Luis Muñoz1,2, Tomas Kron2,3, Marco Petasecca2, Joseph Bucci4,5, Michael Jackson2,6, Peter Metcalfe2, Anatoly B Rosenfeld2, Giordano Biasi2,3. 1. Genesiscare Flinders Private Hospital, Bedford Park, SA, Australia. 2. Centre for Medical Radiation Physics, University of Wollongong, NSW, Australia. 3. Peter MacCallum Cancer Centre, Melbourne, VIC, Australia. 4. St. George Cancer Care Centre, St George Hospital, Kogarah, NSW, Australia. 5. Genesiscare Waratah Private Hospital, Hurstville, NSW, Australia. 6. University of New South Wales, Kensington, NSW, Australia.
Abstract
PURPOSE: Stereotactic radiosurgery (SRS) can be delivered with a standard linear accelerator (linac). At institutions having more than one linac, beam matching is common practice. In the literature, there are indications that machine central axis (CAX) matching for broad fields does not guarantee matching of small fields with side ≤2 cm. There is no indication on how matching for broad fields on axis translates to matching small fields off axis. These are of interest to multitarget single-isocenter (MTSI) SRS planning and the present work addresses that gap in the literature. METHODS: We used 6 MV flattening filter free (FFF) beams from four Elekta VersaHD® linacs equipped with an Agility™ multileaf collimator (MLC). The linacs were strictly matched for broad fields on CAX. We compared output factors (OPFs) and effective field size, measured concurrently using a novel 2D solid-state dosimeter "Duo" with a spatial resolution of 0.2 mm, in square and rectangular static fields with sides from 0.5 to 2 cm, either on axis or away from it by 5 to 15 cm. RESULTS: Among the four linacs, OPF for fields ≥1 × 1 cm2 ranged 1.3% on CAX, whereas off axis a maximum range of 1.9% was observed at 15 cm. A larger variability in OPF was noted for the 0.5 × 0.5 cm2 field, with a range of 5.9% on CAX, which improved to a maximum of 2.3% moving off axis. Two linacs showed greater consistency with a range of 1.4% on CAX and 2.2% at 15 cm off axis. Between linacs, the effective field size varied by <0.04 cm in most cases, both on and off axis. Tighter matching was observed for linacs with a similar focal spot position. CONCLUSIONS: Verification of small-field consistency for matched linacs used for SRS is an important task for dosimetric validation. A significant benefit of concurrent measurement of field size and OPF allowed for a comprehensive assessment using a novel diode array. Our study showed the four linacs, strictly matched for broad fields on CAX, were still matched down to a field size of 1 x 1 cm2 on and off axis.
PURPOSE: Stereotactic radiosurgery (SRS) can be delivered with a standard linear accelerator (linac). At institutions having more than one linac, beam matching is common practice. In the literature, there are indications that machine central axis (CAX) matching for broad fields does not guarantee matching of small fields with side ≤2 cm. There is no indication on how matching for broad fields on axis translates to matching small fields off axis. These are of interest to multitarget single-isocenter (MTSI) SRS planning and the present work addresses that gap in the literature. METHODS: We used 6 MV flattening filter free (FFF) beams from four Elekta VersaHD® linacs equipped with an Agility™ multileaf collimator (MLC). The linacs were strictly matched for broad fields on CAX. We compared output factors (OPFs) and effective field size, measured concurrently using a novel 2D solid-state dosimeter "Duo" with a spatial resolution of 0.2 mm, in square and rectangular static fields with sides from 0.5 to 2 cm, either on axis or away from it by 5 to 15 cm. RESULTS: Among the four linacs, OPF for fields ≥1 × 1 cm2 ranged 1.3% on CAX, whereas off axis a maximum range of 1.9% was observed at 15 cm. A larger variability in OPF was noted for the 0.5 × 0.5 cm2 field, with a range of 5.9% on CAX, which improved to a maximum of 2.3% moving off axis. Two linacs showed greater consistency with a range of 1.4% on CAX and 2.2% at 15 cm off axis. Between linacs, the effective field size varied by <0.04 cm in most cases, both on and off axis. Tighter matching was observed for linacs with a similar focal spot position. CONCLUSIONS: Verification of small-field consistency for matched linacs used for SRS is an important task for dosimetric validation. A significant benefit of concurrent measurement of field size and OPF allowed for a comprehensive assessment using a novel diode array. Our study showed the four linacs, strictly matched for broad fields on CAX, were still matched down to a field size of 1 x 1 cm2 on and off axis.
Authors: Ryan T Hughes; Emory R McTyre; Michael LeCompte; Christina K Cramer; Michael T Munley; Adrian W Laxton; Stephen B Tatter; Jimmy Ruiz; Boris Pasche; Kounosuke Watabe; Michael D Chan Journal: Neurosurgery Date: 2019-08-01 Impact factor: 4.654
Authors: Paul D Brown; Kurt Jaeckle; Karla V Ballman; Elana Farace; Jane H Cerhan; S Keith Anderson; Xiomara W Carrero; Fred G Barker; Richard Deming; Stuart H Burri; Cynthia Ménard; Caroline Chung; Volker W Stieber; Bruce E Pollock; Evanthia Galanis; Jan C Buckner; Anthony L Asher Journal: JAMA Date: 2016-07-26 Impact factor: 56.272
Authors: Dror Limon; Frances McSherry; James Herndon; John Sampson; Peter Fecci; Justus Adamson; Zhiheng Wang; Fang-Fang Yin; Scott Floyd; John Kirkpatrick; Grace J Kim Journal: Adv Radiat Oncol Date: 2017-09-11