Methodology for long-term wear testing of total knee replacements.

P S Walker1, G W Blunn, J P Perry, C J Bell, S Sathasivam, T P Andriacchi, J P Paul, H Haider, P A Campbell.   

Abstract

This article begins to address the validation requirements of wear testing on total knee replacements in a knee simulator. The knee simulator has four stations. The axial force is variable but reaches a maximum of 2.3 kN. Physiologic anteroposterior shear force and rotational torques are supplied to the knee. The forces and displacements are timed to coincide with those of a typical gait cycle. Kinematics of the simulator are dependent on the type of knee being tested. Tests of designs with well known clinical histories were done to 10 million cycles. The relative amounts and types of wear shown by the designs were similar to that found in their clinical histories. Wear tracks on more conforming designs were larger, and the penetration into the plastic appeared to be less. This did not necessarily mean that wear, as measured by loss of material, was reduced on conforming designs. Delamination of the plastic was achieved only after aging the tibial components. Wear particles isolated from the lubricating fluid were similar in size and shape to those isolated from in vivo specimens. However, the relative amounts of wear particle shapes were different depending on the design. At the start of the tests, all of the flexibly mounted tibial components showed more motion than after 5 million cycles, indicating that the surface of the plastic became more conforming. This study showed that knee wear similar to wear observed in vivo can be reproduced in the laboratory. The parameters and methods elucidated in this introductory study should form the basis for use in preclinical wear tests of total knee replacements.

Mesh:

Year:  2000        PMID: 10738439     DOI: 10.1097/00003086-200003000-00032

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  12 in total

1.  Radiation cross-linking in ultra-high molecular weight polyethylene for orthopaedic applications.

Authors:  Ebru Oral; Orhun K Muratoglu
Journal:  Nucl Instrum Methods Phys Res B       Date:  2007-12       Impact factor: 1.377

2.  [A novel knee endoprosthesis with a physiological joint shape. Part 1: Biomechanical basics and tribological studies].

Authors:  K-H Frosch; T Floerkemeier; C Abicht; P Adam; H Dathe; J Fanghänel; K M Stürmer; D Kubein-Meesenburg; H Nägerl
Journal:  Unfallchirurg       Date:  2009-02       Impact factor: 1.000

3.  Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions.

Authors:  Carmen Zietz; Joern Reinders; Jens Schwiesau; Alexander Paulus; Jan Philippe Kretzer; Thomas Grupp; Sandra Utzschneider; Rainer Bader
Journal:  J Mater Sci Mater Med       Date:  2015-02-26       Impact factor: 3.896

4.  The influence of third-body particles on wear rate in unicondylar knee arthroplasty: a wear simulator study with bone and cement debris.

Authors:  Christian Schroeder; Thomas M Grupp; Bernhard Fritz; Christoph Schilling; Yan Chevalier; Sandra Utzschneider; Volkmar Jansson
Journal:  J Mater Sci Mater Med       Date:  2013-02-17       Impact factor: 3.896

5.  Changes in dynamic medial tibiofemoral contact mechanics and kinematics after injury of the anterior cruciate ligament: a cadaveric model.

Authors:  Asheesh Bedi; Tony Chen; Thomas J Santner; Saadiq El-Amin; Natalie H Kelly; Russell F Warren; Suzanne A Maher
Journal:  Proc Inst Mech Eng H       Date:  2013-06-26       Impact factor: 1.617

6.  Comparison of ISO standard and TKR patient axial force profiles during the stance phase of gait.

Authors:  Hannah J Lundberg; Valentina Ngai; Markus A Wimmer
Journal:  Proc Inst Mech Eng H       Date:  2012-03       Impact factor: 1.617

7.  [A new total knee arthroplasty with physiologically shaped surfaces. Part 2: First clinical results].

Authors:  K-H Frosch; H Nägerl; D Kubein-Meesenburg; J Buchholz; J Dörner; H Dathe; O Hellerer; C Dumont; K M Stürmer
Journal:  Unfallchirurg       Date:  2009-02       Impact factor: 1.000

8.  CR TKA UHMWPE wear tested after artificial aging of the vitamin E treated gliding component by simulating daily patient activities.

Authors:  Jens Schwiesau; Bernhard Fritz; Ines Kutzner; Georg Bergmann; Thomas M Grupp
Journal:  Biomed Res Int       Date:  2014-11-20       Impact factor: 3.411

9.  A comparison between electromechanical and pneumatic-controlled knee simulators for the investigation of wear of total knee replacements.

Authors:  Abdellatif Abdelgaied; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2017-03-25       Impact factor: 1.617

10.  Standardized loads acting in knee implants.

Authors:  Georg Bergmann; Alwina Bender; Friedmar Graichen; Jörn Dymke; Antonius Rohlmann; Adam Trepczynski; Markus O Heller; Ines Kutzner
Journal:  PLoS One       Date:  2014-01-23       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.