| Literature DB >> 23516543 |
Ruifang Zhang1, Yi Song, Yuanyuan Zhou, Lulu Sun.
Abstract
BACKGROUND ANDEntities:
Mesh:
Year: 2013 PMID: 23516543 PMCID: PMC3597634 DOI: 10.1371/journal.pone.0058729
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Flowchart of study selection.
Basic characteristics of the 10 included studies involving 581 patients with aortic valve stenosis.
| Author | Country | Year of publication | Numberof patients | Meanage (y) | Sex, men (%) | View on TTE | View on TEE | View on MDCT | Measurementphase | Surgical measurement |
| Wood | Canada | 2009 | 26 | 82±9 | Unavailable (NA) | Para-sternallong axis | Long axis | Coronal andSagittal view | End-diastole | No |
| Messika-Zeitoun | France | 2010 | 45 | 80±8 | 58 | Para-sternallong axis | Long axis | Three chamberview | Mid-systole | No |
| Delgado | Netherlands | 2010 | 53 | 80±8 | 55 | Para-sternallong axis | NA | Coronal andSagittal view | NA | No |
| Hutter | Germany | 2010 | 187 | 81±7 | 37 | Para-sternallong axis | Long axis | Three chamberview | NA | No |
| Altiok | Germany | 2011 | 49 | 81±7 | 33 | Para-sternallong axis | Long axis | Coronal andSagittal view | End-diastole | No |
| Dashkevich | Germany | 2011 | 33 | 77±8 | 55 | NA | Long axis | Coronal andSagittal view | End-diastole | Yes |
| Koos | Germany | 2011 | 58 | 83±6 | 26 | NA | Long axis | Coronal andSagittal view | Diastole | No |
| Mizia-Stec | Poland | 2011 | 20 | 69±6 | 85 | Para-sternallong axis | Long axis | Coronal andSagittal view | NA | Yes |
| Mesa Rubio | Spain | 2011 | 40 | 77±4 | 54 | Para-sternallong axis | Long axis | Coronal andSagittal view | End-diastole | No |
| Tzikas | Netherlands | 2011 | 70 | 82 | 51 | Para-sternallong axis | NA | Coronal andSagittal view | Mid-systole | No |
AS: Aortic valve stenosis; NA: Unavailable; TTE: transthoracic echocardiography; TEE: transesophegeal echocardiography; MDCT: multi-detector computed tomography.
Meta-analysis of the weighted mean differences of aortic annulus measurement between MDCT, TTE, and TEE.
| Methods | n | Weighted Mean Differences | Test of Heterogeneity | |||
| Estimate (95% CI)(mm) | P value | Q-test(P) | I2(95% CI) | |||
| MDCT Coronal View | vs. Sagittal view | 279 | 2.53(1.38 to 3.68) | <0.0001 | <0.001 | 90(83 to 94) |
| vs. TTE | 188 | 2.97(1.98 to 3.96) | <0.0001 | <0.001 | 79(54 to 90) | |
| vs. TEE | 222 | 1.74(1.30 to 2.19) | <0.0001 | 0.39 | 3(0 to 76) | |
| vs. Surgical | 53 | 2.96(0.51 to 5.40) | 0.018 | 0.04 | NA | |
| MDCT Sagittal View | vs. TTE | 188 | 0.17(−0.68 to 1.02) | 0.70 | 0.001 | 76(45 to 89) |
| vs. TEE | 222 | −0.22(−0.93 to 0.50) | 0.55 | 0.02 | 62(7 to 84) | |
| vs. Surgical | 53 | 1.40(−2.80 to 5.60) | 0.51 | <0.001 | NA | |
| MDCT Three Chamber View | vs. TTE | 232 | 0.23(−0.19 to 0.64) | 0.28 | 0.47 | NA |
| vs. TEE | 232 | −0.30(−0.70 to 0.09) | 0.12 | 0.99 | NA | |
| TEE | vs. TTE | 363 | 0.45(0.12 to 0.78) | 0.007 | 0.61 | 0(0 to 75) |
| vs. Surgical | 53 | 1.09(−2.63 to 4.81) | 0.56 | 0.002 | NA | |
MDCT: Multi-detector computed tomography, TTE: Transthoracic echocardiography, TEE: Transesophegeal echocardiography; NA, not available for ≤2 studies included in the meta-analysis.
Figure 2Meta-analysis of the weighted mean difference of aortic annulus diameter measurement between coronal view on MDCT and TTE.
Figure 3Meta-analysis of the weighted mean difference of aortic annulus diameter measurement between coronal view and sagittal view on MDCT.
Figure 4Meta-analysis of the weighted mean difference of aortic annulus diameter measurement between coronal view on MDCT and TEE.
Figure 5Meta-analysis of the weighted mean difference of aortic annulus diameter measurement between TEE and TTE.