| Literature DB >> 23149774 |
Stephen F Kry1, Richard Popple, Andrea Molineu, David S Followill.
Abstract
Ion recombination is approximately corrected for in the Task Group 51 protocol by Pion, which is calculated by a two-voltage measurement. This measurement approach may be a poor estimate of the true recombination, particularly if Pion is large (greater than 1.05). Concern exists that Pion in high-dose-per-pulse beams, such as flattening filter free (FFF) beams, may be unacceptably high, rendering the two-voltage measurement technique inappropriate. Therefore, Pion was measured for flattened beams of 6, 10, 15, and 18 MV and for FFF beams of 6 and 10 MV. The values for the FFF beams were verified with 1/V versus 1/Q curves (Jaffé plots). Pion was also measured for electron beams of 6, 12, 16, 18, and 20 MeV on a traditional accelerator, as well as on the high-dose-rate Varian TrueBeam accelerator. The measurements were made at a range of depths and with PTW, NEL, and Exradin Farmer-type chambers. Consistent with the increased dose per pulse, Pion was higher for FFF beams than for flattening filter beams. However, for all beams, measurement locations, and chambers examined, Pion never exceeded 1.018. Additionally, Pion was always within 0.3% of the recombination calculated from the Jaffé plots. We conclude that ion recombination can be adequately accounted for in high-dose-rate FFF beams using Pion determined with the standard two-voltage technique.Entities:
Mesh:
Year: 2012 PMID: 23149774 PMCID: PMC5718527 DOI: 10.1120/jacmp.v13i6.3803
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
values (at 300 V) for conventional flattened photon beams of different energies measured on several Varian 21 EX accelerators. The mean and the maximum and minimum of the measurements are shown. Measurements were taken at a depth of 10 cm with an Exradin A‐12 chamber.
|
|
| |||
|---|---|---|---|---|
|
|
|
|
| |
| No. of accelerators | 17 | 16 | 7 | 7 |
| Mean | 1.003 | 1.003 | 1.005 | 1.006 |
| Maximum | 1.004 | 1.004 | 1.006 | 1.006 |
| Minimum | 1.002 | 1.002 | 1.004 | 1.006 |
Measured values (at 300 V) for conventional electron beams at in water measured on several accelerators with an Exradin A‐12 chamber.
|
| |||||
|---|---|---|---|---|---|
|
|
|
|
|
| |
| No. of accelerators | 30 | 26 | 4 | 7 | 19 |
| Mean | 1.013 | 1.013 | 1.015 | 1.012 | 1.014 |
| Maximum | 1.015 | 1.016 | 1.017 | 1.015 | 1.015 |
| Minimum | 1.010 | 1.011 | 1.013 | 1.010 | 1.010 |
Measured values at 300 V for FFF beams at a depth of 10 cm in water and at with three ion chambers.
|
|
|
| ||
|---|---|---|---|---|
|
|
|
|
| |
| Exradin A‐12 | 1.006 | 1.009 | 1.010 | 1.014 |
| PTW TN30013 | 1.005 | 1.008 | 1.011 | 1.013 |
| NEL 2571 | 1.008 | 1.013 | 1.015 | 1.018 |
Measured values at 300 V for high‐dose‐rate (1000 MU/min) electron beams at two depths in water using an Exradin A‐12 chamber.
|
|
| ||||
|---|---|---|---|---|---|
|
|
|
|
|
| |
|
| 1.011 | 1.011 | 1.012 | 1.011 | 1.011 |
|
| 1.007 | 1.007 | 1.006 | 1.007 | 1.007 |
Figure 1The inverse of the collected charge versus the inverse of the applied voltage for the 6 MV FFF beam and the 10 MV FFF beam for three ion chambers examined in this study (PTW, Exradin, and NEL). A linear best‐fit line through the data is included for each series.
Recombination factors at 300 V based on the two‐voltage technique (), and based on a Jaffé‐plot (1/V versus 1/Q curve).
|
|
|
| ||
|---|---|---|---|---|
|
|
|
|
| |
| Exradin A‐12 | 1.009 | 1.009 | 1.014 | 1.017 |
| PTW TN30013 | 1.008 | 1.008 | 1.013 | 1.015 |
| NEL 2571 | 1.013 | 1.011 | 1.018 | 1.020 |
Figure 2Dose profiles over the central 4 cm of large 6 MV FFF and 10 MV FFF fields. Decreased dose away from the central axis is characteristic of FFF beams.
Figure 3Measured versus dose per pulse for X‐ray beams of 21 EX accelerators (“21EX X” from Table 1), electron beams of 21 EX accelerators (“21EX e” from Table 2), FFF X‐ray beams from a Varian TrueBeam accelerator (“TB FFF X” from Table 3), and from McEwan. Error bars on the FFF TrueBeam data are based on the accuracy of the two‐voltage technique established in Table 5 (0.15%). Doses per pulse at measurement locations not at (e.g. 10 cm depth or ) were determined from the dose per pulse at and PDDs from either Varian golden beam data or clinical data.