| Literature DB >> 26500404 |
A Saravanakumar1, K Vaideki1, K N Govindarajan2, S Jayakumar3, B Devanand4.
Abstract
In the present work, a pediatric head and body phantom was fabricated using polymethyl methacrylate (PMMA) at a low cost when compared to commercially available phantoms for the purpose of computed tomography (CT) dosimetry. The dimensions of head and body phantoms were 10 cm diameter, 15 cm length and 16 cm diameter, 15 cm length, respectively. The dose from a 128-slice CT machine received by the head and body phantom at the center and periphery were measured using a 100 mm pencil ion chamber and 150 mm CT dose profiler (CTDP). Using these values, the weighted computed tomography dose index (CTDIw) and in turn the volumetric CTDI (CTDIv) were calculated for various combinations of tube voltage and current-time product. A similar study was carried out using standard calibrated phantom and the results have been compared with the fabricated ones to ascertain that the performance of the latter is equivalent to that of the former. Finally, CTDIv measured using fabricated and standard phantoms were compared with respective values displayed on the console. The difference between the values was well within the limits specified by Atomic Energy Regulatory Board (AERB), India. These results indicate that the cost-effective pediatric phantom can be employed for CT dosimetry.Entities:
Keywords: CT dose profiler; CTDI phantom; Pediatric CT; pediatric radiation dose; pencil ion chamber
Year: 2015 PMID: 26500404 PMCID: PMC4594387 DOI: 10.4103/0971-6203.165073
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Routine scan parameters for pediatric patients
Scan parameters for computed tomography (CT) dosimetry
Figure 1CT dose profiler. CT = Computed tomography
Figure 2Piranha 557 reader
Figure 3Pediatric CTDI: (a) Head phantom and (b) body phantom. CTDI = Computed tomography dose index
Figure 4Dose profiles in the standard phantom: (a) Centre (b) periphery
Figure 5Dose profiles in the fabricated phantom: (a) Centre (b) periphery
Mean CTDIc, CTDIp, CTDIw, and CTDIv along with standard deviation using CTDP for fabricated phantoms
Mean CTDIc, CTDIp, CTDIw, and CTDIv along with standard deviation using pencil ion chamber for fabricated phantoms
Percentage difference between displayed and calculated CTDIv values for CTDP
Percentage difference between displayed and calculated CTDIv values for pencil ion chamber
Figure 6Variations in CTDIv with tube potential (a) head (b) body. CTDIv = Volumetric computed tomography dose index, CTDP = CT dose profiler