Gabriella Lakos1, Zainab Mukhtar1, Loredana Masi2, Sabatino Valente2, Fabrizio Papa2.
Abstract
BACKGROUND: Impedance technology has been shown to overestimate platelet (PLT) count in samples with microcytes, while the optical-fluorescence PLT count (PLT-F) by Sysmex has been suggested to be unaffected by microcytosis. The Abbott Alinity hq analyzer employs multi-dimensional optical PLT counting. Our goal was to assess the accuracy of this technology in microcytic samples.Entities:
Keywords: MAPSS™; impedance; microcytosis; optical technology; platelet count
Mesh:
Year: 2022 PMID: 35274768 PMCID: PMC8993633 DOI: 10.1002/jcla.24218
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
FIGURE 1Passing Bablok regression between Alinity hq PLT and Sysmex PLT‐F (A), Sysmex PLT‐I and PLT‐F (B), and Alinity hq PLT and Sysmex PLT‐I (C). r, correlation coefficient
Mean PLT concentration bias between Alinity hq and Sysmex PLT‐F and between PLT‐I and PLT‐F according to MCV
| MCV |
| Mean bias (×109/L; 95% CI) | |
|---|---|---|---|
| PLT‐I vs PLT‐F | Alinity hq vs PLT‐F | ||
| All (40.9–123 fL) | 447 | 9.6 (8.1–11.2) | 7.2 (6.3–8.1) |
| <65 fL | 36 | 35.2 (25.0–45.3) | 9.2 (4.9–13.5) |
| 65–80 fL | 79 | 14.4 (10.0–18.7) | 7.1 (4.3–9.8) |
| >80 fL | 332 | 6.0 (4.8–7.3) | 7.0 (6.1–8.0) |
Abbreviations: CI, Confidence interval; fL, femtoliter; MCV, Mean corpuscular volume; PLT‐F, Fluorescence platelet count; PLT‐I, Impedance platelet count.
FIGURE 2Spearman's correlation between PLT count bias (Sysmex PLT‐I vs PLT‐F) and the MCV of the samples. r s, Spearman's rank correlation coefficient
Number of suppressed (by Sysmex), invalid (by Alinity hq), flagged for review (by Sysmex), or flagged as Suspect (by Alinity hq) MPV results according to MCV
| MCV |
| Suppressed or invalid result | Flagged result | ||
|---|---|---|---|---|---|
| Alinity hq MPV | Sysmex MPV | Alinity hq MPV | Sysmex MPV | ||
| <65 fL | 36 | 0 | 13 | 9 | 5 |
| 65–80 fL | 79 | 0 | 4 | 2 | 8 |
| >80 fL | 332 | 0 | 6 | 6 | 15 |
Abbreviations: fL, femtoliter; MCV, Mean corpuscular volume; MPV, Mean platelet volume.
RDW (median and interquartile ranges) by Alinity hq and Sysmex XN according to MCV
| MCV |
| RDW median and interquartile ranges (%) |
| |
|---|---|---|---|---|
| Alinity hq | Sysmex XN | |||
| <65 fL | 36 | 13.7 (131 – 17.2) | 18.1 (17.3 – 20.3) | <0.0001 |
| 65–80 fL | 79 | 15.5 (14.2 – 18.0) | 17.0 (15.2 – 19.1) | 0.014 |
| >80 fL | 332 | 13.4 (12.8 – 14.6) | 13.6 (12.9 – 14.9) | N.S |
Abbreviations: fL, femtoliter; MCV, Mean corpuscular volume; MPV, Mean platelet volume; N.S., Not significant; RDW, Red cell distribution width.