The influence of third-body particles on wear rate in unicondylar knee arthroplasty: a wear simulator study with bone and cement debris.

Christian Schroeder1, Thomas M Grupp, Bernhard Fritz, Christoph Schilling, Yan Chevalier, Sandra Utzschneider, Volkmar Jansson.   

Abstract

The reduced intraoperative visibility of minimally invasive implanted unicondylar knee arthroplasty makes it difficult to remove bone and cement debris, which have been reported on the surface of damaged and retrieved bearings. Therefore, the aim of this study was to analyze the influence of bone and cement particles on the wear rate of unicompartmental knee prostheses in vitro. Fixed bearing unicompartmental knee prostheses were tested using a knee-wear-simulator according to the ISO standard 14243-1:2002(E) for 5.0 million cycles. Afterwards bone debris (particle size 671 ± 262 μm) were added to the test fluid in a concentration of 5 g/l for 1.5 million cycles, followed by 1.5 million cycles blended with cement debris (particle size 644 ± 186 μm) in the same concentration. Wear rate, knee-kinematics and wear-pattern were analyzed. The wear rate reached 12.5 ± 1.0 mm³/million cycles in the running-in and decreased during the steady state phase to 4.4 ± 0.91 mm³/million cycles. Bone particles resulted in a wear rate of 3.0 ± 1.27 mm³/million cycles with no influence on the wear rate compared to the steady state phase. Cement particles, however, lead to a significantly higher wear rate (25.0 ± 16.93 mm³/million cycles) compared to the steady state phase. The careful removal of extruded cement debris during implantation may help in reducing wear rate. Bone debris are suggested to have less critical influence on the prostheses wear rate.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23417520     DOI: 10.1007/s10856-013-4883-8

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  19 in total

1.  Methodology for long-term wear testing of total knee replacements.

Authors:  P S Walker; G W Blunn; J P Perry; C J Bell; S Sathasivam; T P Andriacchi; J P Paul; H Haider; P A Campbell
Journal:  Clin Orthop Relat Res       Date:  2000-03       Impact factor: 4.176

2.  Rapid recovery after oxford unicompartmental arthroplasty through a short incision.

Authors:  A J Price; J Webb; H Topf; C A Dodd; J W Goodfellow; D W Murray
Journal:  J Arthroplasty       Date:  2001-12       Impact factor: 4.757

3.  Biotribology of alternative bearing materials for unicompartmental knee arthroplasty.

Authors:  Thomas M Grupp; Sandra Utzschneider; Christian Schröder; Jens Schwiesau; Bernhard Fritz; Allan Maas; Wilhelm Blömer; Volkmar Jansson
Journal:  Acta Biomater       Date:  2010-04-04       Impact factor: 8.947

4.  [Medial unicompartmental knee replacement using the "Oxford Uni" meniscal bearing knee].

Authors:  P R Aldinger; M Clarius; D W Murray; J W Goodfellow; S J Breusch
Journal:  Orthopade       Date:  2004-11       Impact factor: 1.087

5.  Free bone cement fragments after minimally invasive unicompartmental knee arthroplasty: an underappreciated problem.

Authors:  S M Hauptmann; P Weber; C Glaser; C Birkenmaier; V Jansson; P E Müller
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2008-05-31       Impact factor: 4.342

Review 6.  Unicompartmental knee arthroplasty: long-term results.

Authors:  R V Deshmukh; R D Scott
Journal:  Clin Orthop Relat Res       Date:  2001-11       Impact factor: 4.176

7.  Polyethylene wear in Oxford unicompartmental knee replacement: a retrieval study of 47 bearings.

Authors:  B J L Kendrick; D Longino; H Pandit; U Svard; H S Gill; C A F Dodd; D W Murray; A J Price
Journal:  J Bone Joint Surg Br       Date:  2010-03

8.  [Preclinical evaluation of coated knee implants for allergic patients].

Authors:  J Reich; L Hovy; H-L Lindenmaier; R Zeller; J Schwiesau; P Thomas; T M Grupp
Journal:  Orthopade       Date:  2010-05       Impact factor: 1.087

9.  Wear of fixed bearing and rotating platform mobile bearing knees subjected to high levels of internal and external tibial rotation.

Authors:  H M McEwen; J Fisher; A A Goldsmith; D D Auger; C Hardaker; M H Stone
Journal:  J Mater Sci Mater Med       Date:  2001 Oct-Dec       Impact factor: 3.896

10.  A comparative study of the medial St Georg sled and kinematic total knee arthroplasties. Ten-year survivorship.

Authors:  C E Ackroyd; S L Whitehouse; J H Newman; C C Joslin
Journal:  J Bone Joint Surg Br       Date:  2002-07
View more
  9 in total

1.  Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions.

Authors:  Carmen Zietz; Joern Reinders; Jens Schwiesau; Alexander Paulus; Jan Philippe Kretzer; Thomas Grupp; Sandra Utzschneider; Rainer Bader
Journal:  J Mater Sci Mater Med       Date:  2015-02-26       Impact factor: 3.896

2.  Exposure of articular chondrocytes to wear particles induces phagocytosis, differential inflammatory gene expression, and reduced proliferation.

Authors:  Michael D Kurdziel; Meagan Salisbury; Lige Kaplan; Tristan Maerz; Kevin C Baker
Journal:  J Mater Sci Mater Med       Date:  2017-05-22       Impact factor: 3.896

3.  Novel osteoconductive β-tricalcium phosphate/poly(L-lactide-co-e-caprolactone) scaffold for bone regeneration: a study in a rabbit calvarial defect.

Authors:  Hanna Pihlman; Pauli Keränen; Kaarlo Paakinaho; Jere Linden; Markus Hannula; Iida-Kaisa Manninen; Jari Hyttinen; Mikko Manninen; Outi Laitinen-Vapaavuori
Journal:  J Mater Sci Mater Med       Date:  2018-10-08       Impact factor: 3.896

4.  PMMA third-body wear after unicondylar knee arthroplasty decuples the UHMWPE wear particle generation in vitro.

Authors:  Alexander Christoph Paulus; Manja Franke; Michael Kraxenberger; Christian Schröder; Volkmar Jansson; Sandra Utzschneider
Journal:  Biomed Res Int       Date:  2015-03-19       Impact factor: 3.411

5.  Increase in the tibial slope reduces wear after medial unicompartmental fixed-bearing arthroplasty of the knee.

Authors:  Patrick Weber; Christian Schröder; Jens Schwiesau; Sandra Utzschneider; Arnd Steinbrück; Matthias F Pietschmann; Volkmar Jansson; Peter E Müller
Journal:  Biomed Res Int       Date:  2015-01-15       Impact factor: 3.411

6.  Wear Performance of Calcium Carbonate-Containing Knee Spacers.

Authors:  Ulrike Mueller; Joern Reinders; Sydney Smith-Romanski; Jan Philippe Kretzer
Journal:  Materials (Basel)       Date:  2017-07-15       Impact factor: 3.623

7.  The influence of a calcium sulphate bone void filler on the third-body damage and polyethylene wear of total knee arthroplasty.

Authors:  R M Cowie; S S Aiken; J J Cooper; L M Jennings
Journal:  Bone Joint Res       Date:  2019-03-02       Impact factor: 5.853

8.  More Anterior in vivo Contact Position in Patients With Fixed-Bearing Unicompartmental Knee Arthroplasty During Daily Activities Than in vitro Wear Simulator.

Authors:  Huiyong Dai; Nan Zheng; Diyang Zou; Zhemin Zhu; Ming Han Lincoln Liow; Tsung-Yuan Tsai; Qi Wang
Journal:  Front Bioeng Biotechnol       Date:  2021-05-20

9.  Influence of third-body particles originating from bone void fillers on the wear of ultra-high-molecular-weight polyethylene.

Authors:  Raelene M Cowie; Silvia Carbone; Sean Aiken; John J Cooper; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2016-06-16       Impact factor: 1.617

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.