Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer.

Kelechi C Ogbuehi1.   

Abstract

BACKGROUND: The reliability of non-contact tonometers has been reported extensively in the literature. This study was designed to assess reliability of the new Topcon CT80 non-contact tonometer in normotensive subjects, using the Goldmann tonometer as the standard.
METHODS: The accuracy of the Topcon CT 80 non-contact tonometer was assessed by comparing its IOP assessments with those of the Goldmann applanation tonometer, on 60 right eyes of young healthy subjects with normal intraocular pressures. Each subject's intraocular pressure was assessed with each technique on two separate occasions, one week apart. The reliability of each technique was determined by the assessment of its inter-session repeatability using the Bland-Altman method. The 95 per cent limits of agreement for the two methods were also determined.
RESULTS: No statistically significant difference was found between the average intraocular pressures measured with the two techniques (p > 0.05). The inter-session repeatability indices for the two techniques did not differ significantly (p > 0.05). The mean difference in intraocular measurements between the two techniques was 0.2 +/- 1.5 mmHg (mean +/- SD) and the 95 per cent limits of agreement were -3.14 and +2.74 mmHg, with the non-contact tonometer returning higher readings than the Goldmann tonometer.
CONCLUSION: In this sample of normotensive subjects, the Topcon CT80 non-contact tonometer proved to be accurate and as reliable as the Goldmann tonometer in the assessment of intraocular pressure. Thus, it can be used as an objective clinical method for the assessment of normal intraocular pressure.

Entities:  

Mesh:

Year:  2006        PMID: 16907669     DOI: 10.1111/j.1444-0938.2006.00068.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  20 in total

1.  Intraocular pressure alterations after visual field testing.

Authors:  Akira Sawada; Hiroki Yamada; Yoshikazu Yamamoto; Tetsuya Yamamoto
Journal:  Jpn J Ophthalmol       Date:  2014-07-08       Impact factor: 2.447

2.  Clinical investigation of the effect of topical anesthesia on intraocular pressure.

Authors:  Turki M Almubrad; Kelechi C Ogbuehi
Journal:  Clin Ophthalmol       Date:  2007-09

3.  Comparison of IOPen rebound tonometer with Goldmann applanation tonometer at different IOP levels.

Authors:  Fereydoun Farrahi; Farideh Sharifipour; Mohammad Malekahmadi; Bahman Cheraghian
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

4.  Performance of the PT100 noncontact tonometer in healthy eyes.

Authors:  Turki M AlMubrad
Journal:  Clin Ophthalmol       Date:  2011-05-19

5.  Two-position measurement of intraocular pressure by PT100 noncontact tonometry in comparison with Goldmann tonometry.

Authors:  Kelechi C Ogbuehi; John C Chijuka; Uchechukwu L Osuagwu
Journal:  Clin Ophthalmol       Date:  2011-09-06

6.  Intraocular pressure measurement by three different tonometers in primary congenital glaucoma.

Authors:  Athar Zareei; Mohammad Reza Razeghinejad; Mohammad Hosein Nowroozzadeh; Yadollah Mehrabi; Mohammad Aghazadeh-Amiri
Journal:  J Ophthalmic Vis Res       Date:  2015 Jan-Mar

7.  Test retest variability of TonoPen AVIA.

Authors:  Shibal Bhartiya; Shveta J Bali; Mathew James; Anita Panda; Tanuj Dada
Journal:  Indian J Ophthalmol       Date:  2013-03       Impact factor: 1.848

8.  Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-Pen XL.

Authors:  Ihsan Yilmaz; Cigdem Altan; Ebru Demet Aygit; Cengiz Alagoz; Okkes Baz; Sibel Ahmet; Semih Urvasizoglu; Dilek Yasa; Ahmet Demirok
Journal:  Clin Ophthalmol       Date:  2014-06-07

9.  Repeatability and reproducibility of measurements using a NT-530P noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure.

Authors:  Fusako Fujimura; Kazutaka Kamiya; Kazuko Fujiwara; Nobuyuki Shoji; Kimiya Shimizu
Journal:  Biomed Res Int       Date:  2013-10-08       Impact factor: 3.411

10.  A Longitudinal Study of Association between Adiposity Markers and Intraocular Pressure: The Kangbuk Samsung Health Study.

Authors:  Di Zhao; Myung Hun Kim; Roberto Pastor-Barriuso; Yoosoo Chang; Seungho Ryu; Yiyi Zhang; Sanjay Rampal; Hocheol Shin; Joon Mo Kim; David S Friedman; Eliseo Guallar; Juhee Cho
Journal:  PLoS One       Date:  2016-01-05       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.