Establishment of diagnostic reference levels in computed tomography for select procedures in Pudhuchery, India.

A Saravanakumar1, K Vaideki1, K N Govindarajan1, S Jayakumar1.   

Abstract

Computed tomography (CT) scanner under operating conditions has become a major source of human exposure to diagnostic X-rays. In this context, weighed CT dose index (CTDIw), volumetric CT dose index (CTDIv), and dose length product (DLP) are important parameter to assess procedures in CT imaging as surrogate dose quantities for patient dose optimization. The current work aims to estimate the existing dose level of CT scanner for head, chest, and abdomen procedures in Pudhuchery in south India and establish dose reference level (DRL) for the region. The study was carried out for six CT scanners in six different radiology departments using 100 mm long pencil ionization chamber and polymethylmethacrylate (PMMA) phantom. From each CT scanner, data pertaining to patient and machine details were collected for 50 head, 50 chest, and 50 abdomen procedures performed over a period of 1 year. The experimental work was carried out using the machine operating parameters used during the procedures. Initially, dose received in the phantom at the center and periphery was measured by five point method. Using these values CTDIw, CTDIv, and DLP were calculated. The DRL is established based on the third quartile value of CTDIv and DLP which is 32 mGy and 925 mGy.cm for head, 12 mGy and 456 mGy.cm for chest, and 16 mGy and 482 mGy.cm for abdomen procedures. These values are well below European Commission Dose Reference Level (EC DRL) and comparable with the third quartile value reported for Tamil Nadu region in India. The present study is the first of its kind to determine the DRL for scanners operating in the Pudhuchery region. Similar studies in other regions of India are necessary in order to establish a National Dose Reference Level.

Entities:  

Keywords:  CTDIv; CTDIw; Computed tomography; dose length product; dose reference level; pencil ionization chamber; polymethylmethacrylate phantom

Year:  2014        PMID: 24600173      PMCID: PMC3931229          DOI: 10.4103/0971-6203.125509

Source DB:  PubMed          Journal:  J Med Phys        ISSN: 0971-6203


Introduction

Computed tomography (CT) was introduced in the early 1970s and soon became a very important tool in medical diagnosis. This modality has become a very strong and flexible examination that has replaced many radiologic techniques.[12] CT applications have improved with the introduction of helical and multidetector row arrangement.[3] However, CT is associated with relatively high radiation doses, causing concerns regarding the risk of carcinogenesis.[45] In addition, the current CT scanners have an extensive choice of exposure factors and employ techniques that can significantly influence the radiation dose given to the patient.[6] Following international basic safety standards for protection against ionizing radiation and safety of radiation sources becomes mandatory whenever a radiological examination has to be performed in the case of a valid clinical indication.[78] During CT examination for a specific clinical objective, a quality image should be recorded without unnecessary dose to the patients. All guidelines therefore, include reference doses that are described as diagnostic reference levels (DRL)[9] or guidance levels[10] which can be thought of as reasonable doses for various clinical procedures that meet the clinical objectives and also can be used for optimizing patient dose. DRL are usually defined for a collection of patient dose data at the 75th percentile point of the dose spread.[11] It means 75% of the dose data are below the DRL value. DRLs are intended to provide guidance on what is achievable with current good practice rather than optimum performance, and helps to identify unusually high radiation doses or exposure levels (as in the rest of the 25% of cases). Hence, regular patient dose monitoring and image quality assessment will lead to optimal doses and meaningful DRLs and reduction of unnecessary patient doses.[12] The dose parameters recommended in the guidelines are weighed CT dose index (CTDIw) and volumetric CT dose index (CTDIv) for a single section and dose length product (DLP) for the entire examination.[13] The purpose of our study is to establish nationwide “Diagnostic Reference Level" for CT scanners for select procedures viz., head, chest, and abdomen. The first step towards setting a national DRL is to arrive at regional DRL for which the entire nation has been divided into five zones namely south, north, east, west, and center. The south zone includes the state of Tamil Nadu, Kerala, Karnataka, Andhra Pradesh, and Pudhuchery which is a union territory. An attempt has been made in the year 2008 to determine the third quartile value for CTDIv for the CT scanners in Tamil Nadu and the values were reported as 557 and 551 mGy for thorax and abdomen procedures, respectively.[14] Since then no studies have been carried out in this line in India. The aim of the present study is to establish DRL for CTDIv and DLP for the CT scanners operating in Pudhuchery region.

Materials and Methods

CT scanners

This study was carried out using six CT scanners out of 10 machines operating in Pudhuchery region. Before initiating measurements in hospitals, a questionnaire was prepared to collect data regarding the CTs’ protocols and clinical practices adopted by the hospital in Pudhuchery region. This data helped to record the CT dose index values for different scanning protocols adopted by the various departments. Table 1 summarizes the make and model of CT scanners used in this study.
Table 1

Details of CT units

Details of CT units

Dose Measurement system

In-site CT dose measurements were carried out using a calibrated 100 mm pencil ionization chamber (DCT10 RS, S/N 1636) [Figure 1a] with Solidose electrometer 400 (S/N 4253) [Figure 1b], from RTI Electronics, Sweden. The electrometer was calibrated to read the dose in phantom directly.
Figure 1

Dose measurement system (a) Pencil ion chamber (b) Electrometer

Dose measurement system (a) Pencil ion chamber (b) Electrometer

Phantom

The average patient was simulated by dedicated cylindrical polymethylmethacrylate (PMMA) phantom. The diameter and length of head-equivalent phantom was 16 and 15 cm, respectively. This was nested into another phantom with dimension 32 cm outer diameter, 16 cm inner diameter, and 15 cm length to use as a body equivalent phantom [Figure 2]. Each cylindrical phantom contains four holes (13 mm diameter) on the periphery at 90° intervals and one at the center. During measurements the unused holes were plugged using PMMA pegs [Figure 2].
Figure 2

Computed tomography dose index head and body phantom with the pencil ion chamber inserted in the center hole of head phantom and the other holes plugged with polymethylmethacrylate pegs

Computed tomography dose index head and body phantom with the pencil ion chamber inserted in the center hole of head phantom and the other holes plugged with polymethylmethacrylate pegs

Experimental technique

The most common procedures performed in most of the radiology departments’ viz., adult head, chest, and abdomen were selected for the study. The head scan procedure was carried out with and without contrast in machines other than Hitachi (Pratico), whereas chest and abdomen scan procedures were carried without contrast in all the machines. From each machine the data pertaining to 50 head, 50 chest, and 50 abdomen procedures (a total of 150 × 6 = 900 procedures) performed over a period of 1 year have been collected. The data included information related to patients such as patient height, weight, sex, age, and lateral diameter and machine operating parameters such as tube voltage and tube current-time product, pitch, scan time, slice thickness, and average scan length used for the purpose of obtaining good quality images. This data abstraction has been done as per ‘Nationwide Evaluation of X-ray Trends’ (NEXT) protocol.[15] The data related to the machine operating parameters for the scanners under study is presented in Table 2.
Table 2

Exposure parameters used for select procedures in CT examinations

Exposure parameters used for select procedures in CT examinations The dosimetry technique was based on the methods proposed by European Guidelines.[9] Before dose measurements were carried out, quality assurance (QA) was performed for each machine and was compared with the CT dose indices displayed on the control console to ensure that the output of the machines were fairly constant for the past 1 year because the data were collected during this period. After QA, each scanner's CTDI values were normalized using standard protocol involving tube potential of 80, 100, and 120 kV, tube current-time product of 100 mAs and 5 mm slice thickness. Then, the phantom was placed in the couch and as per Food and Drug Administration (FDA)'s recommendation the ion chamber was inserted into holes in the phantom such that its center matched with the isocenter of the phantom so as to read the dose received throughout the length of the phantom. The temperature and pressure correction for the chamber were applied at different locations of CT scanner under the prevailing conditions. Thus, the dose received by the phantom at the center (CTDI100, c) and periphery (CTDI100, p) was measured using the pencil ion chamber connected to a solidose electrometer for the machine operating parameters presented in Table 2. Using these dose values, the other CT dose indices viz, CTDIw, CTDIv, and DLP were calculated using equation 1, 2 and 3. CTDIw= 1/3 (CTDI100, c) + 2/3 (CTDI100, p)      (1) CTDIv= CTDIw/pitch      (2) DLP = CTDIv × scan length      (3) The CTDIv thus calculated was compared with the estimated CTDIv obtained from the control console and was ensured that the difference between the two values fell within the limit recommended by Atomic Energy Regulatory Board (AERB). To establish DRL the third quartile value of the calculated CTDIv and DLP has been determined for the dose pertaining to the exposure parameters involved during patient scanning that were routinely used in each center.

Results and Discussion

The dose received by the head and body CTDI phantom at the center and periphery was measured using the pencil ionization chamber connected to solidose electrometer. Weighted CTDI, CTDIv, and DLP have been calculated as per equations 1, 2, and 3. The values are presented in Table 3.
Table 3

The computed tomography dose indices of few scanners operating in Pudhuchery for select procedures

The computed tomography dose indices of few scanners operating in Pudhuchery for select procedures From Table 3, it can be seen that the difference between the estimated and calculated CTDIv is well within 40% which is the maximum tolerance level as per AERB standards.[16] As the mode of scanning for head region was axial in machine D (Hitachi (Pratico)) a least difference between the estimated and calculated CTDIv was observed. The mean, range, and third quartile values have been calculated for the CTDIv and DLP and are presented in Tables 4 and 5.
Table 4

Mean, range, and third quartile values for volumetric CTDI for select procedures

Table 5

Mean, range, and third quartile values for dose length product for select procedures

Mean, range, and third quartile values for volumetric CTDI for select procedures Mean, range, and third quartile values for dose length product for select procedures From Table 4 it can be observed that the range of CTDIv values is almost the same for head, chest, and abdomen scan; whereas, Table 5 reveals the fact that the range of DLP values vary significantly for head, chest, and abdomen scan. To analyze this similarity and variation, the CTDIv and DLP values and their respective third quartile values are represented using a bar chart [Figures 3 and 4].
Figure 3

Comparison between calculated volumetric computed tomography dose index values for the different scanners, third quartile value, and the European Commission Reference level for head, chest, and abdomen scan

Figure 4

Comparison between calculated dose length product values for the different scanners, third quartile value, and the European Commission reference level for head, chest, and abdomen scan

Comparison between calculated volumetric computed tomography dose index values for the different scanners, third quartile value, and the European Commission Reference level for head, chest, and abdomen scan Comparison between calculated dose length product values for the different scanners, third quartile value, and the European Commission reference level for head, chest, and abdomen scan Figures 3 and 4 reveals the fact that the third quartile values of CTDIv and DLP for head, chest, and abdomen scans are well below the European Commission Reference Level. This is because, based on the average European adult patient size (density of the scan region) the machine operating parameters used by the scanners operating in those countries are on the higher side when compared to Indian conditions. Also, the third quartile value of CTDIv for abdomen scan is less than the one reported for Tamil Nadu (521 mGy).[14] Figures 3 and 4 reveal the fact that the CTDIv and DLP value of machines A (Siemens (Somatom sprit)), B (Siemens (Somatom sensation)), and C (Siemens (Somatom sprit)) are low when compared to the other three for chest and abdomen scan. This may be due to a low scan time used in these machines when compared to machine D (Hitachi (Pratico)) and E (Philips (Brilliance Multidetector CT (MDCT))) and low tube voltage and current time product combination when compared to F (Siemens (Somatom sprit)). Due to these facts the radiation output from machines A, B, and C is less when compared to D, E, and F that has resulted in low dose level received by the phantoms and hence low CTDIv when compared to other machines. It can be observed from Figures 3 and 4 that for head scan, the experimentally determined CTDIv and DLP for machine B (Siemens (Somatom sensation)) and E (Philips (Brilliance MDCT)) are above the third quartile values. This is ascribed to the machine operating parameters viz., low pitch value and slice thickness and fairly high scan length for machine B and low pitch value and longer scan time for machine E. Due to these factors the phantom has received higher doses which in turn resulted in a higher CTDIv and DLP. If these machine operating parameters are optimized than the dose indices could be brought down below the third quartile values and that would lead to a good scan practice. As far as chest and abdomen scans are concerned, the dose indices values of machine F (Siemens (Somatom sprit)) are higher than the third quartile values. This is attributed to higher tube voltage, tube current time product, scan time, and mean scan length. These parameters when optimized would result in dose indices values below the third quartile values.

Conclusion

The paper presents the data that are an outcome of the preliminary survey and experiments carried out on more than 50% of CT scanners operating in Pudhuchery region of south India in order to establish regional DRL for select procedures. Based on the dose measurements using five point method, the CT dose indices have been calculated. To establish regional DRL, the third quartile value of CTDIv and DLP has been determined. A comparison between the CTDIv and DLP of individual scanners, third quartile values and EC DRL indicate that the third quartile values are below European Commission (EC) DRL. However, the CTDIv and DLP of certain scanners are higher than third quartile value revealing the fact that the radiation output is high due to the machine operating parameters used in these scanners. Optimization of machine operating parameters in these cases is required to prevent the patients from receiving unnecessary doses. This investigation provides data related to dose and technique to facilitate further initiatives in the optimization of patient safety in select procedures. Performance of such surveys is important in other regions of the country to formulate national reference levels.
  8 in total

1.  Radiation doses in computed tomography. The increasing doses of radiation need to be controlled.

Authors:  M M Rehani; M Berry
Journal:  BMJ       Date:  2000-03-04

2.  A contribution to the establishment of diagnostic reference levels in CT.

Authors:  K Hatziioannou; E Papanastassiou; M Delichas; P Bousbouras
Journal:  Br J Radiol       Date:  2003-08       Impact factor: 3.039

Review 3.  Overview of patient dosimetry in diagnostic radiology in the USA for the past 50 years.

Authors:  Walter Huda; Edward L Nickoloff; John M Boone
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

Review 4.  Strategies for CT radiation dose optimization.

Authors:  Mannudeep K Kalra; Michael M Maher; Thomas L Toth; Leena M Hamberg; Michael A Blake; Jo-Anne Shepard; Sanjay Saini
Journal:  Radiology       Date:  2004-01-22       Impact factor: 11.105

Review 5.  Cancer risks from diagnostic radiology.

Authors:  E J Hall; D J Brenner
Journal:  Br J Radiol       Date:  2008-05       Impact factor: 3.039

6.  Dose reduction in computed tomography by individualized scan protocols.

Authors:  L Jangland; E Sanner; J Persliden
Journal:  Acta Radiol       Date:  2004-05       Impact factor: 1.990

Review 7.  Multislice CT: technical principles and future trends.

Authors:  Mathias Prokop
Journal:  Eur Radiol       Date:  2003-12       Impact factor: 5.315

8.  Radiation safety concerns and diagnostic reference levels for computed tomography scanners in Tamil Nadu.

Authors:  Roshan S Livingstone; Paul M Dinakaran
Journal:  J Med Phys       Date:  2011-01
  8 in total
  5 in total

1.  Assessment of Regional Diagnostic Reference Levels in Dental Radiography in Tamil Nadu.

Authors:  Amal Jose; A Saravana Kumar; K N Govindarajan; Sunil Dutt Sharma
Journal:  J Med Phys       Date:  2022-03-31

2.  Estimation of dose reference levels in computed tomography for select procedures in Kerala, India.

Authors:  A Saravanakumar; K Vaideki; K N Govindarajan; S Jayakumar; B Devanand
Journal:  J Med Phys       Date:  2015 Apr-Jun

3.  Assessment of Regional Pediatric Computed Tomography Dose Indices in Tamil Nadu.

Authors:  A Saravanakumar; K Vaideki; K N Govindarajan; S Jayakumar; B Devanand
Journal:  J Med Phys       Date:  2017 Jan-Mar

4.  Cost-effective pediatric head and body phantoms for computed tomography dosimetry and its evaluation using pencil ion chamber and CT dose profiler.

Authors:  A Saravanakumar; K Vaideki; K N Govindarajan; S Jayakumar; B Devanand
Journal:  J Med Phys       Date:  2015 Jul-Sep

5.  Radiation dose metrics in multidetector computed tomography examinations: A multicentre retrospective study from seven tertiary care hospitals in Kerala, South India.

Authors:  Binoj Varghese; Indu Kandanga; Paul Puthussery; Dhanesh Vijayan; S P Harish Babu; M K Aneesh; Muhammad Noufal; E V Binu; Arun C Babu; Sheen M James; Siva Kumar
Journal:  Indian J Radiol Imaging       Date:  2018 Apr-Jun
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.