| Literature DB >> 28405108 |
A Saravanakumar1, K Vaideki2, K N Govindarajan3, S Jayakumar4, B Devanand1.
Abstract
The aim of this article is to assess Tamil Nadu pediatric computed tomography (CT) diagnostic reference levels (DRLs) by collecting radiation dose data for the most commonly performed CT examinations. This work was performed for thirty CT scanners installed in various parts of the Tamil Nadu region. The patient cohort was divided into two age groups: <1 year, and 1-5 years. CT dose indices were measured using a 10 cm3 pencil ion chamber with pediatric head and body polymethyl methacrylate phantoms. Dose data such as volumetric CT dose index (CTDIv) and dose length product (DLP) on a minimum of twenty average-sized pediatric patients in each category were recorded to calculate a mean site CTDIv and DLP value. The rounded 75th percentile was used to calculate a pediatric DRL for each hospital, and then region by compiling all results. Data were collected for 3600 pediatric patients. Pediatric CT DRL for two age groups: <1 year (CTDIv and DLP of head [20 mGy, 352 mGy.cm], chest [7 mGy, 120 mGy.cm] and abdomen [12 mGy, 252 mGy.cm]), and 1-5 years (CTDIv and DLP of head [38 mGy, 505 mGy.cm], chest [8 mGy, 132 mGy.cm] and abdomen [14 mGy, 270 mGy.cm]) for select procedures have been calculated. Proposed pediatric DRLs of CTDIv and DLP for head procedure were lower, and for chest and abdomen procedures were higher than European pediatric DRLs for both age groups.Entities:
Keywords: Diagnostic reference level; pediatric computed tomography; pediatric computed tomography diagnostic reference level; volumetric computed tomography dose index
Year: 2017 PMID: 28405108 PMCID: PMC5370338 DOI: 10.4103/0971-6203.202425
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Make and model of the computed tomography scanners involved in this study
Lowest, highest and mean of the operating parameters
Pediatric patient physical parameters in Tamil Nadu region
Figure 1Calculated volumetric computed tomography dose index for <1 year and 1–5 years age group of pediatric patients for select procedures.
Figure 2Calculated dose length product for <1 year and 1–5 years age group of pediatric patients for select procedures.
Mean, range and third quartile values for volumetric computed tomography dose index for select procedures
Mean, range and third quartile values for dose length product for select procedures