Aortic annulus evaluation in transcatheter aortic valve implantation.

Andrea Hutter1, Anke Opitz, Sabine Bleiziffer, Hendrik Ruge, Ina Hettich, Domenico Mazzitelli, Albrecht Will, Peter Tassani, Robert Bauernschmitt, Ruediger Lange.   

Abstract

OBJECTIVES: We compared the annulus diameters measured by transthoracic echocardiography (TTE), transesophageal echocardiography (TEE), and dual-source computed tomography (DSCT) before transcatheter aortic valve implantation (TAVI).
BACKGROUND: In TAVI correct evaluation of the aortic annulus is mandatory to choose the correct prosthesis type and size and to prevent complications. There is no gold standard for the assessment of aortic annulus diameters.
METHODS: Preoperative assessment of the aortic annulus with TTE, TEE, and DSCT was performed in 187 consecutive patients referred for TAVI between June 2007 and May 2009.
RESULTS: The mean aortic annuli were 22.6 ± 2.0 mm measured with DSCT, 22.3 ± 2.5 mm with TTE, and 22.9 ± 2.2 mm with TEE. Despite a strong correlation between the measurement techniques, relevant statistical spread occurred with differences up to 3 mm in all measurement methods. Inter- and intraobserver variability was good for TEE and less satisfactory for DSCT measurements. TEE measurements taken as decisive parameter for the implantation changed the implantation strategy in 15.5% of patients and did not show an increased rate of procedural complications.
CONCLUSION: Despite a strong correlation, the measurement techniques for the aortic annulus show relevant statistical spread, consequently one measurement technique cannot definitely predict another. TEE measurements show a more satisfactory intra- and interobserver variability than DSCT. Taking TEE annulus measurements as decisive parameter for the implantation has an impact on the implantation strategy and is safe with a low rate of procedural complications.
Copyright © 2010 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20518012     DOI: 10.1002/ccd.22654

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  14 in total

Review 1.  In vivo imaging and computational analysis of the aortic root. Application in clinical research and design of transcatheter aortic valve systems.

Authors:  Paul Schoenhagen; Alexander Hill; Tim Kelley; Zoran Popovic; Sandra S Halliburton
Journal:  J Cardiovasc Transl Res       Date:  2011-04-12       Impact factor: 4.132

2.  CT evaluation prior to transapical aortic valve replacement: semi-automatic versus manual image segmentation.

Authors:  Borek Foldyna; Camelia Jungert; Christian Luecke; Konstantin von Aspern; Sonja Boehmer-Lasthaus; Eva Maria Rueth; Matthias Grothoff; Stefan Nitzsche; Matthias Gutberlet; Friedrich Wilhelm Mohr; Lukas Lehmkuhl
Journal:  Int J Cardiovasc Imaging       Date:  2015-04-19       Impact factor: 2.357

3.  Reproducibility of aortic annulus measurements by computed tomography.

Authors:  Annika Schuhbaeck; Stephan Achenbach; Tobias Pflederer; Mohamed Marwan; Jasmin Schmid; Holger Nef; Johannes Rixe; Franziska Hecker; Christian Schneider; Michael Lell; Michael Uder; Martin Arnold
Journal:  Eur Radiol       Date:  2014-05-22       Impact factor: 5.315

4.  Assessment of the aortic valve annular geometry by real-time three-dimensional transthoracic echocardiography: comparison with two-dimensional transthoracic echocardiography and multidetector computed tomography.

Authors:  Chikage Oshita; Kazuya Murata; Yasuaki Wada; Tomoko Nao; Kosuke Uchida; Shinichi Okuda; Takehisa Susa; Nobuaki Tanaka; Masunori Matsuzaki; Masafumi Yano
Journal:  J Echocardiogr       Date:  2013-11-23

5.  Multi-detector computed tomography is equivalent to trans-oesophageal echocardiography for the assessment of the aortic annulus before transcatheter aortic valve implantation.

Authors:  Johannes Rixe; Annika Schuhbaeck; Christoph Liebetrau; Helge Moellmann; Holger M Nef; Sebastian Szardien; Roland Brandt; Joern Schmitt; Thomas Neumann; Christian Schneider; Gabriele Krombach; Christian W Hamm; Stephan Achenbach; Andreas Rolf
Journal:  Eur Radiol       Date:  2012-06-12       Impact factor: 5.315

6.  Computer Vision Techniques for Transcatheter Intervention.

Authors:  Feng Zhao; Xianghua Xie; Matthew Roach
Journal:  IEEE J Transl Eng Health Med       Date:  2015-06-18       Impact factor: 3.316

7.  Inter-individual variance and cardiac cycle dependency of aortic root dimensions and shape as assessed by ECG-gated multi-slice computed tomography in patients with severe aortic stenosis prior to transcatheter aortic valve implantation: is it crucial for correct sizing?

Authors:  Lukas Lehmkuhl; Borek Foldyna; Konstantin Von Aspern; Christian Lücke; Matthias Grothoff; Stefan Nitzsche; Jörg Kempfert; Martin Haensig; Ardawan Rastan; Thomas Walther; Friedrich-Wilhelm Mohr; Matthias Gutberlet
Journal:  Int J Cardiovasc Imaging       Date:  2012-09-18       Impact factor: 2.357

8.  Edwards Sapien XT valve placement as treatment option for aortic regurgitation after transfemoral CoreValve implantation: a multicenter experience.

Authors:  Patrick Diemert; Philipp Lange; Martin Greif; Moritz Seiffert; Lenard Conradi; Steffen Massberg; Stefan Blankenberg; Hermann Reichenspurner; Christian Hagl; Christoph Schmitz; Holger Schröfel; Hendrik Treede; Gerhard Schymik; Christian Kupatt
Journal:  Clin Res Cardiol       Date:  2013-11-20       Impact factor: 5.460

9.  Major themes for 2010 in cardiothoracic and vascular anesthesia.

Authors:  H Riha; J Fassl; P Patel; T Wyckoff; J Choi; J G Augoustides
Journal:  HSR Proc Intensive Care Cardiovasc Anesth       Date:  2011

10.  Transesophageal echocardiography measurements of aortic annulus diameter using biplane mode in patients undergoing transcatheter aortic valve implantation.

Authors:  Kambiz Shahgaldi; Cristina da Silva; Magnus Bäck; Andreas Rück; Aristomenis Manouras; Anders Sahlén
Journal:  Cardiovasc Ultrasound       Date:  2013-01-30       Impact factor: 2.062

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.