Bart Vandenberghe1, Reinhilde Jacobs1.
Abstract
OBJECTIVES: The purpose of the present study was to determine the measurement accuracy and subjective image quality for periodontal disease diagnosis when using two X-ray tube voltages with a digital photostimulable storage phosphor sensor.Entities:
Keywords: alveolar process; dental digital radiography; furcation defects.; periodontal diseases; radiographic image enhancitalicent; radiographic phantom
Year: 2010 PMID: 24421961 PMCID: PMC3886041 DOI: 10.5037/jomr.2010.1105
Source DB: PubMed Journal: J Oral Maxillofac Res ISSN: 2029-283X
Figure 1(A) Radiographic protocol of the dry adult skull. The soft tissues were simulated with synthetic material (in white), bone loss fiducials were introduced made from gutta percha fragments and rigid occlusal imprints ensured radiographic reproducibility. (B) Maxillary cadaver jaw with formalin fixed tissues.
Figure 2Standardized PSP radiographs of the dry skull's mandibular molar region. The images were obtained using two different tube voltages, 63 kV and 70 kV, at decreasing exposure times.
Descriptive statistics of the periodontal bone level measurementsa for the variables kV (irrespective exposure time) and exposure time (irrespective kV)
| Descriptives | Variables kV | Exposure time | |||
|---|---|---|---|---|---|
| kV 63 | kV 70 | 80 ms | 120 ms | 160 ms | |
| Sample size | 96 | 96 | 64 | 64 | 64 |
| Minimum | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 |
| Maximum | 1.97 | 2.00 | 1.80 | 2.00 | 1.97 |
| Mean | 0.44 | 0.44 | 0.47 | 0.45 | 0.41 |
| 95% CI | 0.37 - 0.51 | 0.37 - 0.52 | 0.39 - 0.54 | 0.36 - 0.53 | 0.33 - 0.50 |
| SD | 0.33 | 0.35 | 0.32 | 0.34 | 0.35 |
aPeriodontal bone level in mm.
CI = confidence interval of the mean; SD = standard deviation.
Figure 3Bar chart for the absolute differences (abs diff in mm, y-axis) of the radiographic measuritalicents from the Standard. The means and error bars at a 95% confidence interval (CI) are shown for the different exposure times (ms), clustered by the two kV settings (x-axis).
Figure 4Percentage of measuritalicents (y-axis) falling within a specific distance from the Standard (X-axis), ignoring exposure time. At least 90% of measuritalicents were 1 mm deviation. A kV of 63 resulted in a slightly higher accuracy at this threshold level.
Results of the Friedman ANOVA test for the ordinal scale ratings, established for the periapical radiographs parameters visibility evaluation in the six different groupsa
| Parameters | Ordinal scale ratings (mean rank) in six groups | N | Chi-Square | Df | Pb | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | 6 | |||||
| 3.08 | 3.58 | 3.58 | 3.58 | 3.58 | 3.58 | 6 | 5.00 | 5 | 0.42c | |
| Trabecular pattern depiction | 2.33 | 3.25 | 3.25 | 3.75 | 3.75 | 4.67 | 6 | 8.63 | 5 | 0.13c |
| Contrast perception | 3.00 | 3.00 | 3.00 | 3.50 | 4.50 | 4.00 | 6 | 7.50 | 5 | 0.19c |
| Bone crater involvement | 3.33 | 3.33 | 3.33 | 3.33 | 3.83 | 3.83 | 6 | 1.11 | 5 | 0.95c |
| Furcation defect involvement | 2.83 | 2.33 | 3.83 | 3.33 | 4.33 | 4.33 | 6 | 8.00 | 5 | 0.16c |
aSix groups are combination of two kV settings and three exposure times: 1 = 70 kV, 80 ms; 2 = 63, kV 80 ms; 3 = 70 kV, 120 ms; 4 = 63 kV, 120 ms; 5 = 70 kV, 160 ms; 6 = 63 kV, 160 ms.
bTested between the groups.
cThe mean differences are not significant at 95% significance level.
Df = degree of freedom.
Figure 5Bar chart for the subjective quality ratings of trabecular pattern depiction. Ordinal scores of the observers (y-axis) are plotted by exposure times (ms) and clustered by kV (x-axis).