Roshan S Livingstone1, Paul M Dinakaran.
Abstract
Radiation safety in computed tomography (CT) scanners is of concern due its widespread use in the field of radiological imaging. This study intends to evaluate radiation doses imparted to patients undergoing thorax, abdomen and pelvic CT examinations and formulate regional diagnostic reference levels (DRL) in Tamil Nadu, South India. In-site CT dose measurement was performed in 127 CT scanners in Tamil Nadu for a period of 2 years as a part of the Atomic Energy Regulatory Board (AERB)-funded project. Out of the 127 CT scanners,13 were conventional; 53 single-slice helical scanners (SSHS); 44 multislice CT (MSCT) scanners; and 17 refurbished scanners. CT dose index (CTDI) was measured using a 32-cm polymethyl methacrylate (PMMA)-body phantom in each CT scanner. Dose length product (DLP) for different anatomical regions was generated using CTDI values. The regional DRLs for thorax, abdomen and pelvis examinations were 557, 521 and 294 mGy cm, respectively. The mean effective dose was estimated using the DLP values and was found to be 8.04, 6.69 and 4.79 mSv for thorax, abdomen and pelvic CT examinations, respectively. The establishment of DRLs in this study is the first step towards optimization of CT doses in the Indian context.Entities:
Keywords: Computed tomography; diagnostic reference level; effective dose
Year: 2011 PMID: 21430858 PMCID: PMC3048954 DOI: 10.4103/0971-6203.75471
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Scan lengths for patients undergoing thorax, abdomen and pelvis CT examinations
Effective doses, CTDIw and exposure parameters used for thorax CT examinations performed using various CT scanners
Effective doses, CTDIw and exposure parameters used for abdomen CT examinations performed using various CT scanners
Effective doses, CTDIw and exposure parameters used for pelvis CT examinations performed using various CT scanners
Comparison of DLP and effective doses in the current study with those reported in literature
Figure 1Mean effective doses for thorax CT examinations from various CT installations
Figure 2Mean effective doses for abdomen CT examinations from various CT installations
Figure 3Mean effective doses for pelvis CT examinations from various CT installations
Details on availability of radiation safety accessories in various CT centers