Evaluation of the platelet counting by Abbott CELL-DYN SAPPHIRE haematology analyser compared with flow cytometry.

E Grimaldi1, L Del Vecchio, F Scopacasa, C Lo Pardo, F Capone, S Pariante, G Scalia, M De Caterina.   

Abstract

The Abbot Cell-Dyn Sapphire is a new generation haematology analyser. The system uses optical/fluorescence flow cytometry in combination with electronic impedance to produce a full blood count. Optical and impedance are the default methods for platelet counting while automated CD61-immunoplatelet analysis can be run as selectable test. The aim of this study was to determine the platelet count performance of the three counting methods available on the instrument and to compare the results with those provided by Becton Dickinson FACSCalibur flow cytometer used as reference method. A lipid interference experiment was also performed. Linearity, carryover and precision were good, and satisfactory agreement with reference method was found for the impedance, optical and CD61-immunoplatelet analysis, although this latter provided the closest results in comparison with flow cytometry. In the lipid interference experiment, a moderate inaccuracy of optical and immunoplatelet counts was observed starting from a very high lipid value.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18177435     DOI: 10.1111/j.1751-553X.2007.01015.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  Performance of automated platelet quantification using different analysers in comparison with an immunological reference method in thrombocytopenic patients.

Authors:  Ernesto Trabuio; Sara Valverde; Francesco Antico; Fabio Manoni; Gianluca Gessoni
Journal:  Blood Transfus       Date:  2009-01       Impact factor: 3.443

2.  Performance evaluation of platelet counting by novel fluorescent dye staining in the XN-series automated hematology analyzers.

Authors:  Yuzo Tanaka; Yumiko Tanaka; Kazumi Gondo; Yoshiko Maruki; Tamiaki Kondo; Satomi Asai; Hiromichi Matsushita; Hayato Miyachi
Journal:  J Clin Lab Anal       Date:  2014-03-19       Impact factor: 2.352

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.