‐modeling |
| Primary source | All | Steepness penumbra of small field sizes | Auto‐modeling |
| Flattening filter source | All | Out‐of‐field doses of large field sizes | Auto‐modeling |
| Energy correction factor | Fixed/Iris | PDD | Manual |
| Diameter correction factor | Fixed/Iris | FWHM profile | Manual |
| Output correction factor | All | All | Auto‐modeling |
Abbreviations: CC, collapsed cone; PDD, percentage depth doses.
The energy correction factor for the Iris collimator was not available for CC modeling.
FIGURE 1Phantoms used for dose measurement: (a) homogeneous phantom with A1SL insert (Phantom A), (b) homogeneous phantom with films insert (Phantom B), (c) heterogeneous phantom with a lung slab and A1SL insert (Phantom C), (d) homogeneous phantom with A1SL inserts in depth (Phantom D), and (e) Octavius phantom with PTW SRS 100012 (Phantom E)
List of configurations tested for the validation of fixed, Iris, and multileaf collimators (MLC)
| Configuration | Phantom | Irradiation type | Fixed diameter measured (mm) | Iris diameter measured (mm) | MLC field size measured (mm2) | Measured quantity |
|---|---|---|---|---|---|---|
| 1 | A | SB, normal incidence | 20, 60 | ≥15 | 20 × 20, 60 × 60, 80 × 80 | Δ |
| 2 | B | SB, normal incidence | 5, 7.5, 10, 12.5 |
For CC: 12.5 For MC: 7.5, 10, 12.5 | – |
Δ GI FWHM |
| 3 | A | SB, oblique incidence | 20, 60 | 20, 40 | – | Δ |
| 4 | A/E | MB | 20 |
Plan 1: 12.5, 15, 20 Plan 2: 20, 30, 40 | 20 × 20 |
Δ Profiles |
| 5 | C | SB, normal incidence | 20, 60 | 20, 40 | 40 × 45, 60 × 60 | Δ |
| 6 | C | SB, oblique incidence | 60 | 20, 40 | – | Δ |
| 7 | C | MB | 20 |
Plan 1: 20, 25, 30 Plan 2: 30, 40, 50 | 20 × 20 | Δ |
| 8 | D | Depth measurements | 60 | 50, 60 | 38 × 60 | Δ |
| 9 | E | Patient QA | 7.5 | ≥15 | Irregular shapes | GI |
Abbreviations: CC, collapsed cone; GI, gamma index; MB, multiple‐beam; MC, Monte Carlo; SB, single‐beam.
“–” Indicates that no measurements were performed for that configuration.
FIGURE 2Beam eye's view of multileaf collimators (MLC) for (a) irregular field and (b) off‐axis field of beam f
Differences obtained between computed and measured validation curves for all collimators and algorithms
| Fixed | Iris | MLC | ||||
|---|---|---|---|---|---|---|
| Algorithm | CC | MC | CC | MC | CC | MC |
| Δ | 0.0 ± 0.4 | 0.3 ± 0.5 | 0.1 ± 0.3 | 0.1 ± 0.5 | 0.1 ± 0.4 | 0.1 ± 0.3 |
| ΔFWHM (%) | −0.3 ± 0.3 | 0.6 ± 0.6 | 0.0 ± 0.3 | 0.1 ± 0.3 | −0.2 ± 0.2 | 0.1 ± 0.0 |
Abbreviations: CC, collapsed cone; MC, Monte Carlo; MLC, multileaf collimators; PDD, percentage depth doses.
Dose differences obtained with measurements of the different configurations
| Configuration | Fixed | Iris | MLC | ||||
|---|---|---|---|---|---|---|---|
| CC | MC | CC | MC | CC | MC | ||
| 1 | −0.1 ± 0.0 | 0.2 ± 0.5 | 0.0 ± 0.5 | 0.4 ± 0.5 | −0.1 ± 0.6 | −0.2 ± 0.4 | |
| 2 | 0.1 ± 1.4 | −0.6 ± 1.9 | −1.0 | 0.9 ± 1.9 | – | – | |
| 3 | −0.3 ± 0.9 | −0.1 ± 0.5 | −0.4 ± 0.1 | 0.7 ± 0.2 | – | – | |
| 4 | −0.1 | −0.4 | −0.8 ± 1.3 | −0.4 ± 1.4 | 0.3 | 0.8 | |
| 5 | 0.4 ± 0.3 | −0.7 ± 0.7 | 0.4 ± 0.6 | −0.5 ± 1.1 | 0.7 ± 0.2 | 0.4 ± 0.2 | |
| 6 | −0.6 | 0.1 | −0.2 ± 0.1 | 0.0 ± 0.6 | – | – | |
| 7 | −0.7 | −1.4 | 0.7 ± 0.1 | −0.1 ± 0.4 | 0.6 | −0.1 | |
| 8 | 3 cm | 0.3 | −0.1 | 0.9 | 0.2 | 0.0 | −0.5 |
| 6 cm | 0.5 | 0.5 | −0.1 | 0.3 | −0.2 | −0.4 | |
| 9 cm | 1.0 | 1.0 | 1.7 | 1.6 | 1.0 | 0.7 | |
| 15.5 cm | 1.3 | 1.3 | 1.7 | 2.3 | 2.0 | 1.9 | |
| 19.5 cm | 1.8 | 1.7 | 2.8 | 2.9 | 3.2 | 2.5 | |
| Average Δ | 0.3 ± 0.8 | 0.1 ± 0.9 | 0.5 ± 1.1 | 0.7 ± 1.1 | 0.8 ± 1.1 | 0.6 ± 1.0 | |
| 9 | 100 | 100 | 98.2 ± 2.1 | 96.7 ± 2.1 | 99.5 ± 0.3 | 99.1 ± 0.5 | |
Abbreviations: CC, collapsed cone; MC, Monte Carlo; MLC, multileaf collimators.
No standard deviation indicated when only one measurement was performed.
“–” Indicates that no measurements were performed for that configuration.
FIGURE 3Comparison of profiles for multiple‐beam (MB) plan between the 1000 SRS detector (markers) and the RayStation (RS) treatment planning system (TPS) (filled line) calculated with collapsed cone (CC) algorithm with (a) fixed, (b) Iris, and (c) multileaf collimators (MLC). The gray area represents a dose difference of ±2%.
Results of film dosimetry (ΔFWHM and gamma index [GI]) for fixed and iris collimators and GI for irregular and off‐axis multileaf collimators (MLC) beams
| Collimator | Fixed | Iris | MLC | ||||
|---|---|---|---|---|---|---|---|
| Algorithms | CC | MC | CC | MC | CC | MC | |
| Film dosimetry (%) | ΔFWHM | −0.52 ± 1.08 | −1.18 ± 0.73 | −1.0 | 1.87 ± 0.51 | – | – |
| GI | 99.5 ± 0.6 | 99.4 ± 0.4 | 99.8 | 99.7 ± 0.2 | – | – | |
| GI Phantom E (%) | GI irregular field | – | – | – | – | 100 | 98 |
| GI off‐axis beams | – | – | – | – | 99.5 ± 0.8 | 98.8 ± 1.0 | |
Abbreviations: CC, collapsed cone; MC, Monte Carlo.
Indicates that no measurements were performed for that configuration.