Comparisons between Pascal dynamic contour tonometry, the TonoPen, and Goldmann applanation tonometry in patients with glaucoma.

Maria L Salvetat1, Marco Zeppieri, Claudia Tosoni, Paolo Brusini.   

Abstract

PURPOSE: To compare intraocular pressure (IOP) measurements taken with Pascal dynamic contour tonometry (DCT), the TonoPen and the Goldmann applanation tonometry (GAT). The influence of central corneal thickness (CCT) on IOP measurements taken with Pascal DCT and the TonoPen was evaluated.
METHODS: One eye in each of 101 consecutive patients with primary open-angle glaucoma (POAG) underwent ultrasonic CCT measurement and IOP evaluation with GAT, Pascal DCT and the TonoPen in random order. The agreement between results from Pascal DCT and the TonoPen and those of GAT was assessed using the Bland-Altman method. The deviation of Pascal DCT and TonoPen readings from GAT values, corrected for CCT, was calculated and correlated to CCT using a linear regression model.
RESULTS: The mean of the differences in IOP measurements was 3.2 +/- 2.4 mmHg for Pascal DCT minus GAT readings and 0.5 +/- 4.5 mmHg for TonoPen minus GAT readings. The 95% confidence interval of differences in IOP measurements was higher between TonoPen and GAT readings (- 6 to 7 mmHg) than between Pascal and GAT readings (0.1-6.8 mmHg). Pascal DCT significantly overestimated IOP compared with GAT, especially for higher IOP readings. Bland-Altman scatterplots showed reasonable inter-method agreement between Pascal DCT and GAT measurements, and poor agreement between TonoPen and GAT measurements. The deviations of Pascal DCT and TonoPen readings from the corrected GAT values were both highly correlated with CCT values (linear regression analysis, p < 0.0001). The mean change in measured IOP for a 10-microm increase in CCT was 0.48 mmHg for Pascal DCT and 0.74 mmHg for the TonoPen.
CONCLUSIONS: Agreement with GAT measurements was higher for Pascal DCT than for TonoPen readings; however, Pascal DCT significantly overestimated IOP values compared with GAT. Measurements of IOP obtained with both Pascal DCT and the TonoPen appeared to be influenced by CCT, and this influence appeared to be greater for the latter.

Entities:  

Mesh:

Year:  2007        PMID: 17488456     DOI: 10.1111/j.1600-0420.2006.00834.x

Source DB:  PubMed          Journal:  Acta Ophthalmol Scand        ISSN: 1395-3907


  19 in total

1.  Intraocular pressure after Descemet's stripping and non-Descemet's stripping automated endothelial keratoplasty.

Authors:  Yoshiro Mawatari; Akira Kobayashi; Hideaki Yokogawa; Kazuhisa Sugiyama
Journal:  Jpn J Ophthalmol       Date:  2011-03-13       Impact factor: 2.447

2.  Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication.

Authors:  Shikha Gupta; Gautam Sinha; Reetika Sharma; Bhagabat Nayak; Bharat Patil; Bibhuti Kashyap; Abdul Shameer; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

Review 3.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

4.  Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry.

Authors:  Shibal Bhartiya; Shveta Jindal Bali; Reetika Sharma; Neha Chaturvedi; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2011-07-15       Impact factor: 2.031

5.  Goldmann applanation tonometry and dynamic contour tonometry in eyes with elevated intraocular pressure (IOP): comparison in the same eyes after subsequent medical normalization of IOP.

Authors:  Chungkwon Yoo; Young Sub Eom; Yong Yeon Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-07-22       Impact factor: 3.117

6.  Comparison of ocular response analyzer, dynamic contour tonometer and Goldmann applanation tonometer.

Authors:  Charlotte Renier; Thierry Zeyen; Steffen Fieuws; Sofie Vandenbroeck; Ingeborg Stalmans
Journal:  Int Ophthalmol       Date:  2010-05-25       Impact factor: 2.031

7.  Comparison of Goldmann applanation and dynamic contour tonometry in a population of Mexican open-angle glaucoma patients.

Authors:  J Jimenez-Roman; F Gil-Carrasco; A Martinez; O Albis-Donado; J D De la Torre-Tovar
Journal:  Int Ophthalmol       Date:  2012-11-16       Impact factor: 2.031

8.  Correlation of intraocular pressure measured with goldmann and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Tony Realini; Robert N Weinreb; Gerald Hobbs
Journal:  J Glaucoma       Date:  2009-02       Impact factor: 2.503

9.  Comparison of Goldmann and Pascal tonometry in relation to corneal hysteresis and central corneal thickness in nonglaucomatous eyes.

Authors:  G Mangouritsas; S Mourtzoukos; A Mantzounis; L Alexopoulos
Journal:  Clin Ophthalmol       Date:  2011-08-03

10.  Test retest variability of TonoPen AVIA.

Authors:  Shibal Bhartiya; Shveta J Bali; Mathew James; Anita Panda; Tanuj Dada
Journal:  Indian J Ophthalmol       Date:  2013-03       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.