Performance evaluation of the Sysmex haematology XN modular system.

Carol Briggs1, Ian Longair, Punamar Kumar, Deepak Singh, Samuel J Machin.   

Abstract

BACKGROUND: The Sysmex XN haematology instrument performs automatic reflex testing, depending on sample results. A nucleated red blood cell (NRBC) count is provided on all samples. The instrument has a smaller footprint (34%) than previous Sysmex XE analysers.
METHODS: An evaluation comparing all results to the Sysmex XE-2100 and manual microscopic differential and morphology (n=390) was performed followed by a workflow study of 1000 samples to compare speed of operation and number of blood films reviews required from both systems.
RESULTS: The new features on the instrument are: (1) white cell and NRBC channel, all samples include the NRBC count; (2) white cell precursor channel: false positive flags for blasts, abnormal lymphocytes and atypical lymphocytes are reduced significantly without a statistical increase of false negatives; (3) low white cell count mode: suggested setting of <0.5×10(9)/l. An extended count is more precise and provides an accurate differential. Fluorescent platelet count is performed in a dedicated channel. If the red cell or platelet size histograms are abnormal or if the platelet count is low, then a fluorescent platelet count is automatically performed. Good correlation with the XE-2100 and manual differential was found and the improved results compared to the reference flow cytometric analysis for platelet counts, especially below 30×10(9)/l (XE-2100, R(2)=0.500; XN, R(2)=0.875).
CONCLUSION: The XN showed reduced sample turnaround time of 10% and reduced number of blood films for examination, 49% less than the XE-2100 without loss of sensitivity with more precise and accurate results on low cell counts.

Entities:  

Mesh:

Year:  2012        PMID: 22851510     DOI: 10.1136/jclinpath-2012-200930

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  25 in total

1.  Performance evaluation of platelet counting by novel fluorescent dye staining in the XN-series automated hematology analyzers.

Authors:  Yuzo Tanaka; Yumiko Tanaka; Kazumi Gondo; Yoshiko Maruki; Tamiaki Kondo; Satomi Asai; Hiromichi Matsushita; Hayato Miyachi
Journal:  J Clin Lab Anal       Date:  2014-03-19       Impact factor: 2.352

2.  Sample stability for complete blood cell count using the Sysmex XN haematological analyser.

Authors:  Massimo Daves; Elmar M Zagler; Roberto Cemin; Flora Gnech; Alexandra Joos; Stefan Platzgummer; Giuseppe Lippi
Journal:  Blood Transfus       Date:  2015-04-27       Impact factor: 3.443

3.  Clinical significance of cell population data (CPD) on Sysmex XN-9000 in septic patients with our without liver impairment.

Authors:  Sabrina Buoro; Michela Seghezzi; Mauro Vavassori; Paola Dominoni; Sara Apassiti Esposito; Barbara Manenti; Tommaso Mecca; Gianmariano Marchesi; Enrico Castellucci; Giovanna Azzarà; Cosimo Ottomano; Giuseppe Lippi
Journal:  Ann Transl Med       Date:  2016-11

4.  Decreased platelet count in children with epilepsy treated with valproate and its relationship to the immature platelet fraction.

Authors:  Hirokazu Kurahashi; Akiyoshi Takami; Kenta Murotani; Shingo Numoto; Akihisa Okumura
Journal:  Int J Hematol       Date:  2017-09-06       Impact factor: 2.490

5.  Immature granulocytes as biomarkers of inflammation in children with predialysis chronic kidney disease.

Authors:  Nuran Cetın; Evin Kocaturk; Aslı Kavaz Tufan; Zeynep Kusku Kıraz; Ozkan Alatas
Journal:  Pediatr Nephrol       Date:  2022-04-20       Impact factor: 3.714

6.  A Comparison of Mindray BC-6800, Sysmex XN-2000, and Beckman Coulter LH750 Automated Hematology Analyzers: A Pediatric Study.

Authors:  Olga Ciepiela; Iwona Kotuła; Szymon Kierat; Sandra Sieczkowska; Anna Podsiadłowska; Anna Jenczelewska; Karolina Księżarczyk; Urszula Demkow
Journal:  J Clin Lab Anal       Date:  2016-05-17       Impact factor: 2.352

7.  Extended leukocyte differential count and C-reactive protein in septic patients with liver impairment: diagnostic approach to evaluate sepsis in intensive care unit.

Authors:  Sabrina Buoro; Tommaso Mecca; Giovanna Azzarà; Sara Apassiti Esposito; Michela Seghezzi; Mauro Vavassori; Alberto Crippa; Gianmariano Marchesi; Enrico Castellucci; Cosimo Ottomano; Giuseppe Lippi
Journal:  Ann Transl Med       Date:  2015-10

Review 8.  The state of point-of-care testing: a European perspective.

Authors:  Anders Larsson; Roman Greig-Pylypczuk; Albert Huisman
Journal:  Ups J Med Sci       Date:  2015-01-26       Impact factor: 2.384

9.  Accuracy of a New Platelet Count System (PLT-F) Depends on the Staining Property of Its Reagents.

Authors:  Atsushi Wada; Yuri Takagi; Mari Kono; Takashi Morikawa
Journal:  PLoS One       Date:  2015-10-23       Impact factor: 3.240

10.  Immature platelet fraction measured on the Sysmex XN hemocytometer predicts thrombopoietic recovery after autologous stem cell transplantation.

Authors:  Noreen van der Linden; Lieke J J Klinkenberg; Steven J R Meex; Erik A M Beckers; Norbert C J de Wit; Lenneke Prinzen
Journal:  Eur J Haematol       Date:  2014-04-25       Impact factor: 2.997

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.