Fiducial marker-based correction for involuntary motion in weight-bearing C-arm CT scanning of knees. Part I. Numerical model-based optimization.

Jang-Hwan Choi1, Rebecca Fahrig, Andreas Keil, Thor F Besier, Saikat Pal, Emily J McWalter, Gary S Beaupré, Andreas Maier.   

Abstract

PURPOSE: Human subjects in standing positions are apt to show much more involuntary motion than in supine positions. The authors aimed to simulate a complicated realistic lower body movement using the four-dimensional (4D) digital extended cardiac-torso (XCAT) phantom. The authors also investigated fiducial marker-based motion compensation methods in two-dimensional (2D) and three-dimensional (3D) space. The level of involuntary movement-induced artifacts and image quality improvement were investigated after applying each method.
METHODS: An optical tracking system with eight cameras and seven retroreflective markers enabled us to track involuntary motion of the lower body of nine healthy subjects holding a squat position at 60° of flexion. The XCAT-based knee model was developed using the 4D XCAT phantom and the optical tracking data acquired at 120 Hz. The authors divided the lower body in the XCAT into six parts and applied unique affine transforms to each so that the motion (6 degrees of freedom) could be synchronized with the optical markers' location at each time frame. The control points of the XCAT were tessellated into triangles and 248 projection images were created based on intersections of each ray and monochromatic absorption. The tracking data sets with the largest motion (Subject 2) and the smallest motion (Subject 5) among the nine data sets were used to animate the XCAT knee model. The authors defined eight skin control points well distributed around the knees as pseudo-fiducial markers which functioned as a reference in motion correction. Motion compensation was done in the following ways: (1) simple projection shifting in 2D, (2) deformable projection warping in 2D, and (3) rigid body warping in 3D. Graphics hardware accelerated filtered backprojection was implemented and combined with the three correction methods in order to speed up the simulation process. Correction fidelity was evaluated as a function of number of markers used (4-12) and marker distribution in three scenarios.
RESULTS: Average optical-based translational motion for the nine subjects was 2.14 mm (± 0.69 mm) and 2.29 mm (± 0.63 mm) for the right and left knee, respectively. In the representative central slices of Subject 2, the authors observed 20.30%, 18.30%, and 22.02% improvements in the structural similarity (SSIM) index with 2D shifting, 2D warping, and 3D warping, respectively. The performance of 2D warping improved as the number of markers increased up to 12 while 2D shifting and 3D warping were insensitive to the number of markers used. The minimum required number of markers for 2D shifting, 2D warping, and 3D warping was 4-6, 12, and 8, respectively. An even distribution of markers over the entire field of view provided robust performance for all three correction methods.
CONCLUSIONS: The authors were able to simulate subject-specific realistic knee movement in weight-bearing positions. This study indicates that involuntary motion can seriously degrade the image quality. The proposed three methods were evaluated with the numerical knee model; 3D warping was shown to outperform the 2D methods. The methods are shown to significantly reduce motion artifacts if an appropriate marker setup is chosen.

Entities:  

Mesh:

Year:  2013        PMID: 24007156      PMCID: PMC3751969          DOI: 10.1118/1.4817476

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  33 in total

1.  Efficient correction for CT image artifacts caused by objects extending outside the scan field of view.

Authors:  B Ohnesorge; T Flohr; K Schwarz; J P Heiken; K T Bae
Journal:  Med Phys       Date:  2000-01       Impact factor: 4.071

2.  Enhanced 3-D-reconstruction algorithm for C-arm systems suitable for interventional procedures.

Authors:  K Wiesent; K Barth; N Navab; P Durlak; T Brunner; O Schuetz; W Seissler
Journal:  IEEE Trans Med Imaging       Date:  2000-05       Impact factor: 10.048

3.  Effects of hip joint centre mislocation on gait analysis results.

Authors:  R Stagni; A Leardini; A Cappozzo; M Grazia Benedetti; A Cappello
Journal:  J Biomech       Date:  2000-11       Impact factor: 2.712

4.  3D tomographic reconstruction of coronary arteries using a precomputed 4D motion field.

Authors:  Christophe Blondel; Régis Vaillant; Grégoire Malandain; Nicholas Ayache
Journal:  Phys Med Biol       Date:  2004-06-07       Impact factor: 3.609

5.  Image quality assessment: from error visibility to structural similarity.

Authors:  Zhou Wang; Alan Conrad Bovik; Hamid Rahim Sheikh; Eero P Simoncelli
Journal:  IEEE Trans Image Process       Date:  2004-04       Impact factor: 10.856

6.  Computed tomography simulation with superquadrics.

Authors:  Jiehua Zhu; Shiying Zhao; Yangbo Ye; Ge Wang
Journal:  Med Phys       Date:  2005-10       Impact factor: 4.071

7.  CONRAD--a software framework for cone-beam imaging in radiology.

Authors:  Andreas Maier; Hannes G Hofmann; Martin Berger; Peter Fischer; Chris Schwemmer; Haibo Wu; Kerstin Müller; Joachim Hornegger; Jang-Hwan Choi; Christian Riess; Andreas Keil; Rebecca Fahrig
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

8.  The feasibility of head motion tracking in helical CT: a step toward motion correction.

Authors:  Jung-Ha Kim; Johan Nuyts; Zdenka Kuncic; Roger Fulton
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

9.  A mathematical model of the patellofemoral joint.

Authors:  T M van Eijden; E Kouwenhoven; J Verburg; W A Weijs
Journal:  J Biomech       Date:  1986       Impact factor: 2.712

10.  Optimal short scan convolution reconstruction for fanbeam CT.

Authors:  D L Parker
Journal:  Med Phys       Date:  1982 Mar-Apr       Impact factor: 4.071

View more
  14 in total

1.  Motion compensation in extremity cone-beam computed tomography.

Authors:  Alejandro Sisniega; Gaurav K Thawait; Delaram Shakoor; Jeffrey H Siewerdsen; Shadpour Demehri; Wojciech Zbijewski
Journal:  Skeletal Radiol       Date:  2019-06-06       Impact factor: 2.199

2.  Dynamic detector offsets for field of view extension in C-arm computed tomography with application to weight-bearing imaging.

Authors:  Magdalena Herbst; Frank Schebesch; Martin Berger; Jang-Hwan Choi; Rebecca Fahrig; Joachim Hornegger; Andreas Maier
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

3.  Assessment of image quality in soft tissue and bone visualization tasks for a dedicated extremity cone-beam CT system.

Authors:  S Demehri; A Muhit; W Zbijewski; J W Stayman; J Yorkston; N Packard; R Senn; D Yang; D Foos; G K Thawait; L M Fayad; A Chhabra; J A Carrino; J H Siewerdsen
Journal:  Eur Radiol       Date:  2015-01-20       Impact factor: 5.315

4.  Symmetry prior for epipolar consistency.

Authors:  Alexander Preuhs; Andreas Maier; Michael Manhart; Markus Kowarschik; Elisabeth Hoppe; Javad Fotouhi; Nassir Navab; Mathias Unberath
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-07-12       Impact factor: 2.924

5.  Fiducial marker-based correction for involuntary motion in weight-bearing C-arm CT scanning of knees. II. Experiment.

Authors:  Jang-Hwan Choi; Andreas Maier; Andreas Keil; Saikat Pal; Emily J McWalter; Gary S Beaupré; Garry E Gold; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

6.  Marker-free motion correction in weight-bearing cone-beam CT of the knee joint.

Authors:  M Berger; K Müller; A Aichert; M Unberath; J Thies; J-H Choi; R Fahrig; A Maier
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

7.  Motion compensation in extremity cone-beam CT using a penalized image sharpness criterion.

Authors:  A Sisniega; J W Stayman; J Yorkston; J H Siewerdsen; W Zbijewski
Journal:  Phys Med Biol       Date:  2017-03-22       Impact factor: 3.609

8.  Motion compensation for cone-beam CT using Fourier consistency conditions.

Authors:  M Berger; Y Xia; W Aichinger; K Mentl; M Unberath; A Aichert; C Riess; J Hornegger; R Fahrig; A Maier
Journal:  Phys Med Biol       Date:  2017-08-21       Impact factor: 3.609

9.  Correction of patient motion in cone-beam CT using 3D-2D registration.

Authors:  S Ouadah; M Jacobson; J W Stayman; T Ehtiati; C Weiss; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2017-11-09       Impact factor: 3.609

10.  3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT.

Authors:  Jennifer Maier; Andreas Maier; Bjoern Eskofier; Rebecca Fahrig; Jang-Hwan Choi
Journal:  IEEE Access       Date:  2021-05-11       Impact factor: 3.367

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.