E A van IJsseldijk1, M K Harman2, J Luetzner3, E R Valstar1, B C Stoel4, R G H H Nelissen1, B L Kaptein1.
Abstract
INTRODUCTION: Wear of polyethylene inserts plays an important role in failure of total knee replacement and can be monitored in vivo by measuring the minimum joint space width in anteroposterior radiographs. The objective of this retrospective cross-sectional study was to compare the accuracy and precision of a new model-based method with the conventional method by analysing the difference between the minimum joint space width measurements and the actual thickness of retrieved polyethylene tibial inserts.Entities:
Keywords: Knee replacement; Measurement accuracy; Model-based wear measurements; Polyethylene wear; mSJW
Year: 2014 PMID: 25278502 PMCID: PMC4220171 DOI: 10.1302/2046-3758.310.2000304
Source DB: PubMed Journal: Bone Joint Res ISSN: 2046-3758 Impact factor: 5.853
Description of the 15 total knee replacements (TKR) used in this study
| K2004 | TC-Plus | Mild | 41 | 0.7 | Infection |
| K2133 | TC-Plus | Mild | 17 | 0.2 | Pain |
| K2145 | TC-Plus | Mild | 24 | 3.0 | Infection |
| K2154 | Zimmer NK | Mild | 50 | 12.1 | Infection |
| K2171 | TC-Plus | Mild | 34 | 0.0 | Painful flexion, infection |
| K2178 | TC-Plus | Mild | 19 | 2.5 | Infection |
| K2035* | TC-Plus | Moderate | 23 | 0.0 | Infection |
| K2132 | TC-Plus | Moderate | 86 | 0.2 | Infection |
| K2137 | TC-Plus | Moderate | 130 | 23.6 | Suspected osteolysis later diagnosed as metastasis |
| K2144 | TC-Plus | Moderate | 132 | 3.1 | Aseptic loosening |
| K2175 | TC-Plus | Moderate | 60 | 0.0 | Infection |
| K2046* | Encore foundation | Severe | 144 | 4.0 | Aseptic loosening |
| K2156* | Stryker 7000 | Severe | 77 | 0.2 | Infection |
| K2159* | Sulzer Protek | Severe | 108 | 1.6 | Infection |
| K2161 | TC-Plus | Severe | 108 | 0.2 | Infection |
* For these TKRs, double leg standing radiographs were used for measuring mJSW; for all other TKR a single leg standing radiograph was used † List of manufacturers: TC-Plus (Smith & Nephew, London, UK); Zimmer NK (Zimmer, Warsaw, Indiana); Encore Foundation (DJO Surgical, Vista, California); Stryker 7000 (Stryker, Kalamazoo, Michigan); Sulzer Protek (Protek Medical Product Inc., Coralville, Iowa) ‡ The period between the radiograph acquisition and revision surgery
Statistics of the differences between the mJSW measurements (conventional ∆C and model-based ∆MB) with respect to the true minimum insert thickness
| ∆ | ∆ | ∆ | ∆ | ||||
|---|---|---|---|---|---|---|---|
| Mean (mm) | 0.94 | 0.50 | p = 0.00* | 0.34 | 0.06 | p = 0.02* | |
| Standard deviation (mm) | 0.84 | 0.79 | p = 0.77† | 0.46 | 0.46 | p = 0.98† | |
| Mean measurement error (mm) | 1.02 | 0.66 | p = 0.00* | 0.44 | 0.40 | p = 0.31* | |
| N (err < 1 mm) (%) | 9 ( | 11 ( | 13 ( | 15 ( | |||
* Paired t-test for equal means † Levene’s test for homogeneity of variance
The differences in position between the femoral contact locations (pMB) and the minimum insert thickness locations (p0) (as seen in Fig. 4). Values are expressed in millimetres.
| AP difference | ML difference | AP difference | ML difference | ||
| K2004 | Mild | 5.98 | -1.84 | 0.63 | 1.06 |
| K2133 | Mild | 5.98 | -0.22 | -2.06 | 0.81 |
| K2145 | Mild | -7.47 | 3.88 | 5.05 | -0.68 |
| K2154 | Mild | 0.52 | 0.76 | -7.51 | -0.96 |
| K2171 | Mild | 2.02 | -6.38 | -1.29 | 2.26 |
| K2178 | Mild | 1.67 | -1.06 | -0.46 | 1.20 |
| K2035 | Moderate | -1.95 | -7.26 | -0.14 | -3.37 |
| K2132 | Moderate | -8.97 | -8.80 | -16.40 | -2.59 |
| K2137 | Moderate | 2.32 | -6.57 | -8.80 | -3.39 |
| K2144 | Moderate | -9.04 | -2.06 | -9.10 | -0.17 |
| K2175 | Moderate | -12.88 | -1.16 | -17.76 | -2.18 |
| K2046 | Severe | -11.02 | -3.69 | -6.06 | 3.60 |
| K2156 | Severe | 0.14 | -1.07 | -2.51 | 2.42 |
| K2159* | Severe | 0.03 | -0.53 | 1.27 | 1.99 |
| K2161 | Severe | 4.89 | -4.30 | 1.43 | 1.78 |
AP, Anterioposterior direction; ML, mediolateral direction