Abstract

OBJECTIVES: The purpose of this study was to determine the predictors of aortic regurgitation (AR) after transcatheter aortic valve implantation (TAVI).
BACKGROUND: TAVI has been associated with a high rate of paravalvular regurgitation, usually mild. Nevertheless, moderate to severe regurgitations still occur and may have negative clinical consequences.
METHODS: Fifty patients with severe aortic stenosis were recruited and underwent successful TAVI with the Medtronic CoreValve bioprosthesis through the transfemoral route. The end point of this study is the early occurrence of significant AR, defined as the occurrence of grade II or more AR by post-procedural aortography.
RESULTS: The study population's mean age was 80.5 ± 7.9 years, with a mean aortic valve area of 0.64 ± 0.17 cm². Post-procedural AR was absent in 3 patients and was grade I in 27 patients, grade II in 13 patients, and grade III in 7 patients. Using univariate analysis, the chance of significant AR increased with increasing angle of left ventricular outflow tract to ascending aorta (∠LVOT-AO) (odds ratio: 1.24, p < 0.001). For the depth of the device in relation to the noncoronary cusp, there was a minimum chance of AR corresponding to depth = 9.5 mm (odds ratio: 1.1, p = 0.01). Using multivariate analysis, we found a greater chance of significant AR with a greater angle (odds ratio: 1.24, p = 0.001), and that the chance of significant AR is a minimum when depth of the device in relation to the noncoronary cusp is ∼10 mm (odds ratio: 1.1, p = 0.024). A predictive model was generated, and if 2 ×∠LVOT-AO + (depth to noncoronary cusp - 10)² ≥ 50, the likelihood of occurrence of significant AR could be predicted with a sensitivity of 85% and a specificity of 87%.
CONCLUSIONS: The occurrence of significant AR after TAVI can be predicted by anatomic and procedural variables. A model such as that presented can be used to select suitable patients for this procedure and guide operators during implantation of the device.
Copyright © 2010 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21050971     DOI: 10.1016/j.jacc.2010.06.035

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  35 in total

1.  Transcatheter resection of the native aortic valve prior to endovalve implantation - A rational approach to reduce TAVI-induced complications.

Authors:  Parla Astarci; Pierre-Yves Etienne; Benoit Raucent; Xavier Bollen; Kahn Tranduy; David Glineur; Laurent Dekerchove; Philippe Noirhomme; Gébrine Elkhoury
Journal:  Ann Cardiothorac Surg       Date:  2012-07

2.  Aortic annulus angulation does not attenuate procedural success of transcatheter aortic valve replacement using a novel self-expanding bioprosthesis.

Authors:  Giuseppe D'Ancona; Stephan Kische; Mohamed El-Mawardy; Martin Dißmann; Helmut Heinze; Dietlind Zohlnhöfer-Momm; Hakan Gürer; Hüseyin Ince
Journal:  Heart Vessels       Date:  2019-05-27       Impact factor: 2.037

3.  Angles between the aortic root and the left ventricle assessed by MDCT are associated with the risk of aortic regurgitation after transcatheter aortic valve replacement.

Authors:  Vincent Roule; Alexandre Placente; Rémi Sabatier; Mathieu Bignon; Vladimir Saplacan; Calin Ivascau; Paul Milliez; Farzin Beygui
Journal:  Heart Vessels       Date:  2017-08-10       Impact factor: 2.037

Review 4.  Standard imaging techniques in transcatheter aortic valve replacement.

Authors:  Arash Salemi; Berhane M Worku
Journal:  J Thorac Dis       Date:  2017-04       Impact factor: 2.895

Review 5.  TAVI 2012: state of the art.

Authors:  Jochen Reinöhl; Constantin von Zur Mühlen; Martin Moser; Stefan Sorg; Christoph Bode; Manfred Zehender
Journal:  J Thromb Thrombolysis       Date:  2013-05       Impact factor: 2.300

Review 6.  Paravalvular regurgitation following transcutaneous aortic valve replacement: predictors and clinical significance.

Authors:  Rebecca T Hahn; Susheel Kodali; Philippe Généreux; Martin Leon
Journal:  Curr Cardiol Rep       Date:  2014-05       Impact factor: 2.931

7.  The Impact of Size and Position of a Mechanical Expandable Transcatheter Aortic Valve: Novel Insights Through Computational Modelling and Simulation.

Authors:  Giorgia Rocatello; Nahid El Faquir; Ole de Backer; Martin J Swaans; Azeem Latib; Luca Vicentini; Patrick Segers; Matthieu De Beule; Peter de Jaegere; Peter Mortier
Journal:  J Cardiovasc Transl Res       Date:  2019-08-23       Impact factor: 4.132

Review 8.  Transcatheter treatment approaches for aortic valve disease.

Authors:  Alex Willson; John Webb
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-24       Impact factor: 2.357

9.  Balloon- or Self-Expandable TAVI: Clinical Equipoise?

Authors:  John Jose; Gert Richardt; Mohamed Abdel-Wahab
Journal:  Interv Cardiol       Date:  2015-05

Review 10.  Transcatheter Valve Replacement: Risk Levels and Contemporary Outcomes.

Authors:  Manuel Reyes; Michael J Reardon
Journal:  Methodist Debakey Cardiovasc J       Date:  2017 Jul-Sep
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.