A model-based approach to measure the minimum joint space width of total knee replacements in standard radiographs.

E A van Ijsseldijk1, E R Valstar, B C Stoel, R G H H Nelissen, B L Kaptein.   

Abstract

Excessive wear in total knee arthroplasty is detected by measuring the minimum joint space width (mJSW) in anterioposterior radiographs. The accuracy of conventional measurement methods is limited and can be improved using model-based techniques. In this study, the model-based wear measurement (MBWM) is introduced. Its accuracy and reproducibility are assessed and compared to the conventional measurement. Forty anterioposterior radiographs were obtained of a knee prosthesis using a phantom setup. Both measurement methods were applied and the accuracy and precision were compared. The reproducibility was calculated with inter- and intra-observer experiments. Three observers measured the mJSW in 30 clinical radiographs with both the conventional measurement and the MBWM and repeated this after 6 weeks. The experiments were conducted with a NexGen mobile bearing and fixed bearing prostheses. In the phantom experiment, the accuracy (mean of the absolute error) was significantly higher (t-test, p<0.01) for the MBWM as for the conventional measurement (0.15 mm versus 0.43 mm, 0.14 mm versus 0.35 mm for the mobile and fixed bearing, respectively). The standard deviation of the measurements is the smallest for the MBWM measurement for both prosthesis types (0.16 mm versus 0.47 mm, Levene's test, p<0.01). In the reproducibility experiment, both the intra- and inter-observer agreements were higher for the MBWM than for the conventional method. The results show that the MBWM is superior to the conventional measurement in both accuracy and reproducibility. Although the use of a phantom experiment poses some limitations in conveying the findings to clinical practice, this improved mJSW measurement can lead to better wear detection for surgery decisions and research purposes.
Copyright © 2012 Elsevier Ltd. All rights reserved.

Mesh:

Year:  2012        PMID: 22748324     DOI: 10.1016/j.jbiomech.2012.05.012

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  2 in total

1.  Radiostereometric Analysis Permits In Vivo Measurement of Very Small Levels of Wear in TKA.

Authors:  Matthew G Teeter; Jacob Wihlidal; Richard W McCalden; Xunhua Yuan; Steven J MacDonald; Brent A Lanting; Douglas D Naudie
Journal:  Clin Orthop Relat Res       Date:  2019-01       Impact factor: 4.176

2.  Validation of a model-based measurement of the minimum insert thickness of knee prostheses: a retrieval study.

Authors:  E A van IJsseldijk; M K Harman; J Luetzner; E R Valstar; B C Stoel; R G H H Nelissen; B L Kaptein
Journal:  Bone Joint Res       Date:  2014-10       Impact factor: 5.853

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.