OBJECTIVES: To evaluate virtual touch tissue imaging quantification (VTIQ) as a new elastography method concerning its intra- and interexaminer reliability and its ability to differentiate benign from malignant breast lesions in comparison to and in combination with ultrasound (US) B-mode breast imaging reporting and data system (BI-RADS) assessment.Entities:
Mesh:
Year: 2014 PMID: 24800257 PMCID: PMC3988855 DOI: 10.1155/2014/960262
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Absolute and relative frequencies of each BI-RADS category and corresponding mean VTIQ values in m/s.
| Total | Benign | Malignant | |
|---|---|---|---|
| BI-RADS | |||
| 3 | 15 (14.4%) | 15 (30%) | 0 (0%) |
| 4a | 32 (30.8%) | 29 (58%) | 3 (5.6%) |
| 4b | 13 (12.5%) | 4 (8%) | 9 (16.7%) |
| 4c | 18 (17.3%) | 2 (4%) | 16 (29.6%) |
| 5 | 26 (25%) | 0 (0%) | 26 (48.1%) |
| VTIQ Mean Examiner 1 | 6.16 ± 2.21 | 4.46 ± 1.87 | 7.73 ± 1.02 |
| VTIQ Mean Examiner 2 | 5.71 ± 2.37 | 3.82 ± 1.79 | 7.46 ± 1.24 |
Figure 1(a) B-mode and VTIQ images of a 55-year-old woman show an invasive lobular carcinoma of 0.9 cm (G2). The maximum velocity of 8.40 m/s was measured in the center of the lesion. (b) B-mode and VTIQ images of a 20-year-old woman show a fibroadenoma of 1.4 cm. The maximum velocity of 2.35 m/s was measured in the center of the lesion.
Figure 2Range of the mean velocities measured in benign (0–8.13 m/s) and malignant lesions (3.06–8.40 m/s).
Mean velocity values for each BI-RADS category in m/s for examiner 1. The values for examiner 2 are shown in parentheses.
| Total | Malignant | Benign | |
|---|---|---|---|
| BI-RADS | |||
| 3 | 3.13 ± 1.47 (2.90 ± 1.15) | — | 3.13 ± 1.47 (2.90 ± 1.15) |
| 4a | 4.89 ± 1.77 (4.01 ± 1.80) | 5.93 ± 2.58 (5.26 ± 2.94) | 4.78 ± 1.69 (3.88 ± 1.67) |
| 4b | 7.33 ± 1.33 (7.25 ± 1.39) | 7.71 ± 0.93 (7.75 ± 0.73) | 6.47 ± 1.83 (6.13 ± 1.98) |
| 4c | 7.61 ± 1.02 (7.37 ± 1.22) | 7.83 ± 0.79 (7.63 ± 0.60) | 5.81 ± 1.06 (5.24 ± 3.09) |
| 5 | 7.87 ± 0.77 (7.50 ± 1.24) | 7.87 ± 0.77 (7.50 ± 1.24) | — |
Comparison of sensitivity, specificity, PPV and NPV in respect to the proposed clinical cut-off value, standard BI-RADS assessment and the combination of BI-RADS and VTIQ.
| Cut-off 5.18 m/s | BIRADS | Combination VTIQ + BIRADS | |
|---|---|---|---|
| Sensitivity | 98% | 98% | 98% |
| Specificity | 68% | 30% | 82% |
| PPV | 77% | 61% | 86% |
| NPV | 97% | 100% | 98% |
Figure 3ROC analysis of VTIQ alone and the combination of VTIQ and BI-RADS.