Measurements of dose distributions in small beams of 6 MV x-rays.

R K Rice1, J L Hansen, G K Svensson, R L Siddon.   

Abstract

Dose distributions produced by small circular beams of 6 MV x-rays have been measured using ionisation chambers of small active volume. Specific quantities measured include tissue maximum ratios (TMR), total scatter correction factors (St), collimator scatter correction factors (Sc) and off-axis ratios (OAR). Field sizes ranged from 12.5 to 30 mm diameter, and were defined by machined auxiliary collimators with the movable jaws set for a 4 cm x 4 cm field size. Due to the lack of complete lateral electronic equilibrium for these small fields, the accuracy of the measurements was also investigated. This was accomplished by studying dose response as a function of detector size. Uncertainties of 2.5% were observed for the central axis dose in the 12.5 mm field when measuring with an ionisation chamber with a diameter of 3.5 mm. The total scatter correction factor exhibits a strong field size dependence for fields below 20 mm diameter, while the collimator scatter correction factor is constant and is defined by the setting of the movable jaws. Off-axis ratio measurements show larger dose gradients at the beam edges than those achieved with conventional collimator systems. Corrected profiles measured with an ionisation chamber are compared with measurements made with photographic film and LiF thermoluminescent dosemeters.

Entities:  

Mesh:

Year:  1987        PMID: 3671496     DOI: 10.1088/0031-9155/32/9/002

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  15 in total

1.  Development of a stereotactic device for gamma knife irradiation of small animals.

Authors:  Hyun-Tai Chung; Young Seob Chung; Dong Gyu Kim; Sun Ha Paek; Keun-Tae Cho
Journal:  J Korean Neurosurg Soc       Date:  2008-01-20

2.  Total scatter factors of small beams: a multidetector and Monte Carlo study.

Authors:  Paolo Francescon; Stefania Cora; Carlo Cavedon
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

3.  Interventional telemedicine for noninvasive neuroradiosurgery: remote-site high-performance computing, mathematical optimization, and virtual scenario simulation.

Authors:  R von Hanwehr; G F Popescu; H E Taylor; K H Winkler; C E Swenberg
Journal:  J Med Syst       Date:  1995-06       Impact factor: 4.460

4.  Dosimetry challenges for implementing emerging technologies.

Authors:  Fang-Fang Yin; Mark Oldham; Jing Cai; Qiuwen Wu
Journal:  J Phys Conf Ser       Date:  2010

5.  Evaluation of beam modeling for small fields using a flattening filter-free beam.

Authors:  Daisuke Kawahara; Shuichi Ozawa; Takeo Nakashima; Masamichi Aita; Shintaro Tsuda; Yusuke Ochi; Takuro Okumura; Hirokazu Masuda; Yoshimi Ohno; Yuji Murakami; Yasushi Nagata
Journal:  Radiol Phys Technol       Date:  2016-06-21

6.  Study of efficiency in five-field and field-by-field intensity modulated radiation therapy (IMRT) plan using DOSXYZnrc Monte Carlo code.

Authors:  Sitti Yani; Indra Budiansah; Mohamad Fahdillah Rhani; Freddy Haryanto
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-27

7.  Optimisation of dose distribution for linear accelerator-based stereotactic radiosurgery.

Authors:  T S Suh; F Bova; S C Yoon; K S Shinn; Y W Bahk
Journal:  Med Biol Eng Comput       Date:  1993-07       Impact factor: 2.602

8.  3-D dosimetric methodology in multiple isocentres radiosurgery: the influence of dose-volume histograms on the choice of the reference isodose.

Authors:  D Lefkopoulos; M Schlienger; E Touboul; M C Plazas; L Merienne
Journal:  Med Biol Eng Comput       Date:  1994-07       Impact factor: 2.602

9.  Monte carlo simulation of bony heterogeneity effects on dose profile for small irradiation field in radiotherapy.

Authors:  Simone C Cardoso; Victor Gabriel L Alves; Luiz Antonio R da Rosa; Luciana T Campos; Delano V S Batista; Alessandro Facure
Journal:  PLoS One       Date:  2010-05-03       Impact factor: 3.240

10.  American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) Practice Guideline for the Performance of Stereotactic Radiosurgery (SRS).

Authors:  Steven K Seung; David A Larson; James M Galvin; Minesh P Mehta; Louis Potters; Christopher J Schultz; Santosh V Yajnik; Alan C Hartford; Seth A Rosenthal
Journal:  Am J Clin Oncol       Date:  2013-06       Impact factor: 2.339

View more