Dose reduction in computed tomography by individualized scan protocols.

L Jangland1, E Sanner, J Persliden.   

Abstract

PURPOSE: To find a method of adjusting the mAs-value in relation to the size of the patient undergoing computed tomography (CT) examination as a means of minimizing the radiation dose to the patient.
MATERIAL AND METHODS: A correction factor to be applied on the tube charge for each patient was calculated using two mathematical methods. This approach was tested on 4 Perspex phantoms of different sizes and geometries. Noise was measured in the images with and without use of the correction factors. Retrospectively, correction factors were calculated for 12 CT examinations of the abdomen and the dose reduction was estimated for these patient studies.
RESULTS: The variations in noise measured in the images of the different phantoms were dramatically reduced by both methods. The retrospectively performed patient study showed that the largest correction factor was 7 times greater than the smallest, which means that a dose reduction factor of 7 is possible in the extreme case.
CONCLUSION: Our proposed methods of adjusting the applied tube charge (mAs-value) in relation to the size of the patient can be used on the vast majority of CT systems. The potential for dose reduction is great, especially for small patients.

Entities:  

Mesh:

Year:  2004        PMID: 15239426     DOI: 10.1080/02841850410004814

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  4 in total

Review 1.  Multidetector CT of the kidney.

Authors:  E M Coppenrath; U G Mueller-Lisse
Journal:  Eur Radiol       Date:  2006-03-28       Impact factor: 7.034

2.  Establishment of diagnostic reference levels in computed tomography for select procedures in Pudhuchery, India.

Authors:  A Saravanakumar; K Vaideki; K N Govindarajan; S Jayakumar
Journal:  J Med Phys       Date:  2014-01

3.  Assessment of Regional Pediatric Computed Tomography Dose Indices in Tamil Nadu.

Authors:  A Saravanakumar; K Vaideki; K N Govindarajan; S Jayakumar; B Devanand
Journal:  J Med Phys       Date:  2017 Jan-Mar

4.  Establishment of diagnostic reference levels in cardiac computed tomography.

Authors:  Mohammad Rawashdeh; Charbel Saade; Maha Zaitoun; Mostafa Abdelrahman; Patrick Brennan; Haytham Alewaidat; Mark F McEntee
Journal:  J Appl Clin Med Phys       Date:  2019-08-30       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.