Optical artefact characterization and correction in volumetric scintillation dosimetry.

Daniel Robertson1, Cheukkai Hui, Louis Archambault, Radhe Mohan, Sam Beddar.   

Abstract

The goals of this study were (1) to characterize the optical artefacts affecting measurement accuracy in a volumetric liquid scintillator detector, and (2) to develop methods to correct for these artefacts. The optical artefacts addressed were photon scattering, refraction, camera perspective, vignetting, lens distortion, the lens point spread function, stray radiation, and noise in the camera. These artefacts were evaluated by theoretical and experimental means, and specific correction strategies were developed for each artefact. The effectiveness of the correction methods was evaluated by comparing raw and corrected images of the scintillation light from proton pencil beams against validated Monte Carlo calculations. Blurring due to the lens and refraction at the scintillator tank-air interface were found to have the largest effect on the measured light distribution, and lens aberrations and vignetting were important primarily at the image edges. Photon scatter in the scintillator was not found to be a significant source of artefacts. The correction methods effectively mitigated the artefacts, increasing the average gamma analysis pass rate from 66% to 98% for gamma criteria of 2% dose difference and 2 mm distance to agreement. We conclude that optical artefacts cause clinically meaningful errors in the measured light distribution, and we have demonstrated effective strategies for correcting these optical artefacts.

Entities:  

Mesh:

Year:  2013        PMID: 24321820      PMCID: PMC3923433          DOI: 10.1088/0031-9155/59/1/23

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  12 in total

1.  Verification of proton range, position, and intensity in IMPT with a 3D liquid scintillator detector system.

Authors:  L Archambault; F Poenisch; N Sahoo; D Robertson; A Lee; M T Gillin; R Mohan; S Beddar
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

2.  Vignette and exposure calibration and compensation.

Authors:  Dan B Goldman
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  2010-12       Impact factor: 6.226

3.  Radiometric framework for image mosaicking.

Authors:  Anatoly Litvinov; Yoav Y Schechner
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2005-05       Impact factor: 2.129

4.  The three-dimensional scintillation dosimetry method: test for a 106Ru eye plaque applicator.

Authors:  A S Kirov; J Z Piao; N K Mathur; T R Miller; S Devic; S Trichter; M Zaider; C G Soares; T LoSasso
Journal:  Phys Med Biol       Date:  2005-06-22       Impact factor: 3.609

5.  Development of an easy-to-handle range measurement tool using a plastic scintillator for proton beam therapy.

Authors:  Yasuhiro Fukushima; Minoru Hamada; Teiji Nishio; Koichi Maruyama
Journal:  Phys Med Biol       Date:  2006-10-26       Impact factor: 3.609

6.  Acceleration of iterative image restoration algorithms.

Authors:  D S Biggs; M Andrews
Journal:  Appl Opt       Date:  1997-03-10       Impact factor: 1.980

7.  Liquid scintillator for 2D dosimetry for high-energy photon beams.

Authors:  Falk Pönisch; Louis Archambault; Tina Marie Briere; Narayan Sahoo; Radhe Mohan; Sam Beddar; Michael T Gillin
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

8.  Exploration of the potential of liquid scintillators for real-time 3D dosimetry of intensity modulated proton beams.

Authors:  Sam Beddar; Louis Archambault; Narayan Sahoo; Falk Poenisch; George T Chen; Michael T Gillin; Radhe Mohan
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

9.  Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas M.D. Anderson Cancer Center, Proton Therapy Center, Houston.

Authors:  Michael T Gillin; Narayan Sahoo; Martin Bues; George Ciangaru; Gabriel Sawakuchi; Falk Poenisch; Bijan Arjomandy; Craig Martin; Uwe Titt; Kazumichi Suzuki; Alfred R Smith; X Ronald Zhu
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

10.  Quenching correction for volumetric scintillation dosimetry of proton beams.

Authors:  Daniel Robertson; Dragan Mirkovic; Narayan Sahoo; Sam Beddar
Journal:  Phys Med Biol       Date:  2012-12-21       Impact factor: 3.609

View more
  10 in total

1.  Fast range measurement of spot scanning proton beams using a volumetric liquid scintillator detector.

Authors:  CheukKai Hui; Daniel Robertson; Fahed Alsanea; Sam Beddar
Journal:  Biomed Phys Eng Express       Date:  2015-07-30

2.  Ionization quenching correction for a 3D scintillator detector exposed to scanning proton beams.

Authors:  Fahed Alsanea; Chinmay Darne; Daniel Robertson; Sam Beddar
Journal:  Phys Med Biol       Date:  2020-04-06       Impact factor: 3.609

3.  Optical imaging provides rapid verification of static small beams, radiosurgery, and VMAT plans with millimeter resolution.

Authors:  Muhammad Ramish Ashraf; Petr Bruza; Brian W Pogue; Nathan Nelson; Benjamin B Williams; Lesley A Jarvis; David J Gladstone
Journal:  Med Phys       Date:  2019-10-04       Impact factor: 4.071

4.  Performance characterization of a 3D liquid scintillation detector for discrete spot scanning proton beam systems.

Authors:  Chinmay D Darne; Fahed Alsanea; Daniel G Robertson; Narayan Sahoo; Sam Beddar
Journal:  Phys Med Biol       Date:  2017-06-23       Impact factor: 3.609

5.  A novel proton-integrating radiography system design using a monolithic scintillator detector: experimental studies.

Authors:  Chinmay D Darne; Daniel G Robertson; Fahed Alsanea; Charles-Antoine Collins-Fekete; Sam Beddar
Journal:  Nucl Instrum Methods Phys Res A       Date:  2021-12-16       Impact factor: 1.455

6.  Direct in-water radiation dose measurements using Cherenkov emission corrected signals from polarization imaging for a clinical radiotherapy application.

Authors:  Émily Cloutier; Luc Beaulieu; Louis Archambault
Journal:  Sci Rep       Date:  2022-06-10       Impact factor: 4.996

7.  3D reconstruction of scintillation light emission from proton pencil beams using limited viewing angles-a simulation study.

Authors:  CheukKai Hui; Daniel Robertson; Sam Beddar
Journal:  Phys Med Biol       Date:  2014-07-23       Impact factor: 3.609

8.  A proton imaging system using a volumetric liquid scintillator: a preliminary study.

Authors:  Chinmay D Darne; Fahed Alsanea; Daniel G Robertson; Fada Guan; Tinsu Pan; David Grosshans; Sam Beddar
Journal:  Biomed Phys Eng Express       Date:  2019-07-12

9.  Determination of the Range and Spread-Out Bragg Peak Width of Proton Beams Using a Large-Volume Liquid Scintillator.

Authors:  Thomas Henry; Daniel Robertson; François Therriault-Proulx; Sam Beddar
Journal:  Int J Part Ther       Date:  2017-09-26

10.  Quality assurance in proton beam therapy using a plastic scintillator and a commercially available digital camera.

Authors:  Mansour Almurayshid; Yusuf Helo; Andrzej Kacperek; Jennifer Griffiths; Jem Hebden; Adam Gibson
Journal:  J Appl Clin Med Phys       Date:  2017-07-29       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.