Charge collection efficiency in ionization chambers exposed to electron beams with high dose per pulse.

R F Laitano1, A S Guerra, M Pimpinella, C Caporali, A Petrucci.   

Abstract

The correction for charge recombination was determined for different plane-parallel ionization chambers exposed to clinical electron beams with low and high dose per pulse, respectively. The electron energy was nearly the same (about 7 and 9 MeV) for any of the beams used. Boag's two-voltage analysis (TVA) was used to determine the correction for ion losses, k(s), relevant to each chamber considered. The presence of free electrons in the air of the chamber cavity was accounted for in determining k(s) by TVA. The determination of k(s) was made on the basis of the models for ion recombination proposed in past years by Boag, Hochhäuser and Balk to account for the presence of free electrons. The absorbed dose measurements in both low-dose-per-pulse (less than 0.3 mGy per pulse) and high-dose-per-pulse (20-120 mGy per pulse range) electron beams were compared with ferrous sulphate chemical dosimetry, a method independent of the dose per pulse. The results of the comparison support the conclusion that one of the models is more adequate to correct for ion recombination, even in high-dose-per-pulse conditions, provided that the fraction of free electrons is properly assessed. In this respect the drift velocity and the time constant for attachment of electrons in the air of the chamber cavity are rather critical parameters because of their dependence on chamber dimensions and operational conditions. Finally, a determination of the factor k(s) was also made by zero extrapolation of the 1/Q versus 1/V saturation curves, leading to the conclusion that this method does not provide consistent results in high-dose-per-pulse beams.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17148826     DOI: 10.1088/0031-9155/51/24/009

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  Monte Carlo-based determination of radiation leakage dose around a dedicated IOERT accelerator.

Authors:  Hamid Reza Baghani; Seyed Rashid Hosseini Aghdam; Mostafa Robatjazi; Seyed Rabi Mahdavi
Journal:  Radiat Environ Biophys       Date:  2019-04-10       Impact factor: 1.925

2.  Analysis of the long-term stability of the output of electron beams generated by the Novac11™ IORT accelerator.

Authors:  Michał Bijok; Ewelina Gruszczynska; Adam Kowalczyk; Katarzyna Sikorska; Agnieszka Walewska; Pawel F Kukolowicz
Journal:  Rep Pract Oncol Radiother       Date:  2018-08-13

3.  Current saturation in free-air ionization chambers with chopped synchrotron radiation.

Authors:  Nobuteru Nariyama
Journal:  J Synchrotron Radiat       Date:  2013-07-03       Impact factor: 2.616

4.  Ion recombination correction factors (P(ion)) for Varian TrueBeam high-dose-rate therapy beams.

Authors:  Stephen F Kry; Richard Popple; Andrea Molineu; David S Followill
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

Review 5.  Present state and issues in IORT Physics.

Authors:  Frank W Hensley
Journal:  Radiat Oncol       Date:  2017-01-27       Impact factor: 3.481

6.  Monte Carlo Simulation of Electron Beams produced by LIAC Intraoperative Radiation Therapy Accelerator.

Authors:  M Robatjazi; K Tanha; S R Mahdavi; H R Baghani; H R Mirzaei; M Mousavi; N Nafissi; E Akbari
Journal:  J Biomed Phys Eng       Date:  2018-03-01

7.  Comparing the dosimetric characteristics of the electron beam from dedicated intraoperative and conventional radiotherapy accelerators.

Authors:  Hamid Reza Baghani; Seyed Mahmoud Reza Aghamiri; Seyed Rabi Mahdavi; Mohammad Esmail Akbari; Hamid Reza Mirzaei
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

8.  Verification of electron beam parameters in an intraoperative linear accelerator using dosimetric and radiobiological response methods.

Authors:  Agnieszka Dróżdż; Martyna Waluś; Marcin Zieliński; Bożena Malesa; Marta Kruszyna-Mochalska; Katarzyna Kulcenty; Beata Adamczyk; Piotr Nowaczyk; Julian Malicki; Jacek Pracz
Journal:  Rep Pract Oncol Radiother       Date:  2021-12-30

9.  Development of an ultra-thin parallel plate ionization chamber for dosimetry in FLASH radiotherapy.

Authors:  Faustino Gómez; Diego M Gonzalez-Castaño; Nicolás Gómez Fernández; Juan Pardo-Montero; Andreas Schüller; Alessia Gasparini; Verdi Vanreusel; Dirk Verellen; Giuseppe Felici; Rafael Kranzer; Jose Paz-Martín
Journal:  Med Phys       Date:  2022-04-22       Impact factor: 4.506

10.  Dosimetric characteristics of electron beams produced by two mobile accelerators, Novac7 and Liac, for intraoperative radiation therapy through Monte Carlo simulation.

Authors:  Sergio Righi; Evis Karaj; Giuseppe Felici; Fabio Di Martino
Journal:  J Appl Clin Med Phys       Date:  2013-01-07       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.