The Effects on Absorbed Dose Distribution in Intraoral X-ray Imaging When Using Tube Voltages of 60 and 70 kV for Bitewing Imaging.

Kristina Hellén-Halme1, Mats Nilsson2.   

Abstract

OBJECTIVES: Efforts are made in radiographic examinations to obtain the best image quality with the lowest possible absorbed dose to the patient. In dental radiography, the absorbed dose to patients is very low, but exposures are relatively frequent. It has been suggested that frequent low-dose exposures can pose a risk for development of future cancer. It has previously been reported that there was no significant difference in the diagnostic accuracy of approximal carious lesions in radiographs obtained using tube voltages of 60 and 70 kV. The aim of this study was, therefore, to evaluate the patient dose resulting from exposures at these tube voltages to obtain intraoral bitewing radiographs.
MATERIAL AND METHODS: The absorbed dose distributions resulting from two bitewing exposures were measured at tube voltages of 60 and 70 kV using Gafchromic(®) film and an anatomical head phantom. The dose was measured in the occlusal plane, and ± 50 mm cranially and caudally to evaluate the amount of scattered radiation. The same entrance dose to the phantom was used. The absorbed dose was expressed as the ratio of the maximal doses, the mean doses and the integral doses at tube voltages of 70 and 60 kV.
RESULTS: The patient receives approximately 40 - 50% higher (mean and integral) absorbed dose when a tube voltage of 70 kV is used.
CONCLUSIONS: The results of this study clearly indicate that 60 kV should be used for dental intraoral radiographic examinations for approximal caries detection.

Entities:  

Keywords:  bitewing radiography; dental digital radiography; dental radiography; radiation dosage; radiographic image enhancement.

Year:  2013        PMID: 24422035      PMCID: PMC3887573          DOI: 10.5037/jomr.2013.4302

Source DB:  PubMed          Journal:  J Oral Maxillofac Res        ISSN: 2029-283X


  23 in total

1.  A comprehensive in vitro study of image accuracy and quality for periodontal diagnosis. Part 2: the influence of intra-oral image receptor on periodontal measurements.

Authors:  Bart Vandenberghe; Hilde Bosmans; Jie Yang; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2010-05-12       Impact factor: 3.573

Review 2.  A review of dentists' use of digital radiography and caries diagnosis with digital systems.

Authors:  A Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2006-09       Impact factor: 2.419

3.  Effect of different background lighting conditions on diagnostic performance of digital and film images.

Authors:  R A Cederberg; N L Frederiksen; B W Benson; J D Shulman
Journal:  Dentomaxillofac Radiol       Date:  1998-09       Impact factor: 2.419

Review 4.  Digital radiography and caries diagnosis.

Authors:  A Wenzel
Journal:  Dentomaxillofac Radiol       Date:  1998-01       Impact factor: 2.419

5.  Accuracy of radiographic caries diagnosis at different kilovoltages and two film speeds.

Authors:  B Svenson; H G Gröndahl; A Petersson; A Olving
Journal:  Swed Dent J       Date:  1985

6.  Gafchromic film as a fast visual indicator of radiation exposure of first responders.

Authors:  Unal Oren; Christopher L Rääf; Sören Mattsson
Journal:  Radiat Prot Dosimetry       Date:  2011-09-17       Impact factor: 0.972

7.  Effect of two X-ray tube voltages on detection of approximal caries in digital radiographs. An in vitro study.

Authors:  Kristina Hellén-Halme
Journal:  Clin Oral Investig       Date:  2010-02-09       Impact factor: 3.573

8.  Effect of beam energy and filtration on the signal-to-noise ratio of the Dexis intraoral X-ray detector.

Authors:  H Kitagawa; A G Farman
Journal:  Dentomaxillofac Radiol       Date:  2004-01       Impact factor: 2.419

9.  Effect of monitors on approximal caries detection in digital radiographs--standard versus precalibrated DICOM part 14 displays: an in vitro study.

Authors:  Kristina Hellén-Halme; Mats Nilsson; Arne Petersson
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2009-02-06

10.  The influence of tube potential on periodontal bone level measurements and subjective image quality using a digital photostimulable storage phosphor sensor.

Authors:  Bart Vandenberghe; Reinhilde Jacobs
Journal:  J Oral Maxillofac Res       Date:  2010-04-01
View more
  2 in total

1.  Radiation dose from X-ray examinations of impacted canines: cone beam CT vs two-dimensional imaging.

Authors:  Nils Kadesjö; Randi Lynds; Mats Nilsson; Xie-Qi Shi
Journal:  Dentomaxillofac Radiol       Date:  2018-01-16       Impact factor: 2.419

2.  The use of a thyroid shield for intraoral anterior oblique occlusal views-a risk-based approach.

Authors:  Mark Worrall; Alison Menhinick; Donald J Thomson
Journal:  Dentomaxillofac Radiol       Date:  2017-10-27       Impact factor: 2.419

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.