Abstract
The study was conducted to evaluate the intra-session repeatability of Tonopen AVIA (TPA). 180 eyes of 180 patients (50 eyes with glaucoma, 130 eyes of controls) were recruited for this observational study. The mean age of patients enrolled in the study was 43.9 ± 16.7 yrs (84 males, 96 females). Mean IOP recorded with Tonopen AVIA was 19.5 ± 9.5 mmHg, 19.4 ± 9.6 mmHg and 19.3 ± 9.2 mmHg, respectively in the first, second and third instances (P = 0.656). The intraclass correlation coefficient (ICC) ranged from 0.996 (95% CI: 0.956 - 0.998) for glaucoma subjects to 0.958 (95% CI: 0.934 - 0.975) for controls. The coefficient of variation in the study population ranged from 3.47% (glaucoma patients) to 8.10% (healthy controls), being 6.07% overall. The coefficient of repeatability varied between 2.96 (glaucoma patients), 3.35 (healthy controls) to 3.24 (overall). Thus, the Tonopen Avia shows good intrasessional repeatability of IOP in both glaucomatous patients and healthy subjects.Entities:
Mesh:
Year: 2013 PMID: 23514651 PMCID: PMC3665044 DOI: 10.4103/0301-4738.109384
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
The mean IOP recordings in the three instances in glaucoma patients and healthy subjects
Comparison of coefficient of repeatability, coefficient of variation and intrasession within subject standard deviation
Figure 1Bland Altman plot showing agreement between Goldmann applanation tonometer and TonoPen Avia