Fiducial marker-based correction for involuntary motion in weight-bearing C-arm CT scanning of knees. II. Experiment.

Jang-Hwan Choi1, Andreas Maier2, Andreas Keil2, Saikat Pal3, Emily J McWalter2, Gary S Beaupré4, Garry E Gold2, Rebecca Fahrig2.   

Abstract

PURPOSE: A C-arm CT system has been shown to be capable of scanning a single cadaver leg under loaded conditions by virtue of its highly flexible acquisition trajectories. In Part I of this study, using the 4D XCAT-based numerical simulation, the authors predicted that the involuntary motion in the lower body of subjects in weight-bearing positions would seriously degrade image quality and the authors suggested three motion compensation methods by which the reconstructions could be corrected to provide diagnostic image quality. Here, the authors demonstrate that a flat-panel angiography system is appropriate for scanning both legs of subjects in vivo under weight-bearing conditions and further evaluate the three motion-correction algorithms using in vivo data.
METHODS: The geometry of a C-arm CT system for a horizontal scan trajectory was calibrated using the PDS-2 phantom. The authors acquired images of two healthy volunteers while lying supine on a table, standing, and squatting at several knee flexion angles. In order to identify the involuntary motion of the lower body, nine 1-mm-diameter tantalum fiducial markers were attached around the knee. The static mean marker position in 3D, a reference for motion compensation, was estimated by back-projecting detected markers in multiple projections using calibrated projection matrices and identifying the intersection points in 3D of the back-projected rays. Motion was corrected using three different methods (described in detail previously): (1) 2D projection shifting, (2) 2D deformable projection warping, and (3) 3D rigid body warping. For quantitative image quality analysis, SSIM indices for the three methods were compared using the supine data as a ground truth.
RESULTS: A 2D Euclidean distance-based metric of subjects' motion ranged from 0.85 mm (±0.49 mm) to 3.82 mm (±2.91 mm) (corresponding to 2.76 to 12.41 pixels) resulting in severe motion artifacts in 3D reconstructions. Shifting in 2D, 2D warping, and 3D warping improved the SSIM in the central slice by 20.22%, 16.83%, and 25.77% in the data with the largest motion among the five datasets (SCAN5); improvement in off-center slices was 18.94%, 29.14%, and 36.08%, respectively.
CONCLUSIONS: The authors showed that C-arm CT control can be implemented for nonstandard horizontal trajectories which enabled us to scan and successfully reconstruct both legs of volunteers in weight-bearing positions. As predicted using theoretical models, the proposed motion correction methods improved image quality by reducing motion artifacts in reconstructions; 3D warping performed better than the 2D methods, especially in off-center slices.

Entities:  

Mesh:

Year:  2014        PMID: 24877813      PMCID: PMC4008764          DOI: 10.1118/1.4873675

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  29 in total

1.  Patellofemoral kinematics during weight-bearing and non-weight-bearing knee extension in persons with lateral subluxation of the patella: a preliminary study.

Authors:  Christopher M Powers; Samuel R Ward; Michael Fredericson; Marc Guillet; Frank G Shellock
Journal:  J Orthop Sports Phys Ther       Date:  2003-11       Impact factor: 4.751

2.  In-vivo measurement of dynamic joint motion using high speed biplane radiography and CT: application to canine ACL deficiency.

Authors:  Scott Tashman; William Anderst
Journal:  J Biomech Eng       Date:  2003-04       Impact factor: 2.097

3.  A method to estimate in vivo dynamic articular surface interaction.

Authors:  William J Anderst; Scott Tashman
Journal:  J Biomech       Date:  2003-09       Impact factor: 2.712

4.  Image quality assessment: from error visibility to structural similarity.

Authors:  Zhou Wang; Alan Conrad Bovik; Hamid Rahim Sheikh; Eero P Simoncelli
Journal:  IEEE Trans Image Process       Date:  2004-04       Impact factor: 10.856

5.  The mathematical theory of communication. 1963.

Authors:  C E Shannon
Journal:  MD Comput       Date:  1997 Jul-Aug

6.  Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy.

Authors:  S A Banks; W A Hodge
Journal:  IEEE Trans Biomed Eng       Date:  1996-06       Impact factor: 4.538

7.  Weight-bearing radiography in osteoarthritis of the knee.

Authors:  R E Leach; T Gregg; F J Siber
Journal:  Radiology       Date:  1970-11       Impact factor: 11.105

8.  Patellar tracking patterns during active and passive knee extension: evaluation with motion-triggered cine MR imaging.

Authors:  J Brossmann; C Muhle; C Schröder; U H Melchert; C C Büll; R P Spielmann; M Heller
Journal:  Radiology       Date:  1993-04       Impact factor: 11.105

9.  Geometrical calibration of X-ray imaging chains for three-dimensional reconstruction.

Authors:  A Rougée; C Picard; C Ponchut; Y Trousset
Journal:  Comput Med Imaging Graph       Date:  1993 Jul-Oct       Impact factor: 4.790

10.  Predicting the compressive mechanical behavior of bone.

Authors:  T S Keller
Journal:  J Biomech       Date:  1994-09       Impact factor: 2.712

View more
  13 in total

1.  Practical dose point-based methods to characterize dose distribution in a stationary elliptical body phantom for a cone-beam C-arm CT system.

Authors:  Jang-Hwan Choi; Dragos Constantin; Arundhuti Ganguly; Erin Girard; Richard L Morin; Robert L Dixon; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

2.  Motion compensation in extremity cone-beam computed tomography.

Authors:  Alejandro Sisniega; Gaurav K Thawait; Delaram Shakoor; Jeffrey H Siewerdsen; Shadpour Demehri; Wojciech Zbijewski
Journal:  Skeletal Radiol       Date:  2019-06-06       Impact factor: 2.199

3.  Dynamic detector offsets for field of view extension in C-arm computed tomography with application to weight-bearing imaging.

Authors:  Magdalena Herbst; Frank Schebesch; Martin Berger; Jang-Hwan Choi; Rebecca Fahrig; Joachim Hornegger; Andreas Maier
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

4.  Marker-free motion correction in weight-bearing cone-beam CT of the knee joint.

Authors:  M Berger; K Müller; A Aichert; M Unberath; J Thies; J-H Choi; R Fahrig; A Maier
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

5.  Motion compensation in extremity cone-beam CT using a penalized image sharpness criterion.

Authors:  A Sisniega; J W Stayman; J Yorkston; J H Siewerdsen; W Zbijewski
Journal:  Phys Med Biol       Date:  2017-03-22       Impact factor: 3.609

6.  Motion compensation for cone-beam CT using Fourier consistency conditions.

Authors:  M Berger; Y Xia; W Aichinger; K Mentl; M Unberath; A Aichert; C Riess; J Hornegger; R Fahrig; A Maier
Journal:  Phys Med Biol       Date:  2017-08-21       Impact factor: 3.609

7.  Validation of watershed-based segmentation of the cartilage surface from sequential CT arthrography scans.

Authors:  Mary E Hall; Marianne S Black; Garry E Gold; Marc E Levenston
Journal:  Quant Imaging Med Surg       Date:  2022-01

8.  Reference-free learning-based similarity metric for motion compensation in cone-beam CT.

Authors:  H Huang; J H Siewerdsen; W Zbijewski; C R Weiss; M Unberath; T Ehtiati; A Sisniega
Journal:  Phys Med Biol       Date:  2022-06-16       Impact factor: 4.174

9.  Image-Based Motion Compensation for High-Resolution Extremities Cone-Beam CT.

Authors:  A Sisniega; J W Stayman; Q Cao; J Yorkston; J H Siewerdsen; W Zbijewski
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-22

10.  3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT.

Authors:  Jennifer Maier; Andreas Maier; Bjoern Eskofier; Rebecca Fahrig; Jang-Hwan Choi
Journal:  IEEE Access       Date:  2021-05-11       Impact factor: 3.367

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.