Literature DB >> 21461698

A pictographic atlas for classifying damage modes on polyethylene bearings.

Melinda Harman1, Luca Cristofolini, Paolo Erani, Susanna Stea, Marco Viceconti.   

Abstract

Evaluation of medical devices retrieved after in vivo service provides unique evidence related to the physiological environment in which the biomaterials performed. This study implements a training procedure for evaluating polyethylene bearings of joint prostheses obtained after pre-clinical tests or explanted after in vivo function. A total of 161 damage regions on 45 bearings were evaluated by four observers. An illustrated Damage Mode Atlas was developed as a reference guide, inclusive of both photographs and concise written descriptions of 16 specific damage modes that are typical for polyethylene bearings. Utilizing the Damage Mode Atlas to train new researchers improved the damage pattern analysis, including more accurate identification of damage modes and improved inter-rater reliability. This Damage Mode Atlas is a useful supplementary tool for conducting Stage II non-destructive analysis of explanted polyethylene bearings used for joint replacement, in accordance with international guidelines for evaluating explanted medical devices.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21461698     DOI: 10.1007/s10856-011-4303-x

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  39 in total

1.  Tibial interface wear in retrieved total knee components and correlations with modular insert motion.

Authors:  Anand R Rao; Gerard A Engh; Matthew B Collier; Smain Lounici
Journal:  J Bone Joint Surg Am       Date:  2002-10       Impact factor: 5.284

2.  Surface damage on open box posterior-stabilized polyethylene tibial inserts.

Authors:  Mauricio Silva; Christina A Kabbash; John V Tiberi; Sang Hyun Park; Donald T Reilly; Ormonde M Mahoney; Thomas P Schmalzried
Journal:  Clin Orthop Relat Res       Date:  2003-11       Impact factor: 4.176

3.  Observations on retrieved glenoid components.

Authors:  Ralph Hertel; F T Ballmer
Journal:  J Arthroplasty       Date:  2003-04       Impact factor: 4.757

4.  Contact stresses in the glenoid component in total shoulder arthroplasty.

Authors:  W Swieszkowski; P Bednarz; P J Prendergast
Journal:  Proc Inst Mech Eng H       Date:  2003       Impact factor: 1.617

5.  Digital photogrammetry for quantitative wear analysis of retrieved TKA components.

Authors:  J C Grochowsky; L W Alaways; R Siskey; E Most; S M Kurtz
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2006-11       Impact factor: 3.368

6.  Investigation of early surface delamination observed in retrieved heat-pressed tibial inserts.

Authors:  R D Bloebaum; K Nelson; L D Dorr; A A Hofmann; D J Lyman
Journal:  Clin Orthop Relat Res       Date:  1991-08       Impact factor: 4.176

7.  Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses.

Authors:  R W Hood; T M Wright; A H Burstein
Journal:  J Biomed Mater Res       Date:  1983-09

8.  Polyethylene insert damage in unicondylar knee replacement: a comparison of in vivo function and in vitro simulation.

Authors:  M Harman; S Affatato; M Spinelli; M Zavalloni; S Stea; A Toni
Journal:  Proc Inst Mech Eng H       Date:  2010       Impact factor: 1.617

9.  The Bite Mark Standard Reference Scale--ABFO No. 2.

Authors:  W G Hyzer; T C Krauss
Journal:  J Forensic Sci       Date:  1988-03       Impact factor: 1.832

10.  Rapid polyethylene failure of unicondylar tibial components sterilized with gamma irradiation in air and implanted after a long shelf life.

Authors:  Thomas F McGovern; Deborah J Ammeen; John P Collier; Barbara H Currier; Gerard A Engh
Journal:  J Bone Joint Surg Am       Date:  2002-06       Impact factor: 5.284

View more
  3 in total

1.  Validation of a model-based measurement of the minimum insert thickness of knee prostheses: a retrieval study.

Authors:  E A van IJsseldijk; M K Harman; J Luetzner; E R Valstar; B C Stoel; R G H H Nelissen; B L Kaptein
Journal:  Bone Joint Res       Date:  2014-10       Impact factor: 5.853

2.  Peripheral snap-fit locking mechanisms and smooth surface finish of tibial trays reduce backside wear in fixed-bearing total knee arthroplasty.

Authors:  Łukasz Łapaj; Adrian Mróz; Paweł Kokoszka; Jacek Markuszewski; Justyna Wendland; Celina Helak-Łapaj; Jacek Kruczyński
Journal:  Acta Orthop       Date:  2016-10-26       Impact factor: 3.717

3.  A rare case of bilateral large osteolysis following cemented and cementless total knee arthroplasties.

Authors:  Julian Dexel; Stephan Kirschner; Melinda K Harman; Jörg Lützner
Journal:  Acta Orthop       Date:  2012-12-17       Impact factor: 3.717

  3 in total