| Literature DB >> 11242802 |
P Harrison1, K A Ault, S Chapman, L Charie, B Davis, K Fujimoto, B Houwen, J Kunicka, F Lacombe, S Machin, R Raynor, L van Hove, O W van Assendelft.
Abstract
A multinational interlaboratory task force explored the important variables of platelet reference counting and developed a candidate flow cytometric reference method based on the RBC/platelet ratio. A multicenter comparison was performed to determine whether the method met the necessary criteria and was precise enough to be recommended as a new reference method. Each laboratory analyzed serial dilutions of normal specimens, stabilized material, and at least 60 patient specimens with a range of platelet counts from 1 to 400 x 10(3)/microL (1-400 x 10(9)/L). Pooled analysis of the serial dilutions showed that RBC-platelet and RBC-RBC coincidence events became negligible at sufficiently high dilutions (i.e., > 1:1,000). All laboratories demonstrated excellent intra-assay and acceptable interlaboratory precision. Two antibodies (CD61 and CD41) were used for identifying platelets and individually gave acceptable results, but in a minority of samples, staining differences were observed. The optimum method thus uses a double-labeling procedure with a final dilution factor of 1:1,000. The study demonstrated that this method meets the criteria for a reference platelet count.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11242802 DOI: 10.1309/91PR-E4G6-XBAF-N8DY