Effect of beam energy and filtration on the signal-to-noise ratio of the Dexis intraoral X-ray detector.

H Kitagawa1, A G Farman.   

Abstract

OBJECTIVE: To evaluate the signal-to-noise ratio (SNR) of the Dexis charge-coupled device (CCD) intraoral X-ray detector, with special reference to the influence of beam energy and filtration.
MATERIAL AND METHODS: Digital radiographic images were made of a nine-step dental aluminium step wedge using a Dexis intraoral detector. The X-ray generator was a GE 1000 operating at 10 mA with 2.7 mm aluminium filtration. The peak tube voltage settings applied were 50 kVp, 60 kVp, 70 kVp, 80 kVp and 90 kVp. Various exposure times were used at each beam energy and the exposures (in micro C kg(-1)) were determined in each case using a 3 cl beryllium-windowed ionization chamber. SNR was defined using the surrogate measure of mean pixel value (surrogate for signal) divided by the standard deviation of the pixel value (surrogate for noise). SNRs were measured at separate regions of interest (ROIs) at the centre and at both sides of the detector for each aluminium step as well as without the step.
RESULTS: For beam energies of 50 kVp and 60 kVp, the estimated SNR improved both with increased exposure and with increased filtration (thickness of the aluminium step), and image saturation did not occur within the exposure time range permitted by the Dexis software. At 70 kVp and above, the SNR was optimized in the middle of the exposure range: 4 micro C kg(-1) with 290:1 at the ninth step (13.5 mm Al) for 70 kVp; 3.5 micro C kg(-1) with 240:1 at the ninth step (13.5 mm Al) for 80 kVp; 3.5 micro C kg(-1) with 160:1 at the ninth step (13.5 mm Al) for 90 kVp. Saturation did occur at these beam energies within the exposure time range permitted by the Dexis software.
CONCLUSION: For the Dexis intraoral radiographic imaging system, estimated SNR improved both with higher filtration and with lower kVp. The Dexis detector was capable of generating acceptable images of the step wedge at a wide range of kVp settings.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15140818     DOI: 10.1259/dmfr/26493631

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  6 in total

1.  Effect of tube potential and image receptor on the detection of natural proximal caries in primary teeth.

Authors:  Elif Sogur; B Güniz Baksı; Kaan Orhan; S Candan Paksoy; Salih Dogan; Yılmaz S Erdal; Ali Mert
Journal:  Clin Oral Investig       Date:  2010-09-14       Impact factor: 3.573

2.  Alternative X-ray filters for an intra-oral digital radiographic system.

Authors:  J Stecke; A D Cruz; S M Almeida; F N Bóscolo
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

3.  The modulation transfer function and signal-to-noise ratio of different digital filters: a technical approach.

Authors:  D D Brüllmann; B d'Hoedt
Journal:  Dentomaxillofac Radiol       Date:  2011-05       Impact factor: 2.419

4.  Effect of two X-ray tube voltages on detection of approximal caries in digital radiographs. An in vitro study.

Authors:  Kristina Hellén-Halme
Journal:  Clin Oral Investig       Date:  2010-02-09       Impact factor: 3.573

5.  The Effects on Absorbed Dose Distribution in Intraoral X-ray Imaging When Using Tube Voltages of 60 and 70 kV for Bitewing Imaging.

Authors:  Kristina Hellén-Halme; Mats Nilsson
Journal:  J Oral Maxillofac Res       Date:  2013-10-01

6.  The influence of tube potential on periodontal bone level measurements and subjective image quality using a digital photostimulable storage phosphor sensor.

Authors:  Bart Vandenberghe; Reinhilde Jacobs
Journal:  J Oral Maxillofac Res       Date:  2010-04-01