Polyethylene damage and knee kinematics after total knee arthroplasty.

M K Harman1, S A Banks, W A Hodge.   

Abstract

This study characterizes the relationship between in vivo knee kinematics and polyethylene damage by combining fluoroscopic analysis of tibiofemoral contact during dynamic activities and implant retrieval analysis in the same patients. Six patients (eight knees) underwent posterior cruciate ligament-retaining total knee arthroplasty. All patients participated in fluoroscopic analysis during a stair-rise and descent activity and treadmill gait an average of 18 months after arthroplasty, and articular contact was measured. Subsequently, all polyethylene tibial inserts were retrieved after an average of 26 months in vivo function: three at autopsy and five at revision. There was a statistically significant correlation between the damage location on the retrieved inserts and the articular contact location measured fluoroscopically during the activities. The femoral contact and polyethylene damage occurred predominantly on the posterior half of the tibial articular surface, and the damage pattern was largest in the compartment with the greatest range of in vivo femoral contact for each patient. This study showed that in vivo fluoroscopic analysis can predict the damage location on the polyethylene articular surface.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11716412     DOI: 10.1097/00003086-200111000-00050

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  21 in total

1.  Significant effect of the posterior tibial slope and medial/lateral ligament balance on knee flexion in total knee arthroplasty.

Authors:  Eisaku Fujimoto; Yoshiaki Sasashige; Yasuji Masuda; Takashi Hisatome; Akio Eguchi; Tetsuo Masuda; Mikiya Sawa; Yoshinori Nagata
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-05-30       Impact factor: 4.342

2.  Alterations in lower limb multimuscle activation patterns during stair climbing in female total knee arthroplasty patients.

Authors:  G Kuntze; V von Tscharner; C Hutchison; J L Ronsky
Journal:  J Neurophysiol       Date:  2015-09-09       Impact factor: 2.714

3.  Multibody dynamic simulation of knee contact mechanics.

Authors:  Yanhong Bei; Benjamin J Fregly
Journal:  Med Eng Phys       Date:  2004-11       Impact factor: 2.242

4.  Comparing in vivo kinematics of anterior cruciate-retaining and posterior cruciate-retaining total knee arthroplasty.

Authors:  Taka-aki Moro-oka; Marc Muenchinger; Jean-Pierre Canciani; Scott A Banks
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-06-21       Impact factor: 4.342

5.  A pictographic atlas for classifying damage modes on polyethylene bearings.

Authors:  Melinda Harman; Luca Cristofolini; Paolo Erani; Susanna Stea; Marco Viceconti
Journal:  J Mater Sci Mater Med       Date:  2011-04-02       Impact factor: 3.896

6.  Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners.

Authors:  Christopher B Knowlton; Priyanka Bhutani; Markus A Wimmer
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-07-12       Impact factor: 3.368

7.  Effect of tibial slope changes on femorotibial contact kinematics after cruciate-retaining total knee arthroplasty.

Authors:  Xi-Qing Pan; A-Qin Peng; Fei Wang; Feng Li; Xi-Zeng Nie; Xi Yang; Gang Ji; Xiao-Meng Wang
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-11-25       Impact factor: 4.342

8.  [A new total knee arthroplasty with physiologically shaped surfaces. Part 2: First clinical results].

Authors:  K-H Frosch; H Nägerl; D Kubein-Meesenburg; J Buchholz; J Dörner; H Dathe; O Hellerer; C Dumont; K M Stürmer
Journal:  Unfallchirurg       Date:  2009-02       Impact factor: 1.000

9.  Radiostereometric measurement of polyethylene deformation pattern in meniscal bearing TKR at 5 years follow-up.

Authors:  Alessandro Russo; Laura Bragonzoni; Claudia Trozzi; Francesco Iacono; Andrea Visani; Maurilio Marcacci
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-11-16       Impact factor: 4.342

10.  Laser surface processing of Ti6Al4V in gaseous nitrogen: corrosion performance in physiological solution.

Authors:  Raghuvir Singh; S Ghosh Chowdhury; S K Tiwari; Narendra B Dahotre
Journal:  J Mater Sci Mater Med       Date:  2007-10-04       Impact factor: 3.896

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.