Performance evaluation of platelet counting by novel fluorescent dye staining in the XN-series automated hematology analyzers.

Yuzo Tanaka1, Yumiko Tanaka, Kazumi Gondo, Yoshiko Maruki, Tamiaki Kondo, Satomi Asai, Hiromichi Matsushita, Hayato Miyachi.   

Abstract

BACKGROUND: Conventional automated hematology analyzers have limitations in platelet measurements such as poor accuracy and precision in the low count range and interference by nonplatelet particles. In order to improve it, the newly developed XN-Series automated hematology analyzers (Sysmex Corporation, Kobe, Japan) have been installed with a new dedicated channel for platelet analysis (PLT-F), which is based on a fluorescence flow cytometry method with uses of a novel fluorescent dye specifically staining platelets. We evaluated the basic performance of this new PLT-F channel.
METHODS: Basic performance of the PLT-F channel in within-run reproducibility and assay linearity was studied using standard methods. Correlation was studied between PLT-F and a conventional automated hematology analyzer (XE-2100) and immunoplatelet analysis using anti-CD61 monoclonal antibody (Cell-Dyn Sapphire; Abbott Laboratories). The assay interference by nonplatelet particles such as fragmented red and white blood cells was evaluated by using clinical samples, respectively, from burn injury and acute leukemia.
RESULTS: Basic performance of the PLT-F platelet counting was satisfactory in within-run reproducibility, linearity and correlation with the conventional analyzer. The correlation was satisfactory also with the immunoplatelet analysis, even for samples from a patient with burn injury, and those with white blood cell fragments displayed, platelet abnormal flag and low platelet counts (<50 × 10(9)/l).
CONCLUSION: The platelet counting performance of the PLT-F channel of the XN Series had improved accuracy and precision in the low range and in abnormal samples, avoiding the interference by nonplatelet particles.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  PLT-F channel; automated hematology analyzer; nonplatelet particle; platelet counting; thrombocytopenia

Mesh:

Substances:

Year:  2014        PMID: 24648166      PMCID: PMC6807536          DOI: 10.1002/jcla.21691

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  10 in total

1.  An interlaboratory study of a candidate reference method for platelet counting.

Authors:  P Harrison; K A Ault; S Chapman; L Charie; B Davis; K Fujimoto; B Houwen; J Kunicka; F Lacombe; S Machin; R Raynor; L van Hove; O W van Assendelft
Journal:  Am J Clin Pathol       Date:  2001-03       Impact factor: 2.493

2.  A rapid and accurate closed-tube immunoassay for platelets on an automated hematology analyzer.

Authors:  J E Gill; K A Davis; W J Cowart; F U Nepacena; Y R Kim
Journal:  Am J Clin Pathol       Date:  2000-07       Impact factor: 2.493

3.  New quantitative parameters on a recently introduced automated blood cell counter--the XE 2100.

Authors:  C Briggs; P Harrison; D Grant; J Staves; S J MacHin
Journal:  Clin Lab Haematol       Date:  2000-12

4.  The most accurate platelet count on the Sysmex XE-2100. Optical or impedance?

Authors:  C Briggs; S Kunka; S J Machin
Journal:  Clin Lab Haematol       Date:  2004-04

5.  Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion.

Authors:  H C Segal; C Briggs; S Kunka; A Casbard; P Harrison; S J Machin; M F Murphy
Journal:  Br J Haematol       Date:  2005-02       Impact factor: 6.998

6.  Evaluation of the platelet counting by Abbott CELL-DYN SAPPHIRE haematology analyser compared with flow cytometry.

Authors:  E Grimaldi; L Del Vecchio; F Scopacasa; C Lo Pardo; F Capone; S Pariante; G Scalia; M De Caterina
Journal:  Int J Lab Hematol       Date:  2007-12-20       Impact factor: 2.877

7.  Performance of automated platelet quantification using different analysers in comparison with an immunological reference method in thrombocytopenic patients.

Authors:  Ernesto Trabuio; Sara Valverde; Francesco Antico; Fabio Manoni; Gianluca Gessoni
Journal:  Blood Transfus       Date:  2009-01       Impact factor: 3.443

8.  Performance evaluation of the Sysmex haematology XN modular system.

Authors:  Carol Briggs; Ian Longair; Punamar Kumar; Deepak Singh; Samuel J Machin
Journal:  J Clin Pathol       Date:  2012-07-31       Impact factor: 3.411

9.  Spurious elevation of automated platelet counts in secondary acute monocytic leukemia associated with tumor lysis syndrome.

Authors:  S Li; K E Salhany
Journal:  Arch Pathol Lab Med       Date:  1999-11       Impact factor: 5.534

10.  Leukemia burden delays lymphocyte and platelet recovery after allo-SCT for AML.

Authors:  R M Saliba; K V Komanduri; S Giralt; J de Souza; P Patah; B Oran; D Couriel; G Rondon; R E Champlin; M de Lima
Journal:  Bone Marrow Transplant       Date:  2008-11-17       Impact factor: 5.483

  10 in total
  8 in total

1.  Sample stability for complete blood cell count using the Sysmex XN haematological analyser.

Authors:  Massimo Daves; Elmar M Zagler; Roberto Cemin; Flora Gnech; Alexandra Joos; Stefan Platzgummer; Giuseppe Lippi
Journal:  Blood Transfus       Date:  2015-04-27       Impact factor: 3.443

2.  Clinical significance of IPF% or RP% measurement in distinguishing primary immune thrombocytopenia from aplastic thrombocytopenic disorders.

Authors:  Mikiko Sakuragi; Satoru Hayashi; Miho Maruyama; Osamu Kabutomori; Tomoko Kiyokawa; Keisuke Nagamine; Hisashi Kato; Hirokazu Kashiwagi; Yuzuru Kanakura; Yoshiaki Tomiyama
Journal:  Int J Hematol       Date:  2015-01-25       Impact factor: 2.490

3.  Changes in novel haematological parameters following thermal injury: A prospective observational cohort study.

Authors:  R J Dinsdale; A Devi; P Hampson; C M Wearn; A L Bamford; J Hazeldine; J Bishop; S Ahmed; C Watson; J M Lord; N Moiemen; P Harrison
Journal:  Sci Rep       Date:  2017-06-12       Impact factor: 4.379

4.  Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system.

Authors:  Milena Velizarova; Teodora Yacheva; Mariana Genova; Dobrin Svinarov
Journal:  J Med Biochem       Date:  2021-09-03       Impact factor: 3.402

5.  Use of Middleware Data to Dissect and Optimize Hematology Autoverification.

Authors:  Rachel D Starks; Anna E Merrill; Scott R Davis; Dena R Voss; Pamela J Goldsmith; Bonnie S Brown; Jeff Kulhavy; Matthew D Krasowski
Journal:  J Pathol Inform       Date:  2021-04-07

6.  Accuracy of a New Platelet Count System (PLT-F) Depends on the Staining Property of Its Reagents.

Authors:  Atsushi Wada; Yuri Takagi; Mari Kono; Takashi Morikawa
Journal:  PLoS One       Date:  2015-10-23       Impact factor: 3.240

7.  Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers.

Authors:  Sabrina Buoro; Tommaso Mecca; Michela Seghezzi; Barbara Manenti; Lorenzo Cerutti; Paola Dominoni; Gavino Napolitano; Stefano Resmini; Alberto Crippa; Cosimo Ottomano; Giuseppe Lippi
Journal:  Rev Bras Hematol Hemoter       Date:  2016-06-23

8.  Alinity hq platelet count is not impacted by severe microcytosis.

Authors:  Gabriella Lakos; Zainab Mukhtar; Loredana Masi; Sabatino Valente; Fabrizio Papa
Journal:  J Clin Lab Anal       Date:  2022-03-11       Impact factor: 2.352

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.