Evaluation of aortic root for definition of prosthesis size by magnetic resonance imaging and cardiac computed tomography: implications for transcatheter aortic valve implantation.

Ralf Koos1, Ertunc Altiok, Andreas Horst Mahnken, Mirja Neizel, Guido Dohmen, Nikolaus Marx, Harald Kühl, Rainer Hoffmann.   

Abstract

BACKGROUND: This study sought to compare cardiac magnetic resonance imaging (CMR) with dual source computed tomography (DSCT) for analysis of aortic root dimensions prior to transcatheter aortic valve implantation (TAVI). In addition, the potential impact of CMR and DSCT measurements on TAVI strategy defined by 2D-transesophageal echocardiography (TEE) was evaluated.
METHODS: Aortic root dimensions were measured using CMR and DSCT in 58 patients referred for evaluation of TAVI. The TAVI strategy (choice of prosthesis size and decision to implant) was based on 2D-TEE annulus measurements.
RESULTS: CMR and DSCT aortic root measurements showed an overall good correlation (r=0.86, p<0.001 for coronal aortic annulus diameters). There was also a good correlation between TEE and CMR as well as between TEE and DSCT for measurement of sagittal aortic annulus diameters (r=0.69, p<0.001). However, annulus diameters assessed by TEE (22.1±2.3mm) were significantly smaller than coronal aortic annulus diameters assessed by CMR (23.4±1.8mm, p<0.001) or DSCT (23.6±1.8, p<0.001). Regarding TAVI strategy, the agreement between TEE and sagittal CMR (kappa=0.89) as well as sagittal DSCT measurements (kappa=0.87) was statistically perfect. However, decision based on coronal CMR- or MSCT measurements would have modified TAVI strategy as compared to a TEE based choice in a significant number of patients (22% to 24%).
CONCLUSION: In patients referred for TAVI, CMR measurements of aortic root dimensions show a good correlation with DSCT measurements and thus CMR may be an alternative 3D-imaging modality. Aortic annulus measurements using TEE, CMR and DSCT were close but not identical and the method used has important potential implications on TAVI strategy.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21315460     DOI: 10.1016/j.ijcard.2011.01.044

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  23 in total

Review 1.  Multimodality imaging in interventional cardiology.

Authors:  Bas L van der Hoeven; Martin J Schalij; Victoria Delgado
Journal:  Nat Rev Cardiol       Date:  2012-02-14       Impact factor: 32.419

2.  Non-contrast-enhanced magnetic resonance angiography is equal to contrast-enhanced multislice computed tomography for correct aortic sizing before transcatheter aortic valve implantation.

Authors:  Peter Bernhardt; Christoph Rodewald; Julia Seeger; Birgid Gonska; Dominik Buckert; Michael Radermacher; Vinzenz Hombach; Wolfgang Rottbauer; Jochen Wöhrle
Journal:  Clin Res Cardiol       Date:  2015-09-22       Impact factor: 5.460

Review 3.  Guidance of transcatheter aortic valve replacement by echocardiography.

Authors:  Rebecca T Hahn
Journal:  Curr Cardiol Rep       Date:  2014-01       Impact factor: 2.931

4.  A non-contrast self-navigated 3-dimensional MR technique for aortic root and vascular access route assessment in the context of transcatheter aortic valve replacement: proof of concept.

Authors:  Matthias Renker; Akos Varga-Szemes; U Joseph Schoepf; Stefan Baumann; Davide Piccini; Michael O Zenge; Wolfgang G Rehwald; Edgar Müller; Jeremy D Rier; Helge Möllmann; Christian W Hamm; Daniel H Steinberg; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2015-07-20       Impact factor: 5.315

5.  Multi-parametric approach to predict prosthetic valve size using CMR and clinical data: insights from SAVR.

Authors:  Federico E Mordini; Conor F Hynes; Richard L Amdur; Jeffrey Panting; Dominic A Emerson; Jason Morrissette; Erin Goheen-Thomas; Michael D Greenberg; Gregory D Trachiotis
Journal:  Int J Cardiovasc Imaging       Date:  2021-03-10       Impact factor: 2.357

6.  Assessment of a transcatheter heart valve prosthesis with multidetector computed tomography: in vitro and in vivo imaging characteristics.

Authors:  Linda M de Heer; Jesse Habets; Jolanda Kluin; Pieter R Stella; Willem P Th M Mali; Lex A van Herwerden; Ricardo P J Budde
Journal:  Int J Cardiovasc Imaging       Date:  2012-08-09       Impact factor: 2.357

Review 7.  Paravalvular regurgitation following transcutaneous aortic valve replacement: predictors and clinical significance.

Authors:  Rebecca T Hahn; Susheel Kodali; Philippe Généreux; Martin Leon
Journal:  Curr Cardiol Rep       Date:  2014-05       Impact factor: 2.931

8.  Exact monitoring of aortic diameters in Marfan patients without gadolinium contrast: intraindividual comparison of 2D SSFP imaging with 3D CE-MRA and echocardiography.

Authors:  Simon Veldhoen; Cyrus Behzadi; Thorsten Derlin; Meike Rybczinsky; Yskert von Kodolitsch; Sara Sheikhzadeh; Frank Oliver Henes; Thorsten Alexander Bley; Gerhard Adam; Peter Bannas
Journal:  Eur Radiol       Date:  2014-10-15       Impact factor: 5.315

9.  Practical update on imaging and transcatheter aortic valve implantation.

Authors:  Gisela Feltes; Iván J Núñez-Gil
Journal:  World J Cardiol       Date:  2015-04-26

Review 10.  The role of echocardiography in transcatheter aortic valve implantation.

Authors:  Toshinari Onishi; Kaoruko Sengoku; Yasuhiro Ichibori; Isamu Mizote; Koichi Maeda; Toru Kuratani; Yoshiki Sawa; Yasushi Sakata
Journal:  Cardiovasc Diagn Ther       Date:  2018-02