Literature DB >> 18091174

Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure.

Aimee Teo Broman1, Nathan G Congdon, Karen Bandeen-Roche, Harry A Quigley.   

Abstract

PURPOSE: To estimate the relationships between ocular parameters and tonometrically measured intraocular pressure (IOP), to determine the influence of ocular parameters on different instrument measurements of IOP, and to evaluate the association of ocular parameters with a parameter called hysteresis.
METHODS: Patients presenting at a glaucoma clinic were recruited for this study. Subjects underwent IOP measurement with the Goldmann applanation tonometer (GAT), the TonoPen, and the Reichert Ocular Response Analyzer (ORA), and also measurements of central corneal thickness (CCT), axial length, corneal curvature, corneal astigmatism, central visual acuity, and refractive error. Chart information was reviewed to determine glaucoma treatment history. The ORA instrument provided a measurement called corneal hysteresis. The association between measured IOP and the other ocular characteristics was estimated using generalized estimating equations.
RESULTS: Among 230 patients, IOP measurements from the TonoPen read lowest, and ORA read highest, and GAT measurements were closest to the mean IOP of the 3 instruments. In a multiple regression model adjusting for age, sex, race, and other ocular characteristics, a 10 microm increase in CCT was associated with an increase of 0.79 mm Hg measured IOP in untreated eyes (P<0.0001). Of the 3 tonometers, GAT was the least affected by CCT (0.66 mm Hg/10 mum, P<0.0001). Hysteresis was significantly correlated with CCT with a modest correlation coefficient (r=0.20, P<0.0007).
CONCLUSIONS: Among parameters related to measured IOP, features in addition to CCT, such as hysteresis and corneal curvature, may also be important. Tonometric instruments seem to be affected differently by various physiologic characteristics.

Entities:  

Mesh:

Year:  2007        PMID: 18091174     DOI: 10.1097/IJG.0b013e3180640f40

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  42 in total

1.  Cornea plana associated with open-angle glaucoma: a case report.

Authors:  Bilge Ozturk Sahin; Goktug Seymenoglu; Esin F Baser
Journal:  Int Ophthalmol       Date:  2011-12-11       Impact factor: 2.031

2.  Corneal biomechanical characteristics in children with diabetes mellitus.

Authors:  Pınar Nalcacioglu-Yuksekkaya; Emine Sen; Semra Cetinkaya; Veysel Bas; Zehra Aycan; Faruk Ozturk
Journal:  Int Ophthalmol       Date:  2014-01-23       Impact factor: 2.031

3.  Distribution of intraocular pressure and its determinants in an Iranian adult population.

Authors:  Hassan Hashemi; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Abbasali Yekta; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2016-08-18       Impact factor: 1.779

4.  Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry.

Authors:  Shibal Bhartiya; Shveta Jindal Bali; Reetika Sharma; Neha Chaturvedi; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2011-07-15       Impact factor: 2.031

5.  Corneal biomechanical changes and intraocular pressure in patients with thyroid orbitopathy.

Authors:  Zofia Pniakowska; Anna Klysik; Roman Gos; Piotr Jurowski
Journal:  Int J Ophthalmol       Date:  2016-03-18       Impact factor: 1.779

6.  Combining corneal hysteresis with central corneal thickness and intraocular pressure for glaucoma risk assessment.

Authors:  D Pensyl; M Sullivan-Mee; M Torres-Monte; K Halverson; C Qualls
Journal:  Eye (Lond)       Date:  2012-08-10       Impact factor: 3.775

7.  Intraocular pressure changes following Descemet's stripping with endothelial keratoplasty.

Authors:  Edgar M Espana; Zachary M Robertson; Bo Huang
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-11-20       Impact factor: 3.117

8.  Central corneal thickness measurements in premature infants.

Authors:  Murat Gunay; Gokhan Celik; Betul Onal Gunay; Mahmut Dogru; Tugba Gursoy; Husnu Fahri Ovali
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

9.  Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer.

Authors:  Mana Hagishima; Kazutaka Kamiya; Fusako Fujimura; Tetsuya Morita; Nobuyuki Shoji; Kimiya Shimizu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-09-29       Impact factor: 3.117

10.  Factors affecting corneal hysteresis in normal eyes.

Authors:  Kazutaka Kamiya; Mana Hagishima; Fusako Fujimura; Kimiya Shimizu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-06-11       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.