| Literature DB >> 23149785 |
Caroline Descamps1, Mercedes Gonzalez, Edgardo Garrigo, Alejandro Germanier, Daniel Venencia.
Abstract
The computed tomography dose index (CTDI) measured with a 10 cm long pencil ionization chamber placed in a 14 cm long PMMA phantom is typically used to evaluate the doses delivered during CT procedure. For the new generation of CT scanners, the efficiency of this methodology is low because it excludes the contribution of radiation scattered beyond the 100 mm range of integration along z. The AAPM TG111 Report proposes a new measurement modality using a small volume ionization chamber positioned in a phantom long enough to establish dose equilibrium at the location of the chamber. In this work, the AAPM report was implemented. The minimum scanning length needed to obtain cumulative dose equilibrium was evaluated. The equilibrium dose was determined and compared to CTDI values informed by the CT scanner, and the dose values were confirmed with TLD measurements. The difference between doses measured with TLD and with the ionization chamber (IC) was below 1% and the repeatability of the measurements' setup was 0.4%. The measurements showed that the scanning lengths needed to reach the cumulated dose equilibrium were 450 mm and 380 mm for the central and peripheral axes, respectively, which justifies the phantom length. For the studied clinical protocols, the doses measured were about 30% higher than those informed by the CT scanner. For the new generation of CT systems with wider longitudinal detector size or cone-beam technology, the current CTDI measurements may no longer be adequate, and the informed CTDI tends to undervalue the dose delivered. It is therefore important to evaluate CT radiation doses following the AAPM TG111 methodology.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23149785 PMCID: PMC5718533 DOI: 10.1120/jacmp.v13i6.3934
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Water phantom with the ionization chamber placed in the central insert.
Figure 2a) Positioning of the charge‐collection volume chamber with the CT z=0 plane; b), c) and d) Alignment of the phantom center with the CT central axis (image center) and verification of the position of the peripheral inserts.
Figure 3Scanning lengths realized to determine the cumulative central dose.
Parameters of the clinical protocols used.
|
|
|
|
|
|
|---|---|---|---|---|
| Scanning mode | Helical | Axial | Axial | Helical |
| kVp | 130 | 130 | 130 | 130 |
| mA | 100 | 100 | 100 | 100 |
| n | 2 | 2 | 2 | 2 |
| T | 5 | 1.5 | 2.5 | 2.5 |
|
| 1 | 1 | 1 | 1.5 |
| Time per tube rotation | 1s | 1s | 1s | 1s |
Notes: , , sections, in a sequence of axial scanning or continuous table advance per rotation during helical scanning.
Figure 4TLD rod and insert to realize measurements in the water phantom.
Figure 5Comparison of doses delivered during both axial and helical scanning modes.
Figure 6Approach to equilibrium of central and peripheral axes central cumulative doses.
Fitted parameters Leq, Deq and a obtained with Origin software for both ionization chamber (IC) positions.
|
|
|
|
|---|---|---|
|
|
|
|
|
|
|
|
| α | 0.9 | 0.5 |
Planar average equilibrium dose and integral dose obtained for four sets of operating condition. Comparison between and CTDI informed by the CT scanner.
|
|
|
|
|
|
|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 9.1 | 9.3 | 9.3 | 9.3 |
| Variation (%) | 35.0% | 32.4% | 33.9% | 33.3% |
|
|
|
|
|
|
TLD results and comparison with ionization chamber (IC) measurements.
|
|
|
|
|
|
|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
|
|
| Variation (%) | 3.3% | 0.7% | 0.3% | 0.6% |
|
|
|
|
|
|
|
|
|
|
|
|
| Variation (%) | 0.7% | 0.8% | ‐4.0% | 1.1% |
|
|
|
|
|
|
|
|
|
|
|
|
| Variation (%) | 1.0% | 0.7% | ‐1.9% | 1.1% |